Understanding the Laureate™ Digital Panel Meter for True RMS AC Voltage & Current
The Laureate™ 1/8 DIN Digital Panel Meter for True RMS is a 60,000-count meter with exceptional performance at half the price of a benchtop meter. It measures true RMS voltage or current of AC signals with considerable waveform distortion, such as rectifier outputs or waveforms chopped by an SCR, TRIAC, or transistor circuit.
Fast Response and Overrange Behavior
True RMS readings arrive 0-16 ms after completion of one input signal cycle, allowing anomalies to be detected and alarmed before they become expensive problems. Crest factor (Vp/Vrms) is 3.0 at full-scale Vrms of 20,000 counts and Vp of 60,000 counts. The meter begins flashing overrange at 132% of full scale Vrms (available crest factor 2.27), with meaningful readings continuing to 212% of full scale (42,433 counts, crest factor 1.414) for sinusoidal signals. Maximum RMS input should never exceed 600V or 5A; ETL certification covers a maximum of 300 Vrms.
Accuracy and Input Resistance
Accuracy is 0.03% of full scale with the standard 1 MΩ input resistance, signals from DC to 5 kHz, and amplitudes down to 0.5% of full scale. A 10 MΩ input resistance version is available as a factory special, but decreases maximum frequency from 5 kHz to 1 kHz on three voltage ranges. Maximum applied voltage is 600 Vac for the 2V/20V/200V/600V ranges, but only 35 Vac for the 0.2V range; current protection is 25x for 2 mA, 8x for 20 mA, 2.5x for 200 mA, 1x for 5A.
AC/DC Coupling and Capture Rate
AC or DC coupling is jumper selectable — AC coupling suits applications like measuring ripple on a DC power supply. Multiple integral cycles are averaged for signals above 50/60 Hz; a single cycle is captured for signals from 3 Hz to 50/60 Hz; below 3 Hz and at DC, capture rate is every 333 ms.
Use With Current Transformers
High common-mode rejection allows stable readings with current shunts on the high side of the line. Built-in 5A input capability lets 5A current transformer outputs be applied directly, with no stepdown transformer needed; readings are easily scaled for the transformer ratio via front-panel pushbuttons.
Synchronizing Two Motor Generators
Successful synchronization requires the two generator voltages be close, the two frequencies identical, and the waveshapes in phase. Two Laureate AC RMS voltmeters (or one multiplexed via toggle switch) measure the two voltages; a Laureate dual-channel counter measures both frequencies to six-figure accuracy, toggling between channels or displaying A-B or A/B directly; a Laureate phase meter measures phase angle.
Factory-Calibrated Accuracy
All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM. Field replacement of the signal conditioner board doesn't require recalibrating the meter. Factory recalibration is recommended annually.
Where True RMS AC Digital Panel Meters Are Used
- Motor & VFD Current Monitoring — accurate current readout despite PWM switching harmonics via direct or CT-based input.
- Generator Synchronization & Paralleling — three-instrument voltage/frequency/phase matching before connecting to a shared bus.
- UPS & Backup Power Systems — true RMS readout of square-wave or stepped inverter output waveforms.
- SCR/TRIAC-Controlled Heating & Lighting — accurate RMS readout of phase-chopped control waveforms.
- Substation & Switchgear Instrumentation — direct CT-fed current readout without a separate stepdown transformer.
- DC Power Supply Ripple Testing — AC-coupled measurement of ripple riding on a DC bus.
- High Voltage Line Monitoring — voltage-transformer-fed readout for measurements exceeding 600V.
True RMS AC Digital Panel Meter Frequently Asked Questions
What actually happens to the meter's reading once a signal exceeds 132% of full scale?
At 132% of full scale Vrms, the meter begins flashing to indicate overrange — at that point available crest factor has dropped to 2.27 from the rated 3.0 at full scale, meaning the meter's headroom for peak excursions above the RMS value is shrinking as the signal approaches its true limits.
Can the meter still give a meaningful reading above the 132% overrange flash point?
Yes, to a further point — for sinusoidal signals specifically, meaningful readings with rated resolution continue up to 212% of full scale Vrms, where available crest factor is 1.414 (the theoretical minimum for a pure sine wave); beyond that there's no crest factor margin left for any peak above the RMS value.
Why is maximum applied voltage only 35 Vac on the 0.2V range when it's 600 Vac on the higher ranges?
The 0.2V range's front-end circuitry is optimized for very small signals and correspondingly has much less voltage headroom before component ratings are exceeded — applying anywhere near 600V to that sensitive low range would risk immediate damage, so its documented maximum applied voltage is set much lower than the higher-voltage ranges' protection circuitry allows.
Does choosing the 10 MΩ input resistance factory special always make sense if higher impedance is generally better?
Not necessarily — the documented tradeoff specifically cuts maximum frequency from 5 kHz down to 1 kHz on three of the voltage ranges, so this option is worth choosing specifically when an application's source genuinely needs the higher input impedance (to avoid loading a high-impedance source) and doesn't need signal content above 1 kHz; for signals needing the full 5 kHz bandwidth, the standard 1 MΩ version is the better fit.
Why does AC coupling matter specifically for measuring ripple on a DC supply?
AC coupling removes the DC component of a signal before RMS calculation, isolating just the AC ripple riding on top of the DC bus — without it, the large DC offset would dominate the reading and the small ripple component of interest would be effectively invisible.
Does the meter's capture rate stay constant regardless of input signal frequency?
No — capture rate is tied to the signal itself: multiple integral cycles are averaged above 50/60 Hz, a single cycle is captured between 3 Hz and 50/60 Hz, and below 3 Hz (down to DC) the capture rate becomes a fixed 333 ms interval rather than being tied to signal cycles at all.
Why can a 5A current transformer be connected directly to this meter without a stepdown transformer?
The meter's built-in 5A input range is specifically designed to accept the standard secondary output of common current transformers directly — since 5A is a widely used CT secondary rating, this avoids the extra cost and complexity of an intermediate stepdown stage.
Why does the motor generator synchronization application need three separate instruments instead of one meter reading everything?
Each of the three conditions for safe synchronization — matched voltage, matched frequency, and matched phase — requires its own dedicated measurement technique: two AC RMS voltmeters (or one multiplexed) for voltage, a dual-channel counter for six-figure frequency accuracy, and a separate phase meter, since combining all three differing measurement types into a single instrument isn't how this documented application is structured.
Does the dual-channel frequency counter used in synchronization need to be manually toggled to compare the two generators, or can it show the difference directly?
Both options are documented as available — a front-panel key toggles the reading between Channel A and Channel B individually, but the meter can also display frequency A minus frequency B, or frequency A divided by frequency B, directly without needing to toggle back and forth to compare the two.
Do the current protection multipliers (25x for 2 mA, down to 1x for 5A) mean the lower current ranges can survive a proportionally larger overcurrent event?
Yes — these multipliers describe how far above each range's own full-scale current the input protection can tolerate before damage; the lower-current ranges have proportionally more overcurrent headroom (25x on the 2 mA range) because an accidental overcurrent relative to that small full-scale rating is still a modest absolute current, while the 5A range's 1x rating reflects that its full-scale current is already substantial, leaving little additional margin before the protection circuitry's own limits are reached.
Current Transformer Burden & Saturation Questions From the Field
What is CT "burden," and why does it matter for accurate current measurement?
Documented technical guidance specifically defines burden as the total impedance connected to a CT's secondary winding — including the connected meter or relay plus the resistance of the connecting leads — expressed in ohms or volt-amperes (VA); exceeding a CT's rated burden causes a voltage drop across that impedance that introduces real measurement error, since the CT's accuracy specification only holds within its rated burden range.
Why is it specifically dangerous to open-circuit a CT's secondary while primary current is still flowing?
Documented safety guidance specifically identifies this as life-threatening — with the secondary open, there's no opposing magnetomotive force to counteract the primary current's magnetic field, causing core flux to increase dramatically and secondary voltage to rise to potentially thousands of volts; a documented example calculation for a 1000:5 CT shows open-circuit secondary voltage could exceed 1500V RMS.
What's the documented correct procedure before disconnecting an instrument from an energized CT secondary?
Documented practice specifically requires short-circuiting the CT secondary before removing any connected instrument, per NEMA STD 1 and IEC 61869-2 — using shorting blocks or test switches with integral shorting contacts is the documented method, ensuring the secondary circuit is never left open while primary current is flowing.
Does excessive CT burden cause a gradual accuracy decline, or a more abrupt failure mode?
Documented analysis specifically describes CT saturation as a genuine loss of linearity once the core magnetically saturates — rather than a smooth, gradual accuracy decline, a saturated CT "goes blind," failing to accurately track the primary current right at the moment it matters most (such as during a fault), which is documented as a specifically dangerous failure mode for protective relaying applications.
How can excessive CT burden actually be detected or verified in a real installation?
Documented troubleshooting guidance specifically recommends checking for a distorted, "flat-topped" secondary waveform using an oscilloscope as a sign of saturation, alongside independently verifying the calculated total burden (VA) doesn't exceed the CT's rating at expected current levels, and inspecting connections for loose or corroded terminals that could contribute unexpected resistance to the burden.
Does lead wire length and gauge genuinely factor into total CT burden, or is burden mostly about the connected instrument itself?
Lead resistance is documented as a real, calculable contributor to total burden — a documented example specifically calculates #14 AWG copper leads contributing roughly 1.27Ω round-trip resistance, which must be subtracted from the CT's total allowable burden impedance budget alongside the connected instrument's own burden, meaning long or thin lead runs can meaningfully eat into a CT's available burden margin.
Is there a documented safety margin recommendation for how close actual burden should get to a CT's rated maximum?
Yes — documented guidance specifically recommends keeping total burden under roughly 80% of rated VA, providing margin specifically to prevent saturation during high-current fault conditions, when the CT's demands on its burden capacity are at their most severe and least forgiving.
Does connecting a CT's output directly to this meter's built-in 5A range change any of the standard CT burden considerations?
No — the same underlying burden principles documented for any CT installation still apply when feeding a 5A CT output directly into this meter's 5A range; the meter's own input burden (impedance) needs to be accounted for within the CT's total rated burden budget the same way any other connected instrument's burden would be, alongside lead resistance and any other devices sharing that CT's secondary circuit.
The crest factor (Vp / Vrms) is 3.0 at full-scale Vrms of 20,000 counts and a Vp of 60,000 counts. The digital panel meter starts to flash overrange at 132% of full scale Vrms, at which point the available crest factor is 60,000 / 26,400 = 2.27. Meaningful readings with rated resolution continue to be obtained up to 212% of full scale Vrms (42,433 counts) for sinusoidal signals, at which point the available crest factor is 1.414. For safety reasons, the maximum RMS input signal should never exceed 600V or 5A. ETL certification is for a maximum voltage of 300 Vrms.






















Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.