Understanding the Laureate™ Digital Panel Meter Ohmmeter for Resistance in Ohms
The Laureate™ 1/8 DIN Digital Panel Meter Ohmmeter is ideal for high-speed, high-accuracy resistance measurements in a production environment, such as contact resistance measurements. It is factory calibrated for five jumper-selectable resistance ranges from 20.000 ohm to 200.00 kohm, plus three fixed factory-special ranges of 2.0000 ohm, 2.0000 Mohm, and 20.000 Mohm. Accuracy is ±0.01% of reading ± 2 counts. Resolution is one part in 20,000 — 0.1 milliohm on the 2 ohm range, suitable for contact resistance measurements.
Fixed vs. Jumper-Selectable Ranges
Ranges R1 through R5 (20Ω through 200kΩ) are jumper-selectable and precalibrated, letting one ordered meter be reconfigured in the field across those five ranges. Ranges R0 (2Ω), R6 (2MΩ), and R7 (20MΩ) are factory-set fixed ranges, chosen at time of order and not field-reconfigurable to another range.
Excitation Current by Range
Excitation current scales inversely with resistance range: 5 mA on the 2Ω (R0) and 20Ω (R1) ranges, 500 µA on the 200Ω (R2) range, 50 µA on 2kΩ (R3), 5 µA on 20kΩ (R4), 500 nA on both 200kΩ (R5) and 2MΩ (R6), and 80 nA on 20MΩ (R7). The applied excitation current is sensed by the meter, which operates ratiometrically and automatically compensates for any changes in excitation.
Signal and Protection Specifications
Maximum applied voltage across the unknown resistance is 100 mV; overvoltage protection is rated to 125 Vac. Span tempco is ±0.003% of reading/°C. An open sensor causes the display to flash full-scale.
2, 3, or 4-Wire Connection
4-wire hookup uses separate lead pairs for excitation and voltage sensing, eliminating IR drop across excitation leads. 3-wire hookup senses the combined resistance-plus-two-lead voltage drop, separately senses one excitation lead's drop, and subtracts twice that value — canceling lead resistance if both excitation leads are identical. 2-wire hookup measures lead resistance once by shorting the resistance during setup and subtracts it from subsequent readings, without compensating for later ambient-driven lead resistance changes.
QA Passband Mode
A deviation limit (such as 50 mΩ) is set around both sides of a setpoint. The relay closes (or opens) when the reading falls within the deviation band, and opens (or closes) when the reading falls outside it — setting up a passband around the setpoint specifically suited to contact resistance testing.
Extended DPM Capability
The optional Extended main board allows up to 180 data points to linearize and scale nonlinear resistance sources, and can display rate derived from successive readings.
Factory-Calibrated Accuracy
All resistance ranges are factory-calibrated, with calibration factors stored in EEPROM that can be scaled via software to accommodate external shunts, enabling field replacement of the signal conditioner board without recalibrating the meter. Factory recalibration is recommended annually.
Where Ohmmeter Digital Panel Meters Are Used
- Connector & Cable Assembly QA — sub-milliohm contact resistance testing on production lines using the 2Ω range and QA passband relay mode.
- Relay & Switch Manufacturing — go/no-go contact resistance screening during production testing.
- Bonding & Grounding Verification — earth bond and static-dissipative resistance checks.
- Winding & Coil Testing — motor, transformer, and solenoid resistance for quality control.
- Cable & Harness Testing — continuity and resistance verification with 3- or 4-wire compensation.
- Insulation & High-Resistance Monitoring — megohm-range measurement using the fixed 2MΩ/20MΩ ranges.
- Thermistor & Resistance-Sensor Readout — linearized display via the Extended board's custom curve capability.
Ohmmeter Digital Panel Meter Frequently Asked Questions
Why are R0, R6, and R7 factory-fixed while R1 through R5 can be jumper-reconfigured?
The three special ranges (very low 2Ω and very high 2MΩ/20MΩ) likely require internal circuit optimizations — very fine resolution for R0, very high input impedance handling for R6/R7 — that aren't practical to combine with the simple jumper-selection scheme used across the five standard mid-range options; ordering the correct fixed range up front is necessary since these can't be changed via jumpers later.
Does "one part in 20,000" resolution mean the same absolute precision on every range?
No — it's a relative resolution figure, so absolute resolution scales with the range itself: 0.1 mΩ on the 2Ω range, but 1000Ω on the 20MΩ range, since one part in 20,000 of a much larger full-scale value is itself a much larger absolute increment.
Why does excitation current drop so dramatically between the lowest range (5 mA) and the highest range (80 nA)?
Since the meter measures a voltage developed across the unknown resistance capped at 100 mV, a much higher resistance needs proportionally less current to develop a comparable, accurately measurable voltage — using the same higher current on a high-resistance range would either exceed the 100 mV ceiling or require handling a much larger voltage range than the front end is designed for.
Does the meter's documented tare functions (auto-tare and manual tare) actually apply to a resistance-measurement use case?
These are documented as standard Laureate meter features carried across the product family, originally purposed for weighing applications (subtracting an empty container's weight) — for a pure ohmmeter application, tare isn't a typical requirement, though the underlying offset-subtraction mechanism could conceivably be repurposed to zero out a known fixture or lead-resistance baseline if an application called for it.
Can the QA passband deviation limit be set to different widths on each side of the setpoint, or is it symmetric?
The documented example specifically describes a deviation limit (such as 50 mΩ) set up around both sides of the setpoint, implying a single band width applied symmetrically — an application needing asymmetric high/low limits around a target value would need to confirm whether the current firmware supports independently configured upper and lower deviation limits.
Does the 100 mV maximum applied voltage limit apply equally across all eight resistance ranges, or does it vary?
Documented specification lists 100 mV as the maximum applied voltage across the unknown resistance as a single figure spanning the full ohmmeter product — since excitation current is independently scaled per range specifically to keep the resulting voltage within a consistent, measurable window, this 100 mV ceiling is documented as the same across all eight ranges rather than varying range-by-range.
Why is overvoltage protection rated to 125 Vac when the meter's own working signal never exceeds 100 mV?
These describe different things — the 100 mV figure is the meter's own intended working excitation voltage during normal measurement, while the 125 Vac overvoltage protection rating describes the input circuitry's survival threshold against an accidental fault voltage being applied to the input terminals, a safety margin far beyond the meter's own normal operating signal level.
Can custom curve linearization correct for a resistance-based sensor whose relationship to the measured physical quantity is genuinely nonlinear?
Yes — the documented 180-point Extended board capability is specifically described as linearizing and scaling nonlinear resistance sources, meaning it can correct the displayed reading to accurately track a resistance sensor's real, nonlinear relationship to whatever physical quantity it's measuring, rather than assuming a simple linear relationship.
Does field-replacing the signal conditioner board on this ohmmeter genuinely require no recalibration of the meter itself?
Documented capability specifically confirms this — calibration factors are stored in EEPROM on the signal conditioner board itself, so swapping that board (for a different range, or as a repair) transfers its own factory calibration data along with it, meaning the meter as a system doesn't need separate recalibration purely because the board was swapped.
Does span tempco (±0.003% of reading/°C) matter much for a typical indoor QA testing environment?
In a temperature-stable indoor test environment, this drift is usually a small contributor to overall measurement uncertainty, but for applications where ambient temperature swings meaningfully — near ovens, outdoor test stations, or unconditioned production floors — this spec becomes more relevant to account for when evaluating overall measurement uncertainty against a tight resistance tolerance, particularly on the lowest ranges where absolute resolution is already fine enough that small percentage drifts can matter.
Precision Reference Resistor & Calibration Standard Questions From the Field
What's the practical difference between a resistor's tolerance and its TCR when selecting a calibration reference?
Documented guidance specifically distinguishes these as separate specifications: tolerance is the maximum percentage deviation from nominal resistance value as delivered, typically measured at 25°C, while TCR (temperature coefficient of resistance) specifies how much that value drifts per degree of temperature change away from the reference point — a resistor can have tight initial tolerance but still drift meaningfully if its TCR is high, or vice versa.
How much can a resistor's value actually shift due to TCR over a realistic temperature swing?
Documented worked examples specifically quantify this: a resistor with a TCR of +100 ppm/°C changes +0.1% total over a 10°C swing and +1% total over a 100°C swing, while a lower-grade 50 ppm/°C resistor experiencing a 100°C rise shows roughly 0.5% change — meaning TCR-driven drift is a genuinely calculable, sometimes significant error source distinct from the resistor's initial tolerance.
Why do the highest-precision reference resistors specifically use 4-wire (Kelvin) construction rather than standard 2-terminal packaging?
Documented calibration-resistor manufacturers specifically note that 4-wire technology enables loss-free measurement of current and voltage separately, meaning the reference resistor's own terminal and lead resistance doesn't corrupt the precision value being used as a calibration standard — this is the same underlying Kelvin principle that makes 4-wire measurement essential for the device under test, applied to the reference itself.
Does using a very tight-tolerance reference resistor (e.g., ±0.005%) automatically mean the calibration performed with it achieves that same accuracy?
No — documented metrology guidance specifically notes that the measuring instrument itself must also have sufficient resolution and traceable calibration to make the reference resistor's tight tolerance meaningful; a documented example specifically illustrates using acceptance limits tightened to account for the measuring instrument's own uncertainty (e.g., 0.009% acceptance limits for a 0.01% tolerance resistor measured with 0.001% instrument uncertainty), since the reference alone doesn't guarantee the overall calibration's achieved accuracy.
What does it mean for a calibration certificate to be "traceable to national standards," and does every reference resistor need one?
Documented calibration-resistor suppliers specifically describe offering certificates compliant with recognized accreditation schemes (such as ISO 17025) specifically to establish a documented, traceable chain connecting a given reference resistor's certified value back to national measurement standards — whether a specific application genuinely needs this level of documented traceability (versus a simpler manufacturer calibration certificate) depends on the regulatory or quality-system requirements of that specific application.
Is a lower TCR always worth paying for, or does it depend on the application's actual temperature exposure?
Documented guidance specifically frames TCR selection around the resistor's role in the overall system error budget — for applications with genuinely wide or uncontrolled temperature swings, a lower TCR meaningfully reduces drift-driven error, but for a stable, temperature-controlled lab environment where ambient swings are already small, paying for an unnecessarily tight TCR may add cost without proportional real-world accuracy benefit.
Why did one documented case specifically choose a resistor made of multiple integrated elements rather than a single resistive element for a demanding reference application?
A documented calibration standard case specifically selected a resistor integrating 11 elements into a single housing, explaining this allowed for lower TCR and longer-term drift stability than could be achieved from any single resistive element alone — illustrating that construction technique, not just raw material choice, genuinely affects achievable long-term stability for the most demanding reference applications.
Does a reference resistor's rated tolerance remain valid indefinitely, or does long-term drift eventually require recalibration regardless of the original tolerance?
Documented guidance specifically notes long-term stability (drift per year) as a separate specification from initial tolerance — high-grade laboratory reference resistors are documented with stability figures as tight as ±0.001%/year, but this figure explicitly acknowledges that some drift accumulates over time even for the most stable references, which is why periodic recalibration against a higher-tier standard remains standard metrology practice rather than treating a reference resistor's original certified value as permanently valid.

























Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.