Understanding the Laureate™ 1/8 DIN Panel Meters for Process and Ratiometric Applications
The Laureate™ 1/8 DIN Panel Meters for digital process applications are a cost-effective solution for process signals such as 4-20 mA, 0-10V, or 0-5V. Full-scale voltage input ranges from ±200 mV to ±600V and current ranges from ±2 mA to ±5A are jumper selectable, all precalibrated at the factory so recalibration isn't needed when changing ranges or signal conditioners.
Ratiometric (Potentiometer Follower) Mode
The meter can be set to absolute or ratiometric mode by software selection. In ratiometric mode, the meter tracks a ratio of the applied excitation voltage and is unaffected by changes in that excitation voltage — providing exceptional accuracy of 0.01% of reading ± 2 counts. This capability is used with the meter's 5V or 10V excitation output for load cells and Wheatstone bridges, and with the 5V excitation output for potentiometers that track a wiper's position along a resistive element. Linear potentiometers measure linear displacement (inches, mm); rotary potentiometers measure angular displacement (degrees, radians).
Powering Two-Wire Transmitters
The isolated 24 Vdc, 50 mA excitation output standard on all Laureate meters is designed for powering two-wire, 4-20 mA transmitters — the same two wires apply voltage and carry the output current. Inside the meter, the 4-20 mA current is dropped across a 10-ohm resistor, setting up a 40-200 mV voltage that's sensed and scaled to engineering units.
Custom Curve Linearization and Rate
The optional Extended main board allows exceptionally accurate custom curve linearization — up to 180 data points can be entered into a spreadsheet, with the system creating multiple nonlinear spline-fit segments for much better accuracy than linear segments alone. This supports readouts like volume in irregularly shaped tanks based on measured liquid level or pressure, with altimeters and thermistors as further applications. The Extended board also displays rate based on successive readings — for example, flow rate based on changes in liquid level or static pressure in a tank; since only the already-linearized readings are compared to determine rate, the raw input itself can be nonlinear.
Peak Detection for Destructive Testing
Destructive testing is a well-suited application for this meter: peak readings are automatically captured at up to 60 per second while the display updates at a legible 3.5 readings per second. The peak reading can be recalled at the push of a button or transmitted via RS232 or RS485. The meter provides isolated 10 Vdc power for up to four strain gauges and can be scaled to read out directly in engineering units from -99,999 to +99,999.
Accuracy Distinction: Absolute vs. Ratiometric
Error at 25°C is 0.01% of full scale ± 2 counts for absolute measurements, but 0.01% of reading ± 2 counts for ratiometric measurements — reflecting that ratiometric mode's accuracy tracks the actual signal level rather than the full-scale range, since excitation-voltage variation is already compensated for mathematically.
Factory-Calibrated Accuracy
All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM that can be scaled via software to accommodate external shunts, enabling field replacement of the signal conditioner board without recalibrating the meter. Factory recalibration is recommended annually.
Where Process & Ratiometric Panel Meters Are Used
- Chemical & Process Manufacturing — reading 4-20 mA transmitters for tank level, line pressure, or flow, with custom curve linearization correcting for irregularly shaped vessels.
- Material Testing Labs — peak-capture readout for destructive testing of wire, cable, fasteners, and composites, powered directly by the meter's excitation output for strain gauge fixtures.
- Plastics & Extrusion — potentiometer-based position feedback for draw rolls, die adjustment, or web tension, using ratiometric mode to stay accurate despite excitation drift.
- Water & Wastewater Treatment — level-to-volume conversion in storage and clarifier tanks, and rate display for fill/drain monitoring derived from successive level readings.
- Machine Tool & Motion Control — rotary or linear potentiometer position feedback on slides, gates, and dampers where absolute position, not just switch state, is needed.
- HVAC & Building Automation — damper and valve position indication from potentiometer-equipped actuators, alongside 4-20 mA transmitter readout for duct pressure or airflow.
- OEM Instrumentation Panels — a single field-configurable meter (process or ratiometric, absolute or excitation-tracking) standardized across multiple product lines to simplify sourcing and panel design.
Process & Ratiometric Panel Meter Frequently Asked Questions
Why does ratiometric mode need the meter's own excitation output specifically, rather than any external power source?
Ratiometric operation works by using the same excitation voltage that powers the sensor (potentiometer, bridge, or load cell) as the reference for the meter's own analog-to-digital converter — this specific pairing is what lets excitation variations cancel out mathematically. An external, independently-sourced power supply for the sensor would break that reference relationship and reintroduce excitation-variation error.
Can this meter track a rotary potentiometer's angular position as easily as a linear potentiometer's displacement?
Yes — the same ratiometric follower principle applies to both; the difference is purely in how the meter is scaled (degrees or radians for rotary displacement, versus inches or mm for linear displacement) rather than any difference in the underlying measurement technique.
What determines whether custom curve linearization is worth using versus a simple two-point scale-and-offset?
Linearization is specifically valuable when the relationship between the input signal and the desired reading is genuinely nonlinear — such as level-to-volume in an irregularly shaped tank — where a simple linear scale-and-offset would introduce real error across the range. For a genuinely linear relationship, the added complexity of entering up to 180 calibration points wouldn't provide meaningful benefit over simpler two-point scaling.
Does the "rate from successive readings" feature need the input signal itself to be linear?
No — this is specifically documented as an advantage of pairing rate calculation with linearization: since rate is calculated by comparing already-linearized readings against each other, the raw input signal feeding the meter can be nonlinear, with the linearization step correcting it before the rate calculation ever sees it.
Why is peak capture rate (60/s) so much faster than the visible display update rate (3.5/s)?
This is intentional — a genuinely fast peak event (such as the instant of failure in a destructive test) needs to be caught by internal sampling at up to 60 readings per second, but a display updating that fast would be illegible to a human operator. The meter captures peaks internally at full speed while keeping the visible display at a comfortably readable rate, with the captured peak recallable on demand.
How many strain gauges can this meter's excitation output power simultaneously for a peak-capture testing setup?
Up to four, using the meter's isolated 10 Vdc excitation output — the same excitation-sharing arrangement used for multi-load-cell weighing platforms applies here for multi-gauge testing setups.
Why does absolute-mode accuracy reference full scale while ratiometric-mode accuracy references the actual reading?
In absolute mode, the meter measures an independent voltage or current against its own internal reference, so its error is tied to the range's full-scale value regardless of where in that range the actual signal falls. In ratiometric mode, the measurement is inherently a ratio of signal-to-excitation, so accuracy naturally scales with the actual reading rather than the nominal full-scale range.
Can this meter be used with a potentiometer that has more than 3 wire connections, like a 4-wire or 6-wire pot?
The standard ratiometric potentiometer follower application described for this meter uses a straightforward 3-wire connection (excitation, excitation return tied to signal return, and wiper signal) — applications needing additional sense wires to compensate for lead resistance in a potentiometer circuit specifically would be a more specialized configuration beyond this standard setup.
Does custom curve linearization need to be redone if I later replace the sensor with a different unit of the same type?
Since the 180-point linearization is specific to the actual calibration data entered (which reflects the real behavior of the specific sensor and physical setup it was measured against, such as an actual tank's geometry), replacing the sensor could introduce enough variation that the linearization should be re-verified, particularly for high-accuracy applications, even if the replacement is nominally the same model.
If I only need approximate rate information, is there a simpler alternative to the Extended board's rate-from-successive-readings feature?
The rate-from-successive-readings feature specifically requires the Extended main board; there isn't a documented simpler built-in alternative on the Standard board for deriving rate from a process signal's changing value over time — for basic process display without rate calculation, the Standard board configuration is sufficient and less costly.
Potentiometer Position Sensor Questions From the Field
Why does my potentiometer-based position reading come in consistently lower than the true position?
This is a well-documented phenomenon called loading error: when the wiper is connected to a circuit whose input resistance isn't considerably higher than the potentiometer's own resistance, current flows out through the wiper arm, effectively "robbing" current from the lower portion of the resistive element and causing the reading to come in low. Using a much higher input impedance on the receiving circuit relative to the potentiometer's resistance is the documented way to minimize this effect.
What practical steps reduce potentiometer loading error in a real installation?
Documented mitigations include increasing the receiving circuit's load resistance relative to the potentiometer's own resistance, using a buffer circuit to isolate the potentiometer from the loading effect of downstream circuitry, or selecting a potentiometer with a higher overall resistance value to begin with — any of these reduces the proportion of current diverted through the measurement path relative to the potentiometer's own resistive element.
Why won't my potentiometer-based reading go all the way down to true zero, even at the wiper's zero-ohm end position?
This is documented as a real, physical limitation rather than necessarily a fault: potentiometers inherently have a nonzero minimum wiper contact resistance — wirewound types typically 0.5 to 5 ohms, conductive plastic types under 0.1 ohm — due to the physical contact interface itself, which can produce a small residual offset even at the theoretical zero-resistance end of travel.
Does temperature or humidity actually affect potentiometer accuracy over time, separate from mechanical wear?
Yes — this is specifically documented: temperature changes and humidity can alter the properties of the resistive element itself, causing drift in the potentiometer's performance distinct from mechanical wear at the wiper contact. Using higher-quality components and temperature-compensating materials is the documented mitigation, along with periodic recalibration to catch and correct for any accumulated drift.
Is linearity error the same thing as loading error, or a separate issue?
These are documented as distinct error sources: loading error results from the external circuit's resistance affecting the measurement (an interaction between the pot and what it's connected to), while linearity error results from imperfections in the potentiometer's own resistive element or mechanical design causing the output voltage to deviate from a perfectly linear relationship with wiper position — a potentiometer can suffer from either, both, or neither independent of the other.
Can wiper contact resistance itself change meaningfully over the life of a potentiometer, even without obvious mechanical damage?
Yes — documented technical analysis specifically ties wiper contact resistance to contact pressure at the interface between the wiper and the resistive element, noting that contact resistance generally decreases as contact pressure increases and vice versa. This means gradual mechanical changes (such as very slight wear altering that contact pressure) can shift contact resistance over time even without a dramatic, obviously visible failure.
Is there a "dead zone" in a rotary potentiometer's travel where position can't be reliably measured?
Yes, and this is a specifically documented mechanical characteristic: the physical rotational travel between a rotary potentiometer's two hard stops is often somewhat greater than its actual effective electrical travel, meaning there's a small region near one end of physical travel — sometimes called a deadspace — where the potentiometer has already reached its resistance extreme even though the shaft can still be turned slightly further.
Does regular calibration actually catch potentiometer drift before it becomes a significant measurement error?
Documented guidance specifically recommends regular calibration and testing as an ongoing practice to maintain accuracy and minimize the accumulation of error over time — since drift mechanisms like temperature/humidity effects on the resistive element and gradual contact resistance changes at the wiper are typically slow and progressive rather than sudden, periodic verification against a known reference is the documented way to catch and correct for this drift before it becomes operationally significant.






















Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.






