Understanding the Laureate™ 1/8 DIN Panel Meters for RTD Temperature
The Laureate™ 1/8 DIN Panel Meters for RTDs are factory calibrated for four Resistance Temperature Detector types: 100-ohm platinum (Pt100) with DIN alpha of 0.00385, 100-ohm platinum (Pt100) with ANSI alpha of 0.003902, 10-ohm copper with alpha of 0.00427, and 120-ohm nickel with alpha of 0.00672 — not Pt1000, which is not offered on this meter. The entire span of each RTD type is presented in a single range, with RTD type, unit of measure, and resolution (1°, 0.1°, or 0.01°) all selectable from the front panel or serial interface. Typical accuracy is better than ±0.1°C (±0.2°F).
Excitation Current and Wiring
The meter applies an excitation current of 196 µA for Pt100 and Ni120 sensors, or 5 mA for Cu10 sensors. RTD hookup can be via 2, 3, or 4 wires. In 4-wire hookup, separate lead pairs apply excitation current and sense the voltage drop across the RTD ratiometrically, so IR drop across the excitation leads isn't a factor. In 3-wire hookup, the meter senses the combined voltage drop across the RTD plus two excitation leads, separately senses the drop across one excitation lead, and subtracts twice that value — effectively canceling lead resistance and compensating for ambient temperature changes, provided both excitation leads are identical. In 2-wire hookup, lead resistance is measured once by shorting the RTD during setup and then subtracted from all readings, but subsequent changes in lead resistance due to ambient temperature won't be compensated.
Calibration Standards and Accuracy
Calibration follows IEC 751 (ITS-90) for Pt100 DIN, NIST Monograph 126 for Pt100 ANSI, and DIN 43760 for Ni120. Max error at 25°C for Pt100 is ±0.04°C (±0.07°F) ±0.01% of reading. Span tempco is ±0.003% of reading/°C, and zero tempco is ±0.03 deg/deg. Sensor lead resistance compensation is rated for up to 10Ω per conductor in 2-wire mode and up to 100Ω per conductor in 3- and 4-wire mode.
Open Sensor and Overvoltage Protection
Overvoltage protection is rated to 125 Vac. If a sensor becomes open, the display flashes full-scale — an unambiguous indication distinct from a valid reading.
Three Operating Modes
- 4-20 mA Transmitter — with the optional analog output board, the meter serves as an isolated 4-20 mA transmitter, scaled to the linearized °C or °F display and tracking the meter's high read rate (up to 60 readings/second at 60 Hz) — beneficial for closed-loop and PID control.
- Fast Controller — dual contact or solid-state relay boards enable fast, accurate ON/OFF closed-loop control with multiple setpoint modes and programmable hysteresis/time delays.
- Supervisory Monitor — relay outputs alarm or shut down when limits are exceeded, using a band deviation mode that alarms whenever the reading is a set number of counts above or below setpoint. Latching outputs remain in alarm until reset via front panel, serial interface, or rear connector.
Factory-Calibrated Accuracy
All RTD types are factory-calibrated, with calibration factors stored in EEPROM that can be scaled via software to accommodate external shunts, enabling field replacement of the signal conditioner board without recalibrating the meter. Factory recalibration is recommended annually.
Where RTD Panel Meters Are Used
- Pharmaceutical & Bioprocessing — tight-tolerance RTD monitoring for sterilizers, bioreactors, and cold-chain storage where Pt100 precision matters.
- HVAC & Building Automation — chilled water, condenser, and duct temperature monitoring using low-cost Ni120 or Cu10 sensors.
- Cryogenic & Materials Testing — RTD linearity advantage across moderate sub-zero ranges for cold-chamber and materials science work.
- Semiconductor & Cleanroom Processing — precise process-temperature control with 4-wire RTD connections over long cable runs.
- Food & Beverage Processing — pasteurization and CIP/SIP cycle monitoring where repeatable, drift-resistant RTD accuracy supports food-safety compliance.
- Power Generation & Rotating Equipment — bearing and winding temperature monitoring with supervisory alarm/shutdown relay modes.
- Laboratory & Calibration Benches — NIST/IEC-traceable reference-grade RTD readout for verifying other temperature instrumentation.
RTD Panel Meter Frequently Asked Questions
Why does this meter support Pt100 but not Pt1000, when Pt1000 is common in many other applications?
This meter's signal conditioner is specifically factory-calibrated for the four listed RTD types (Pt100 DIN, Pt100 ANSI, Ni120, Cu10) — Pt1000 simply isn't one of the offered calibrated options, so an application specifically requiring Pt1000 would need a different meter configuration or an external transmitter converting Pt1000 to a signal this meter does support.
Why does the excitation current differ so much between Pt100/Ni120 (196 µA) and Cu10 (5 mA)?
Copper RTDs have a much lower nominal resistance (roughly 9-19Ω across their range) compared to Pt100's 100-390Ω or Ni120's 120-380Ω — a much higher excitation current is needed on the low-resistance Cu10 element to develop a comparably measurable voltage signal for the meter's front end to resolve accurately.
Does 3-wire compensation work as well as 4-wire, or is there a real accuracy difference?
3-wire compensation is documented as effective specifically when both excitation leads are identical in resistance, since the technique relies on subtracting twice the measured single-lead voltage drop — if the two excitation leads aren't well matched, some residual error remains that true 4-wire connection (which doesn't depend on lead-matching) avoids entirely.
Can 2-wire connection be used for a precision application if I compensate for lead resistance once during setup?
The one-time compensation (shorting the RTD and measuring lead resistance during setup) corrects for the lead resistance present at that moment, but it specifically doesn't track subsequent changes in that lead resistance due to ambient temperature shifts on the cable run — for installations with meaningful ambient temperature swings along the cable path, 3- or 4-wire connection is the more robust choice.
Why do the three different RTD calibration standards (IEC 751, NIST Monograph 126, DIN 43760) matter if they're all just "RTD calibration"?
Different RTD types and even different alpha values within the same base metal were historically standardized by different bodies — Pt100 DIN (alpha 0.00385) follows IEC 751, Pt100 ANSI (alpha 0.003902) follows NIST Monograph 126, and Ni120 follows DIN 43760 — using the correct reference standard for the specific sensor type installed is what ensures the meter's internal resistance-to-temperature conversion actually matches that sensor's real behavior.
What does the difference between max lead resistance limits (10Ω for 2-wire vs. 100Ω for 3-/4-wire) actually mean for cable run length?
This reflects how much residual, uncompensated lead resistance each wiring method can tolerate before it meaningfully affects accuracy — since 3- and 4-wire connections actively compensate for lead resistance, they can tolerate ten times more total lead resistance (and therefore much longer cable runs or thinner wire gauges) than 2-wire connection, which has comparatively little built-in compensation margin.
Does the band deviation alarm mode work the same way on this RTD meter as on other Laureate temperature meters?
Yes — band deviation mode alarms whenever the reading moves a set number of counts above or below setpoint, covering deviation in either direction from a single configured band, the same underlying relay alarm mechanism used across the Laureate temperature meter family regardless of sensor type.
Can the RTD meter's analog output be used to drive a PID controller without introducing significant additional lag?
The analog output specifically tracks the meter's own high read rate (up to 60 readings/second at 60 Hz power), so it adds minimal lag on top of the meter's own measurement rate — documented as specifically beneficial for closed-loop and PID control applications needing timely feedback.
If a latched alarm relay trips on this RTD meter, does it clear automatically once temperature returns to normal?
No — a latched output remains in the alarm condition until deliberately reset via front panel pushbutton, the serial interface, or the rear connector, even after the underlying condition returns to normal, ensuring a brief excursion doesn't go unnoticed.
Does zero tempco (±0.03 deg/deg) mean the meter's zero reading drifts by 3% of the actual temperature per degree of ambient change?
No — zero tempco describes drift in the meter's own zero-point calibration as ambient temperature around the instrument changes, expressed as degrees of reading error per degree of ambient temperature change, not as a percentage of the measured temperature itself; it's a separate, generally small error source layered on top of the meter's stated overall accuracy spec.
RTD Self-Heating & Excitation Current Questions From the Field
Why does using a higher excitation current on an RTD actually make the temperature reading less accurate?
This is a well-documented phenomenon called self-heating — an RTD is fundamentally a resistor, and current flowing through it dissipates power as heat, which warms the element and biases the reading high. Documented technical analysis specifically shows that a lower excitation current directly reduces this self-heating error, which is why high-accuracy measurement applications favor lower currents even though that also produces a smaller, harder-to-resolve signal.
Is there a documented rule of thumb for how much self-heating error is acceptable versus problematic?
Yes — documented guidance based on standards such as IEC 60751 specifies that self-heating error must not exceed 25% of the sensor's own rated tolerance specification, giving a concrete, standards-based ceiling for how much excitation-driven heating is acceptable before it becomes a meaningful contributor to overall measurement error.
Does self-heating error appear instantly when excitation current is applied, or does it take time to develop?
Documented real-world testing specifically notes that self-heating effects can take several minutes to settle to their final value — meaning a reading taken immediately after excitation current is first applied may not yet reflect the full self-heating error that will eventually develop, which matters for both calibration procedures and any application using intermittent rather than continuous excitation.
Can I quantify self-heating error myself if I know my RTD's specifications?
Yes — a documented formula expresses self-heating temperature change as ΔT = I²R × θ, where I is excitation current, R is the RTD's resistance, and θ is the sensor's self-heating factor (typically expressed in °C or K per mW, and specified by the sensor manufacturer) — plugging in a specific sensor's rated self-heating factor and the actual excitation current and resistance gives a concrete estimate of expected error for that specific combination.
Does self-heating error stay constant across an RTD's full measurement range, or does it change with temperature?
It changes, because self-heating error depends on the RTD's resistance at the measurement point, and RTD resistance itself changes with temperature — a documented example for a Pt100 at 850°C (roughly 390Ω) shows measurably different self-heating behavior than the same sensor's behavior near 0°C (100Ω), since higher resistance at the same excitation current produces more power dissipation and thus more self-heating.
Is there a fundamental tradeoff between minimizing self-heating and getting a strong, easily measurable signal?
Yes, and this tradeoff is specifically documented as inherent to RTD measurement design — lower excitation current reduces self-heating error but produces a smaller voltage signal that's harder to resolve accurately against amplifier noise and offset, while higher excitation current improves signal strength at the direct cost of more self-heating bias. Some documented approaches use adaptive or variable excitation current specifically to balance this tradeoff across an RTD's full resistance range.
Does the RTD's physical construction and mounting affect how much self-heating error actually shows up in the final reading?
Yes — documented analysis specifically notes that the magnitude of resulting self-heating error depends not just on how much heat is generated internally, but on how well that heat can be dissipated to the surrounding environment, which in turn depends on the RTD's internal construction and how it's physically mounted or immersed. An RTD in good thermal contact with a fast-moving fluid dissipates self-heating more effectively than one in still air or poor thermal contact, for the same excitation current.
Is self-heating error something that needs to be actively corrected, or can it just be minimized by choosing a low excitation current and ignored?
Documented practice varies by required precision — historically, the conventional approach has simply been to keep excitation current as low as practical while still maintaining a sufficiently strong signal, treating self-heating as something to minimize by design choice rather than actively correct after the fact. However, documented research specifically describes both linear and nonlinear correction techniques as an alternative path to further improve accuracy in applications where minimizing current alone isn't sufficient.
























Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.