Understanding the Laureate™ 1/8 DIN Panel Meters for DC Voltage and Current
Laureate™ 1/8 DIN Panel Meters for DC voltage and DC current are programmable DC panel meters with a DC signal conditioner board, combining high accuracy with a high read rate and a wide range of isolated Plug-and-Play output option boards for computer interface and control. DC meter accuracy is 0.01% of reading ± 2 counts.
DC Voltmeter Operation
Selected by jumpers, DC voltmeter operation provides six full-scale DC voltage ranges from ±200.00 mV (10 µV resolution) to ±600.0V (100 mV resolution). The 200.00 mV and 2.0000V ranges provide high input impedance of 1 GΩ to minimize loading on the voltage signal being measured; the 20V, 200V, 300V, and 600V ranges use 10 MΩ input resistance.
DC Ammeter Operation
Selected by jumpers, DC ammeter operation provides four full-scale DC current ranges from ±2.0000 mA (0.1 µA resolution) to ±5.000 A (1 mA resolution). The 5.000 A range measures the IR drop across a built-in 10 milliohm current shunt — input resistance drops accordingly across the current ranges, from 100Ω on the 2 mA range down to 0.01Ω on the 5A range.
Noise Rejection
Common-mode rejection (CMR) is 130 dB from DC to 60 Hz, and normal-mode rejection (NMR) is 90 dB at 50/60 Hz with minimal filtering — both specifications matter for measuring small DC signals in electrically noisy industrial environments where ground potential differences or line-frequency interference could otherwise corrupt a low-level reading.
Maximum Signal and Overcurrent Protection
Maximum applied voltage is 600 Vac for the 20, 200, and 300V ranges, and 125 Vac for other ranges. Overcurrent protection ratios are 25x for the 2 mA range, 8x for 20 mA, 2.5x for 200 mA, and 1x for 5A — reflecting the very different fault tolerance margins built into each current range's protection circuitry.
Extended Capabilities
The optional Extended Laureate computer board enhances the meter by displaying rate derived from successive readings and enabling highly accurate custom curve linearization — for example, calculating liquid volume or flow rate in a horizontal cylindrical tank using levels from a 4-20 mA transmitter. Up to 180 data points can be input for spline-fit linearization.
Tare, Filtering, and Peak/Valley Capture
Two tare functions are available: auto-tare (triggered by grounding an input line via external pushbutton) and manual tare (entered via front panel or Instrument Setup Software). Three digital filtering modes — unfiltered, batch average (16 conversions), and adaptive moving average (8 selectable time constants from 80 ms to 9.6 seconds) — help maintain stable readings in electrically noisy environments. Peak and valley values are automatically captured at up to 60 conversions per second and can be displayed or transmitted as serial data.
Factory-Calibrated Accuracy
All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM that can be scaled via software to accommodate external shunts, enabling field replacement of the signal conditioner board without recalibrating the meter. Factory recalibration is recommended annually.
DC Voltage & Current Panel Meter — Additional Technical Questions
Why does input impedance differ so much between the low voltage ranges and the higher voltage ranges?
The two lowest voltage ranges (200.00 mV and 2.0000V) use 1 GΩ input impedance specifically to minimize the meter's own loading effect on a sensitive low-level signal source, while the higher voltage ranges (20V and above) use 10 MΩ — a lower impedance that's still high enough to avoid meaningful loading at those signal levels, without requiring the more expensive circuitry needed for gigaohm-level input impedance across the board.
What does the 130 dB common-mode rejection specification actually protect against?
Common-mode rejection describes the meter's ability to reject a voltage that appears identically on both input signal leads relative to ground — such as a ground potential difference between the meter and a remote signal source — without that common voltage corrupting the actual differential measurement. 130 dB from DC to 60 Hz means the meter strongly rejects this class of interference across a wide frequency range including DC ground offsets and line-frequency noise.
Why does overcurrent protection differ so dramatically between the mA ranges and the 5A range?
The lower current ranges (2 mA up to 200 mA) use higher-value, more current-limited internal circuitry, which can tolerate a proportionally larger overcurrent fault (up to 25x rated current on the 2 mA range) before failing. The 5A range's protection is rated for only 1x its own current, reflecting that a range already handling significant current has much less margin before a fault becomes damaging.
How does auto-tare differ from manual tare in a DC voltage or current application?
Auto-tare captures whatever the meter is currently reading as an offset the moment an external pushbutton grounds a designated input line — useful for zeroing out a baseline reading in the field without needing a computer connection. Manual tare instead lets a specific numeric offset be entered directly via the front panel or Instrument Setup Software, useful when the desired tare value is already known rather than needing to be captured live.
When would the adaptive moving average filter be preferable to a fixed batch average filter?
The adaptive filter is designed to reduce noise-driven reading variation during steady-state conditions while still responding quickly when a genuine, significant signal change occurs — briefly switching to its shortest time constant to follow the change, then reverting back. A fixed batch average filter (16-conversion average) doesn't have this adaptive behavior, so it will smooth out both noise and genuine rapid changes equally, which can make it less suitable for applications needing fast response to real signal transitions.
Can the meter be scaled to display a completely different unit than volts or amps, like PSI or gallons?
Yes — scaling from -99,999 to +99,999 with any decimal point position lets the underlying voltage or current reading be displayed directly in whatever engineering unit is relevant to the actual sensor or transducer connected, such as PSI from a pressure transducer's voltage output, using scale-and-offset, two-point, or system-level calibration methods.
What does peak and valley capture actually add beyond the normal displayed reading?
At up to 60 conversions per second, the meter can capture the true maximum (peak) and minimum (valley) values that occurred during a measurement period — useful for catching a brief voltage spike or current dip that would be missed if only the final settled reading were recorded, and these captured values can be reviewed via the front panel, a rear-connector signal, or transmitted as serial data.
Does the meter's excitation output share a common reference with the measurement input?
No — the isolated 5, 10, 12, or 24 Vdc excitation output is isolated to 50 Vdc from signal ground, keeping the power supplied to an external transducer electrically separate from the measurement circuit itself, which helps avoid ground loop issues when powering an external sensor.
What's the difference between ratiometric operation and standard operation for a DC voltage measurement?
Ratiometric operation, which is jumper-selectable, uses the meter's own excitation voltage as the reference for the analog-to-digital converter, so any drift or variation in the excitation level is automatically compensated for in the reading — this is specifically useful for bridge-type or potentiometer-type sensors where the measured signal is inherently proportional to the excitation voltage rather than being an independent absolute value.
Can custom curve linearization correct for a genuinely nonlinear sensor, or only fine-tune an already-linear one?
It's specifically designed for genuinely nonlinear correction — up to 180 user-entered data points are spline-fit into segments and downloaded to the meter via the Extended board, letting the displayed reading track a nonlinear relationship (such as tank volume vs. level in an irregularly shaped vessel) rather than only applying a simple linear scale-and-offset correction.
DC Current Measurement & Shunt Resistor Questions From Engineering Sources
Why does my current reading drift upward or downward the longer the circuit runs at high current, even though nothing else changed?
This is a well-documented shunt resistor phenomenon called self-heating — as current flows through a shunt's resistance, it dissipates power as heat, and that heat changes the shunt's actual resistance value according to its temperature coefficient (TCR). Documented examples show a shunt with a 100 ppm/°C TCR operating 50°C above ambient can drift by roughly 0.5%, with drift approaching 1% at 100°C rise — often larger than the shunt's own rated tolerance. Allowing the shunt to reach thermal steady-state before trusting a precision reading, or derating the shunt to roughly 60-70% of its rated continuous current, are documented mitigations.
Why does a non-Kelvin (2-wire) connection to a current shunt read systematically low?
This is a specifically documented and quantified error source: a non-Kelvin connection includes the resistance of the connecting leads themselves in the measurement, and documented examples show this can add 1-3 milliohms of unwanted lead resistance, producing readings that come in roughly 5-15% low compared to a proper 4-wire (Kelvin) connection. For precision current measurement or currents above roughly 10A, a genuine 4-terminal Kelvin-sensed shunt connection is the documented recommendation rather than a simple 2-terminal connection.
Why is my low-current reading (well below the range's full scale) noisier or less trustworthy than the resolution spec would suggest?
This is a documented and specific limitation distinct from the meter's own quoted resolution: at low currents relative to a shunt's full-scale rating, the actual voltage developed across the shunt becomes very small, and thermal EMF plus amplifier offset drift in the measurement chain can dominate over the theoretical resolution — a documented example shows a meter technically capable of 0.2 mA resolution actually only trustworthy to about ±10 mA at very low trickle currents. This is a fundamental limitation of low-level shunt-based measurement, not necessarily a fault in a specific meter.
What is thermal EMF, and why does it specifically matter for very low-level DC voltage or current measurements?
Thermal EMF is a small, unwanted voltage generated whenever dissimilar metals meet at a junction with a temperature difference across it (a thermocouple-like effect occurring incidentally in ordinary wiring and connectors) — and it's specifically documented as a limiting factor for very sensitive resistance and low-level DC measurements, since this stray voltage can be comparable in size to the actual signal being measured. Documented compensation techniques include measuring the offset at zero current/signal and subtracting it, or reversing current polarity and averaging two readings to cancel out the thermal EMF component.
Does increasing test current always improve current measurement accuracy?
Not universally — this is a documented tradeoff: while a larger test or measurement current does increase the developed signal (improving resolution relative to a fixed measurement noise floor), pushing current too high increases power dissipation and self-heating in the shunt or device under test, which changes its resistance and increases thermal-EMF-related error. Selecting a shunt and operating current where the nominal load sits around 40-70% of the shunt's full-scale rating is a documented sizing guideline that balances these competing effects.
Should I expect a shunt-based ammeter to need periodic recalibration, or is factory calibration permanent?
Periodic recalibration is documented as standard practice even for high-precision shunts — environmental changes, mechanical stress, and aging are all cited as causes of long-term drift, with annual calibration against a traceable current reference documented as a common recommendation for lab-grade or precision applications, similar in spirit to any other precision measurement component.
Could corrosion or a loose connection at the shunt introduce measurement error even if the shunt itself is undamaged?
Yes — documented shunt troubleshooting guidance specifically flags checking for corrosion or loose connections as a source of unpredictable added resistance, distinct from the shunt element's own condition. This added, unpredictable resistance can introduce measurement error that looks like drift or inaccuracy in the shunt itself but actually originates at the connection point.
Is a shunt-based current measurement approach always the right choice, or are there situations where a different current sensing technology is better documented as more suitable?
Shunt-based sensing is documented as generally preferred for currents up to roughly 100A on lower-voltage rails, valued for precision and stability, while alternative approaches like Hall-effect sensing offer different tradeoffs — inherent galvanic isolation and the ability to measure both AC and DC, but with documented disadvantages including lower bandwidth, output drift over temperature, and self-heating limitations of their own at high currents. The right choice depends on which specific tradeoffs (isolation, bandwidth, precision, self-heating behavior) matter most for a given application.






















Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.