Digital Panel Meters for Quadrature Encoder Input and Bidirectional Position or Rate Applications P/N L61007QD

Digital Panel Meters for Quadrature Encoder Input and Bidirectional Position or Rate Applications

Price: $644.00
  • P/NL61007QD
- +

Features

  • Accepts low-level differential or single-ended 5V logic level signals from shaft encoders, linear encoders, incremental encoders or optical encoders
  • 6-digit scalable display for position, length or rate
  • Programmable for position, angle or rate
  • Maximum pulse rates of 250 kpulses/sec at X1, 125 kpulses/sec at X2, 62.5 kpulses/sec at X4
  • Display update rate up to 25/s
  • Zero channel input
  • Digital span adjust from 0 to ±999,999, zero adjust from -999,999 to +999,999
  • Front panel scalable: to ±999,999 for use with current shunts
  • 1/8 DIN size with bright red or green 0.56" (14.2mm), high LED digits
  • Transducer excitation output, 5, 10, 12, or 24 Vdc (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • Operating temperature from -40°C to 70°C  (-40°F to 158°F)
  • Wide choice of Plug-in-Play options:
    - 2 or 4 relays, mechanical or solid state, for alarm or control (isolated)
    - 1 or 2 Analog output, 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V (isolated)
    - Communications: Ethernet, WiFi, USB, RS232, RS485 (isolated)
    - Extended DPM includes set up for scaled position or rate, but not forsimultaneous position and rate

The Laureate™ 1/8 DIN Digital Panel Meters for bidirectional position, length or angle measurement

with the Standard counter main board accepts the A & B quadrature signals from shaft encoders or linear encoders to provide a highly accurate, scaled display of position, length, or angle in engineering units, such as ft, cm or degrees. The A & B quadrature signals are 90° out of phase, and their phase relationship determines whether up counts (+) or down counts (-) are counted. The meter totalizes the counts and then scales the total in software for display and control. A zero index signal, or Z signal, can be added as a third input to the A & B signals.

Bidirectional Rate Measurement

The Laureate™ quadrature meter with the Extended counter main board can be set up for scaled position or rate, but not for simultaneous position and rate. For example, it can display the speed of a moving slab in ft/sec. The display and control output update rate for position or rate is normally set to a maximum of 25/sec, as determined by a user-programmable gate time.

Quadrature Meter Capabilities

  • One, two or four quadrature transitions may be counted at a maximum combined rate of 250 kHz and be mathematically scaled for display in engineering units from -999,999 to +999,999. Maximum pulse rates are 250 kpulses/sec at X1, 125 kpulses/sec at X2, and 62.5 kpulses/sec at X4.
  • Single-ended input signals up to 5V or balanced line driver signals up to 10V are accommodated by the quadrature signal conditioner board. Anti-jitter circuitry eliminates errors due to vibration of the encoder.
  • An excitation output from the meter can be used to power the encoder, thus avoiding the need for an external powr supply. Jumper selectable levels are 100 mA at 5V or 120 mA at 10V.
  • A zero index pulse, if available from the encoder, is interpreted by the meter as a zero reference for an integral number of revolutions of a rotary encoder, or as the home position of a linear encoder. It is used by the meter for initializing and correcting any cumulative pulse count errors. Special circuitry corrects for width of the zero index pulse.
  • In the event of a power failure, the latest total can be stored in non-volatile memory and be used as the starting point for counting when power resumes. Power fail save or zero index capabilities are alternate meter setup choices.

Unidirectional Position and Rate

If the counts are only for one direction, for example for extrusions measured by an encoder wheel, only the encoder's "A" channel can be used and be output to an Extended Laureate totalizer. This model accommodates very high pulse rates up to 1 MHz, and unlike the quadrature meter, it can simultaneously track rate and total. It can also be programmed for batch control, which allows it to simultaneously track rate, batch (or current) total, and grand total or number of batches.

Laureate Digital Panel Meters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated digital panel meters. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via a front panel pushbutton command or control signal at the rear connector, or be transmitted as serial data.

Two rear panel control Inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 meter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Modular Design for Maximum Flexibility at Minimum Cost

All boards are isolated from meter and power grounds. Optional Plug-in-Play boards for communications and control include Ethernet, WiFi, serial communication boardsdual or quad relay boards, and an analog output board. Laureates may be powered from 85-264 Vac or optionally from 12-32 Vac or 10-48 Vdc. The display is available with bright red or green 0.56" (14.2mm) high LED digits. The 1/8 DIN case meets NEMA 4X (IP65) specifications from the front when panel mounted. Any setup functions and front panel keys can be locked out for simplified usage and security. A built-in 5, 10, 12, or 24 Vdc excitation supply can power transducers, eliminating the need for an external power supply. All power and signal connections are via UL / VDE / CSA rated screw clamp plugs.

The Laureate™ Series features modular design with up to 7 isolated plug-in boards, applicable to all Laureate 1/8 DIN Digital Panel Meters.

Schematic for Digital Panel Meter

Modular Hardware

The design of the Laureate™ Series is modular for maximum flexibility at minimum cost. All boards are isolated from meter and power grounds. The base configuration for a panel meter or counter consists of a main module (with computer and plug-in display boards), a power supply board, and a signal conditioner board. Optional plug-in-play boards include an isolated setpoint controller board, an isolated analog output board, and an isolated digital interface board. Modular design and a choice of plug-in options allow the Laureate to be customized for a broad range of applications from simple monitoring to control and computer interface. There can be up to five plug-in boards in a 1/8 DIN Laureate.

Dual Board sets

Connecting Laureate Digital Panel Meters to a Local Area Network (LAN)

Up to 30 Laureate Digital Panel Meters and/or LT Transmitters can be configured for RS485 and daisy-chained to an LT Transmitter using Laurel’s High Speed Ethernet-to-RS485 converter board for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for Digital Panel Meters

Laureate Digital Panel Meters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

Display
Readout 6 LED digits, 7-segment, 14.2 mm (.56"), red or green.
Display Range -999,999 to +999,999, XXXXEX notation beyond 999,999
Zero Adjust -999,999 to +999,999
Span Adjust 0 to ±999,999
Indicators Four LED lamps
Inputs
Types Differential or single-ended quadrature
Transitions Monitored Inputs X1, X2 or X4 (A,B and Z)
Max Pulse Rate 250 kpulses/sec at X1, 125 kpulses/sec X2, 62.5 kpulses/sec at X4
Position Error No error contributed by meter
Differential High Threshold +200 mV
Differential Low Threshold -200 mV
Differential Limits -11V to +14V
Single-ended High Voltage 2.5V to 10V
Single-ended Low Voltage -1V to +1V
Input Resistance, Typ. 17 kOhm
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Quadrature Position Mode
Zero Adjust -999,999 to +999,999
Span Adjust 0 to ±999,999
Quadrature Rate Mode
Freq. Technique Inverse period
Conversion Time Gate time + 30 ms + 0-2 signal periods
Gate Time Selectable 10 ms to 199.99 s
Time Before Zero Out Selectable 10 ms to 199.99 s
Output & Display Update Same as conversion time
Time Base Accuracy Calibrated to ±2 ppm
Power Supply Boards (one required)
Voltage, standard 85-264 Vac or 90-300 Vdc
Voltage, optional 12-32 Vac or 10-48 Vdc
Frequency DC or 47-63 Hz
Power Consumption (typical, base meter) 1.2W @ 120 Vac, 1.5W @ 240 Vac, 1.3W @ 10 Vdc, 1.4W @ 20 Vdc, 1.55W @ 30 Vdc, 1.8W @ 40 Vdc, 2.15W @ 48 Vdc
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Analog Output Boards (one optional)
Output levels 4-20 mA, 0-20 mA, 0-10V, -10 to +10V (single-output option)
Current compliance 2 mA at 10V ( > 5 kΩ load)
Voltage compliance 12V at 20 mA (< 600 Ω load)
Scaling Zero and full scale adjustable from -99999 to +99999
Resolution 16 bits (0.0015% of full scale)
Isolation 250V rms working, 2.3 kV rms per 1 min test
(dual analog outputs share the same ground)
Relay Output Boards (one optional)
Dual magnetic relays 2 Form C, 10A max, 440Vac or 125Vdc max, 2500VA or 300W
Quad magnetic relays 4 Form A (NO), 10A max, 440Vac or 125Vdc max, 2500VA or 300W
Dual solid state relays 2 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Quad solid state relays 4 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Relay commons Isolated commons for dual relays or each pair of quad relays
Relay isolation 250V rms working, 2.3 kV rms per 1 minute test
Relay latching modes Latching or non-latching
Relay active modes Active on or off, active high or low
Hysteresis modes QA passband mode, split hysteresis, span hysteresis
Communication Boards (one optional)
Board selections RS232RS485 with dual RJ11 connectors, RS485 with dual RJ45 connectors, USB, Ethernet, USB-to-RS485 gatewayEthernet-to-RS485 gateway, WiFi with built-in antenna plus USB & RS485, WiFi with external antenna plus USB & RS485
Protocols Laurel Custom ASCII (serial), Modbus RTU (serial), Modbus TCP (Ethernet or WiFi)
Digital addresses 247 (Modbus), 31 (Laurel ASCII),
Isolation 250V rms working, 2.3 kV rms per 1 min test
Environmental
Operating temperature -40°C to 70°C (-40°F to 158°F)
Storage temperature. -40°C to 85°C (-40°F to 185°F)
Relative humidity 95% at 40°C, non-condensing
Protection NEMA-4X (IP-65) when panel mounted
Mechanical
Enclosure 1/8 DIN, high impact plastic, UL 94V-0, color: black
Mounting 1/8 DIN panel cutout required: 3.622" x 1.772" (92 mm x 45 mm).
Dimensions 4.68" x 2.45" x 5.64" (119 mm x 62 mm x 143 mm) (W x H x D)
Maximum panel thickness 4.5 mm (0.18")
Tightening Torque - Connectors Screw terminal connectors: 5 lb-in (0.56 Nm)
Tightening Torque - Pawls Digital Panel Meter Case Pawls: 5 lb-in (0.56 Nm)
Weight of base meter 210 g (7.4 oz) typical (DPM, counter, timer, 6-digit remote display)
Weight of option boards 30 g (1.0 oz) typical per board (analog output, relay output, communications)
General
Programming Methods Four front panel buttons or via Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options include using the front panel buttons, the free Instrument Setup Software, or a hardware jumper.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Laureate™ 1/8 DIN Case For Laureate Digital Panel Meters, Counters, Timers & Remote Displays

Laurel panel meter case

Key Features

  • Meets 1/8 DIN Standard.
  • Installs from front of panel.
  • Short depth behind the panel: only 4" (102 mm) plus connectors.
  • Understated 0.157" (4 mm) thick bezel.
  • Meets NEMA 4X (IP-65) for high-pressure wawshdon when panel mounted.
  • Screw clamps connectors meet VDE / IEC / UL / CSA safety standards.
  • Rugged GE Lexan® housing material.
  • Safety certified per EN 61010-1.
CE CertifiedRoHs CertifiedETL Certified

Dimensions

Mechanical specifications of Laureate digital panel meters and electronic counters

Maximum panel thickness: 4.5 mm (0.18")
Weight of base meter:
210 g (7.4 oz) typical (DPM, counter, timer, 6-digit remote display)
Weight of option boards: 30 g (1.0 oz) typical per board (analog output, relay output, communications)
Tightening Torque - Connectors: Screw terminal connectors: 5 lb-in (0.56 Nm)
Tightening Torque - Pawls: Digital Panel Meter Case Pawls: 5 lb-in (0.56 Nm)
Dimensioned CAD assembly drawings in EPRT, STEP, x_t. dwg, pdf file formats: Laureate-meter-case.zip (zipping prevents browser from opening CAD files as text files).

Panel Mounting

Mechanical Panel Mounting GuideSlide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel. The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.

Turning each screw counterclockwise loosens the pawl and retracts it into its well. This position allows installed meter to be removed from their panel, or new meters to be installed in a panel. Do not remove the screws from their pawls. Doing so would cause the screw and pawl to fall off and likely get lost. Do not overtighten so as not to damage the plastic parts.

 

Using Quadrature for Cutting to Length
Using a Laureate quadrature panel meter to cut material to length Controlling the repetitive cutting of material to length is an excellent application of the Laureate quadrature meter. The quadrature encoder shares the shaft of a sensing wheel, whose rotation corresponds to lineal displacement of material. The meter compares the displacement reading against setpoint information, and then uses its dual relays to first slow down and then cut the material.
Using Quadrature for X-Y Positioning
Using a Laureate quadrature panel meter for X-Y positioning Accurate X-Y position or rate can be obtained from two shaft encoders, which convert linear position to quadrature signals as a shaft turns. In addition to serving as a display, each Laureate can use its optional dual relay setpoint capability for closed loop control. It can also transmit data via RS232, RS485, or a 4-20 mA analog signal.
Using Quadrature to Monitor a Drilling Operation
Using a Laureate quadrature meter to monitor an oil drilling operation Quadrature can be used to track position and vertical drilling speed of the bit in an oil drilling operation. A shaft encoder is rotated by a cable that moves with the drilling shaft. In this application, the same encoder signal is applied to a Laureate quadrature meter for position, and to a second quadrature meter for rate. Both meters can be scaled to read out in appropriate engineering units, such as feet and inches per minute, and can be alarmed. A Laureate 6-digit remote display could be added to read out peak rate.

 

 

CAL-Digital

Certificate of Calibration

$65.00

DLS-XLOG2

XLog2 Data logging Software

$495.00

IPC

Splashproof Cover

$55.00

CON01

CON01 Connector

$75.00

CBL01

RS232 Cable for Meters

$35.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL05

USB Data Cable for Meters

$47.00

CBL06

USB-to-RS485 Adapter Cable

$47.00

CBL07

USB Programming & Data Cable

$47.00

CBL08

RS485 Splitter Cable

$33.00

CBL6

6-foot Power Cable

$41.00

CBL12

12-foot Power Cable

$47.00
Ordering Guide
Part Number as Configured: L61007QD
Price as Configured: $644.00

Click on the Option Board Links for More Product Information

Base Item
$169.00
Display Color
$0.00
$0.00
$41.00
Extended allows up to 180 data points for custom curve linearization and a rate derived from consecutive readings.
$41.00
Power (Isolated) $75.00
$115.00
Relay Output (Isolated)
$0.00
$101.00
$70.00
$128.00
$96.00
Analog Output (Isolated)
$0.00
$115.00
$172.00
Data Interface (Isolated)
$0.00
$81.00
$109.00
$135.00
$81.00
$135.00
$237.00
$259.00
$182.00
$204.00
$171.00
$193.00
$204.00
$226.00
Signal Input (Isolated)
$123.00
Part Number as Configured:
L61007QD
Price as Configured:
$644.00
Quantity:
- +
Extended Price:
$644.00

Understanding 1/8 DIN Digital Panel Meters for Quadrature Encoder Input and Bidirectional Position or Rate

1/8 DIN Digital Panel Meters for Quadrature Encoder Input and Bidirectional Position or Rate measurement are specialized instruments used in various industrial and automation applications. These devices offer precise monitoring and display of position, speed, or rate of movement in systems that use quadrature encoders.

What Are 1/8 DIN Digital Panel Meters?

The term DIN refers to the Deutsches Institut für Normung, a German standards organization. In the context of Digital Panel Meters, DIN size indicates the dimensions of the Digital Panel Meters' front panels. The "1/8 DIN" size means that the Digital Panel Meters have dimensions of approximately 96mm x 48mm (3.78" x 1.89"). This compact size makes them suitable for various control panels where space is a concern.

Digital Panel Meters are electronic devices that display input signals in digital formats. These Digital Panel Meters are used to monitor parameters such as voltage, current, temperature, and in this case, position or rate derived from quadrature encoders. The digital readouts provide easy-to-read and precise indications of the measured parameters, which are crucial for monitoring and control in industrial settings.

Quadrature Encoder Input

A quadrature encoder is a type of incremental encoder used to measure the position and direction of a rotating shaft. It generates two output signals (A and B), which are 90 degrees out of phase with each other. By interpreting these signals, the direction of rotation (clockwise or counterclockwise) can be determined.

The Quadrature Encoder Input capability of Digital Panel Meters allows them to read these signals and accurately determine the position or rate of movement. These Digital Panel Meters decode the pulses from the encoders to calculate the position of the shaft or object. This functionality is crucial in applications requiring precise motion control, such as in CNC machines, robotic arms, and conveyor systems.

Bidirectional Position or Rate Measurement

The Bidirectional Position capability refers to the Digital Panel Meters' ability to track the movement in both directions, typically forward and backward. This is essential in applications where the movement is not unidirectional, and accurate position tracking in both directions is required.

Similarly, Bidirectional Rate Measurement allows the devices to measure the speed or rate of movement in both directions. This can be critical in applications like motor control, where the speed and direction of rotation need to be monitored and controlled accurately.

Applications of 1/8 DIN Digital Panel Meters with Quadrature Encoder Input

These Digital Panel Meters are widely used in various industries due to their precision and reliability. Some common applications include:

  • CNC Machines: Accurate position measurement is vital for the precise cutting and shaping of materials.
  • Robotics: Monitoring the position and speed of robotic arms ensures accurate and repeatable movements.
  • Conveyor Systems: Tracking the position and speed of conveyors helps in material handling and sorting processes.
  • Motor Control: Controlling the speed and direction of motors in various industrial machines.

Key Features and Benefits

  • High Accuracy: Provide precise measurement and display of position or rate.
  • Versatility: Can be used with various types of quadrature encoders.
  • Bidirectional Measurement: Track both forward and reverse movements.
  • Compact Size: The 1/8 DIN size makes them suitable for use in space-constrained control panels.
  • Easy Integration: Can be integrated into existing control systems with minimal modifications.

Where Are 1/8 DIN Digital Panel Meters for Quadrature Encoder Input and Bidirectional Position or Rate Used?

1/8 DIN Digital Panel Meters designed for quadrature encoder input and bidirectional position or rate measurement are specialized instruments widely used across various industries for precise monitoring, control, and feedback in motion and position control systems.

1. Industrial Automation

In industrial automation, accuracy and reliability are paramount. 1/8 DIN Digital Panel Meters with quadrature encoder input are often used to monitor and control the position and speed of machinery components. These Digital Panel Meters provide real-time data, enabling precise adjustments to ensure the smooth operation of automated systems. Typical applications include:

  • CNC Machines: Monitoring the position of the cutting tool relative to the workpiece with high precision.
  • Robotics: Tracking the movement and orientation of robotic arms or conveyors.
  • Assembly Lines: Ensuring that components are positioned accurately for assembly, reducing defects and improving efficiency.

2. Motion Control Systems

Motion control systems, particularly in the manufacturing and automotive industries, rely heavily on quadrature encoders for accurate position and speed feedback. 1/8 DIN Digital Panel Meters play crucial roles in these systems by converting the quadrature encoder signals into readable position or rate data. Key applications include:

  • Servo Motors: Monitoring and controlling the position and speed of servo motors used in various machinery.
  • Conveyor Systems: Tracking the speed and position of items on conveyors, crucial for synchronized operations.
  • Packaging Machines: Ensuring that products are correctly positioned for wrapping, labeling, or sealing.

3. Process Control in Chemical and Pharmaceutical Industries

In process control applications, particularly in the chemical and pharmaceutical sectors, precise measurement and control of the position and speed of pumps, mixers, and other critical equipment are essential. 1/8 DIN Digital Panel Meters are used to provide accurate feedback, enabling operators to maintain the desired process parameters, ensuring product quality and consistency.

  • Dosing Systems: Monitoring the position and speed of dosing pumps to ensure accurate chemical or ingredient delivery.
  • Mixing and Blending: Controlling the speed and position of mixing equipment to achieve consistent product quality.

4. Renewable Energy Sector

In the renewable energy sector, particularly in wind and solar power generation, 1/8 DIN Digital Panel Meters are used to monitor and control the position of components like solar panels or wind turbine blades. These Digital Panel Meters help in optimizing the orientation of panels or blades for maximum energy capture.

  • Solar Tracking Systems: Monitoring the position of solar panels to ensure they are optimally oriented towards the sun.
  • Wind Turbines: Controlling the angle of turbine blades to maximize energy output under varying wind conditions.

5. Test and Measurement

In laboratories and test environments, 1/8 DIN Digital Panel Meters are used to monitor the position or speed of various components in test rigs, ensuring accurate and reliable measurements. These Digital Panel Meters are often used in:

  • Material Testing Machines: Monitoring the position and speed of testing heads during tensile, compression, or fatigue tests.
  • Environmental Chambers: Controlling the position of components or samples within chambers for consistent testing conditions.

Conclusion

1/8 DIN Digital Panel Meters for quadrature encoder input and bidirectional position or rate measurement are versatile tools used across multiple industries. Their ability to provide precise, real-time data on position and speed makes them indispensable in applications ranging from industrial automation to renewable energy.

Quadrature Encoder Input Digital Panel Meter Frequently Asked Questions

What makes a quadrature encoder signal different from a single-channel pulse signal?

A quadrature encoder produces two channels (A and B) offset by 90 degrees from each other, rather than a single pulse train. This phase relationship lets the meter determine not just how many pulses have occurred, but which direction the shaft is turning, based on which channel leads the other.

How does the meter determine direction from the A and B channels?

The meter tracks the sequence of state transitions between the two channels. Because A and B change state at different, predictable points relative to each other, the order in which they transition reveals whether the shaft is rotating clockwise or counterclockwise, which the meter uses to increment or decrement its position count accordingly.

What is the index (or marker/Z) channel, and is it required?

The index channel produces a single pulse once per revolution and is used to establish an absolute reference position, useful for homing or zeroing a position count. It's optional on many applications that only need relative position or rate tracking, but is important where an absolute reference point matters, such as machine homing.

Can this meter be scaled to display position in real-world units like inches or millimeters?

Yes. These meters are typically user-scalable, so the raw quadrature count can be converted and displayed directly in linear or rotational engineering units based on the encoder's counts-per-revolution and the mechanical relationship between the encoder and the moving component being tracked.

What happens to the count if the shaft reverses direction mid-motion?

A properly functioning quadrature meter correctly decrements the count when direction reverses, since the quadrature decoding is based on the actual sequence of A/B transitions rather than assuming a fixed direction — this bidirectional tracking is the core advantage of quadrature over a simple single-channel pulse count.

What alarm and output options are available for position or rate monitoring?

These meters commonly support programmable high/low alarm relays tied to position or rate, an isolated analog output, and serial communications such as RS-232 or RS-485, allowing the meter to trigger a local alarm or feed position/rate data to a PLC or motion controller.

What is the maximum encoder frequency these meters can accurately decode?

Maximum input frequency varies by model, but it's important to confirm the meter's rated maximum against the actual pulse rate expected from the encoder at full operating speed, since exceeding the meter's rated frequency can result in missed or undercounted pulses.

Does the position count get retained if the meter loses power?

Many models store the accumulated position or count in non-volatile memory, so the last known position is retained through a power interruption, though depending on the application, a homing routine using the index channel may still be recommended after power-up to confirm absolute position accuracy.

Can this meter be used with encoders from different manufacturers, or does it require a specific encoder brand?

These meters are generally designed to accept standard quadrature output signals (TTL, differential line driver, open collector, etc.) rather than being tied to a specific encoder brand, so compatibility depends on matching the encoder's electrical output type to the meter's input specification rather than the manufacturer itself.

Is isolation available between the encoder input and the meter's other outputs?

Isolated input and output configurations are commonly available and help protect the encoder signal from noise introduced by relays, analog outputs, or communication activity elsewhere in the same meter or panel, which is particularly relevant in electrically noisy motor-drive environments.

Quadrature Encoder Questions From the Field

Why does my position count run backward instead of forward when the shaft moves in the expected direction?

This is one of the most commonly reported quadrature setup issues, and the fix is almost always a simple channel swap — reversing the physical A and B channel connections (or inverting one channel's logic in software, where that option exists) flips the sensed direction to match reality. If swapping A and B makes the problem worse instead of better, the specific encoder may use a different phase relationship convention than initially assumed, in which case both possible transition-table configurations should be tried.

Why does my encoder count correctly in one direction but double-count in the other direction?

This has been documented in real troubleshooting cases, and a frequent root cause is the index (marker) pulse being wired into one of the A or B differential input channels instead of its own dedicated index input — this corrupts the quadrature decoding specifically when the index pulse occurs, which shows up as inconsistent or doubled counting in one direction but not the other. Verifying the marker/index signal is connected to its own correct input, separate from A and B, resolves this.

Why is my encoder occasionally reporting an incorrect number of counts right around when the index pulse occurs?

This is a documented and fairly subtle issue reported in motion control forums, where extensive high-resolution logging was needed to catch it happening only during certain direction changes near the index pulse. Careful review of the exact A/B/index transition sequence at the moment of the anomaly, rather than assuming a simple wiring fault, was needed to isolate this class of problem — it's a good example of why an oscilloscope trace, not just a multimeter check, is often necessary for intermittent quadrature issues.

My encoder signal looks clean and stable on a scope, but my counts still run backward — what else could it be?

This has been traced in real cases to a configuration setting rather than a wiring or signal quality issue — for example, a timer/counter peripheral configured for a different encoder mode (such as Hall sensor XOR decoding) instead of standard quadrature decoding can produce counts that appear to run backward or behave strangely even though the physical signal itself is completely clean. Double-checking the counting mode configuration, not just the wiring, is worth doing when a clean signal still produces wrong-direction counts.

Should I use an oscilloscope or a simple voltmeter to troubleshoot an encoder signal problem?

An oscilloscope is strongly preferred and commonly recommended as the primary diagnostic tool for encoder troubleshooting, since it's the only practical way to see the actual timing relationship and phase relationship between the A and B channels — a voltmeter can only confirm a signal is present, not whether its phasing and quadrature relationship are correct.

Can long cable runs cause an encoder to lose counts even if the encoder itself is working correctly?

Yes — this is a commonly cited cause of missed counts, since long cable runs can attenuate the signal to the point where the counting device can no longer reliably distinguish valid pulse transitions from noise. Confirming the encoder's output type (differential line driver signals tolerate longer runs better than single-ended outputs) and cable length against the manufacturer's specifications is a standard first check for count-loss issues.

Does electrical noise near motor drives commonly cause quadrature miscounting?

Yes, this is a frequently reported issue in industrial installations, and it's often difficult to isolate since noise sources can be hard to pin down without proper test equipment. Checking for noise directly at the counting device's input terminals, and confirming the encoder's differential signal type (which offers better noise immunity than single-ended output) is compatible with the wiring in use, are standard troubleshooting steps.

How do I know whether my encoder uses the standard A-leads-B or the opposite B-leads-A convention for forward direction?

This isn't standardized across all encoder manufacturers, so it has to be determined empirically for a specific encoder rather than assumed — the practical approach reported in the field is to test the encoder's actual behavior in each direction and adjust the decoding logic or physical wiring to match what's observed, rather than relying on a single universal state table that may not apply to every encoder.