Understanding the Laureate™ 1/8 DIN Panel Meters, Ohmmeter for Resistance in Ohms
The Laureate™ 1/8 DIN Panel Meters Ohmmeter is ideal for high-speed, high-accuracy resistance measurements in a production environment, such as contact resistance measurements. It is factory calibrated for five jumper-selectable resistance ranges from 20.000 ohm to 200.00 kohm, plus three fixed factory-special ranges: 2.0000 ohm, 2.0000 Mohm, and 20.000 Mohm. Accuracy is ±0.01% of reading ± 2 counts. Resolution is one part in 20,000 — in the 2 ohm range, that works out to 0.1 milliohm resolution, making the meter well suited to contact resistance measurement.
Excitation Current by Range
Excitation current is fixed per range and scales inversely with resistance — from 5 mA on the 2Ω and 20Ω ranges down to 500 µA on the 200Ω range, 50 µA on the 2kΩ range, 5 µA on the 20kΩ range, 500 nA on both the 200kΩ and 2MΩ ranges, and 80 nA on the 20MΩ range. The applied excitation current is sensed by the meter, which operates ratiometrically and automatically compensates for any changes in that excitation.
Maximum Applied Voltage
Maximum applied voltage across the unknown resistance is 100 mV, with overvoltage protection rated to 125 Vac. If a sensor is open, the display flashes full-scale.
2, 3, or 4-Wire Connection
In 4-wire hookup, separate lead pairs apply excitation current and sense the voltage drop across the unknown resistance, eliminating IR drop across the excitation leads as a factor. In 3-wire hookup, the meter senses the combined voltage drop across the resistance plus two excitation leads, separately senses the drop across one excitation lead, and subtracts twice that value — effectively canceling lead resistance and compensating for ambient temperature changes if both excitation leads are identical. In 2-wire hookup, lead resistance is measured once by shorting the unknown resistance during setup and subtracted from all subsequent readings, but changes in lead resistance due to ambient temperature afterward won't be compensated.
QA Passband Mode for Contact Resistance Testing
A deviation limit (such as 50 mΩ) can be set on both sides of a setpoint. The relay closes (or opens) when the reading falls within that deviation band, and opens (or closes) when the reading falls outside it — setting up a passband around the setpoint specifically suited to contact resistance testing.
Custom Curve Linearization
The optional Extended main board allows up to 180 data points to linearize and scale nonlinear resistance sources — useful for resistance-based sensors (such as thermistors) whose resistance-to-parameter relationship isn't linear.
Factory-Calibrated Accuracy
All resistance ranges are factory-calibrated, with calibration factors stored in EEPROM that can be scaled via software to accommodate external shunts, enabling field replacement of the signal conditioner board without recalibrating the meter. Factory recalibration is recommended annually.
Where Ohmmeter Panel Meters Are Used
- Connector & Cable Assembly QA — contact resistance and low-level circuit resistance (LLCR) testing on production lines, using the 2Ω range's 0.1 mΩ resolution and the QA passband relay mode.
- Relay & Switch Manufacturing — go/no-go contact resistance screening against a setpoint band during 100% or sampled production testing.
- Bonding & Grounding Verification — earth bond and static-dissipative resistance checks in aerospace, electronics, and cleanroom manufacturing.
- Winding & Coil Testing — resistance measurement of motor windings, transformers, and solenoids for quality control and fault detection.
- Cable & Harness Testing — continuity and resistance verification across long production or field cable runs using 3- or 4-wire compensation.
- Insulation & High-Resistance Monitoring — megohm-range measurement of insulation resistance or high-value sensing elements using the fixed 2MΩ/20MΩ ranges.
- Thermistor & Resistance-Sensor Readout — linearized display of nonlinear resistance sensors via the Extended board's custom curve capability.
Ohmmeter Panel Meter Frequently Asked Questions
Why does maximum applied voltage matter for a resistance measurement, and why is it capped at 100 mV?
Keeping the voltage applied across the unknown resistance low is specifically important for measuring delicate or contaminated contact surfaces — a higher applied voltage risks breaking down thin oxide or contamination films on a contact surface, which would give an artificially low reading that doesn't represent the contact's true resting resistance. The 100 mV ceiling keeps the meter suitable for this kind of sensitive contact measurement.
Why does excitation current drop so dramatically between the lowest range (5 mA) and the highest range (80 nA)?
Since the meter is measuring a voltage developed across the unknown resistance at a fixed maximum voltage ceiling, a much higher resistance requires proportionally less current to develop a comparable, accurately measurable voltage — using the same higher current on a high-resistance range would either exceed the 100 mV limit or require the meter to handle a much larger voltage range than its front end is designed for.
What's the difference between the jumper-selectable ranges and the "special fixed" ranges like R0, R6, and R7?
The five standard ranges (R1-R5) can be jumper-selected and reconfigured by the user across a single ordered meter, while the three special ranges (R0 at 2Ω, R6 at 2MΩ, R7 at 20MΩ) are factory-set and fixed at time of order — reflecting that these three ranges likely require different internal circuitry optimizations (very high resolution at 2Ω, very high impedance handling at 20MΩ) that aren't practical to combine with simple jumper selection across the whole family.
How does the QA passband mode actually differ from a standard high/low alarm setpoint?
A standard high or low alarm triggers when a reading crosses a single threshold in one direction. Passband mode instead defines a band around a setpoint on both sides — the relay is active specifically while the reading stays within that band, and changes state when the reading moves outside it in either direction — which is specifically suited to contact resistance QA, where both "too high" (poor contact) and unexpectedly low or unstable readings can both be worth flagging.
Can this meter's custom curve linearization correct for a thermistor's nonlinear resistance-to-temperature relationship?
Yes — this is specifically documented as an application of the Extended board's 180-point linearization capability, which was originally illustrated for tank volume calculations but applies generally to any nonlinear resistance source, including thermistors, letting the displayed reading track the sensor's actual nonlinear behavior rather than a simple linear approximation.
Does "one part in 20,000" resolution mean the same absolute precision on every range?
No — it's a relative (ratiometric) resolution figure, so the actual absolute resolution scales with the range itself: 0.1 mΩ on the 2Ω range but 1000Ω on the 20MΩ range, since one part in 20,000 of a much larger full-scale value is itself a much larger absolute increment.
Is 3-wire connection a reasonable compromise for contact resistance testing, or should it always be 4-wire?
3-wire compensation works well specifically when both excitation leads have matched resistance, since the technique relies on doubling one measured lead's drop to cancel both — for very precise contact resistance work where lead matching can't be guaranteed, true 4-wire connection removes this dependency entirely and is the more robust choice.
Can the meter's analog output retransmit a resistance reading as a 4-20 mA signal for a PLC or SCADA system?
Yes, with the optional isolated analog output board — the resistance reading (already scaled if desired) can be output as 4-20 mA, 0-20 mA, 0-10V, or -10 to +10V, letting a separate control system monitor the resistance measurement independently of the meter's own relay-based alarming.
Does span temperature coefficient (±0.003% of reading/°C) matter much for a typical indoor QA testing environment?
In a temperature-stable indoor test environment, this drift is usually a very small contributor to overall measurement uncertainty, but for applications where ambient temperature swings meaningfully (near ovens, outdoor test stations, or unconditioned production floors), this spec becomes more relevant to account for when evaluating overall measurement uncertainty against a tight resistance tolerance.
If I need to measure both very low resistance (contact testing) and very high resistance (insulation testing) applications, do I need two separate meters?
Given the meter's fixed excitation current per range and factory-set special ranges (R0 for 2Ω, R6/R7 for megohm-range), a single meter ordered with a specific range configuration is optimized for one general resistance region — an application genuinely needing both very low and very high resistance measurement with full rated accuracy on each end would typically need two meters configured for their respective ranges, or a meter reconfigured between applications via jumper changes on the standard ranges.
Dry Circuit & Low-Level Contact Resistance Testing Questions From the Field
What is "dry circuit" testing, and why does it specifically matter for contact resistance measurement?
Documented industry standards define a dry circuit as one in which test voltage and current are deliberately limited to levels that can't cause changes in the physical or electrical state of the contact being tested — historically, this term contrasts with older "wetting voltage" practices that intentionally broke down surface films to get a lower reading. A true dry circuit measurement instead reveals the contact's actual resting metal-to-metal resistance, including any films or contamination present.
What are the standard voltage and current limits documented for dry circuit / LLCR testing?
Documented industry specifications (including ASTM standards and MIL-STD-1344 Method 3002.1) commonly stipulate a maximum open-circuit voltage of 20 mV (with 50 mV allowed under some specifications) and a maximum short-circuit current of 100 mA — these limits are specifically chosen to be too low to break down surface films on the contact, ensuring the measurement reflects true resting contact condition rather than an artificially improved reading.
Why is dry circuit testing typically performed using a 4-wire (Kelvin) measurement setup?
Documented testing practice specifically identifies Low Level Circuit Resistance (LLCR) measurements as generally requiring 4-wire Kelvin measurement circuits — since dry circuit test currents are deliberately very low, the resulting voltage drop across the contact under test is correspondingly tiny (often in the microvolt range), and a 4-wire connection is what allows that small voltage to be measured accurately without lead resistance error swamping the true reading.
Why is dry circuit testing typically performed before other electrical tests on a connector or contact?
Documented test sequencing guidance specifically recommends performing dry circuit measurements first, before other electrical tests, because other test methods can themselves cause physical or electrical changes to the contact surface — running a higher-voltage or higher-current test first could alter the very surface condition that the dry circuit test is meant to characterize.
Why is dry circuit testing typically paired specifically with environmental stress testing?
Documented practice notes that dry circuit testing is often performed in conjunction with environmental stress tests intended to deliberately introduce contamination or oxide films on contact surfaces — the dry circuit measurement isn't particularly meaningful until such stresses have actually been applied, since its purpose is specifically to reveal how much those contaminants have degraded the true contact resistance.
Does exceeding the recommended test current during contact resistance testing actually damage the contact being measured?
Yes — documented analysis specifically describes excessive current during testing as capable of causing microscopic heating that softens or melts the contact area, which enlarges the effective contact area and produces an artificially reduced (better-than-real) resistance reading. This is precisely the failure mode dry circuit current limits are documented as designed to prevent.
Does the correct test current for measuring contact resistance depend on the type of contact or application being tested?
Yes — documented guidance specifically distinguishes testing approaches by application: high-power switches and relays are documented as appropriately tested with higher currents that mimic their actual working conditions, while dry circuit connectors specifically require low currents (under 100 mA) to avoid altering delicate contact surfaces — using the wrong test regime for a given contact type can produce a reading that doesn't represent real-world performance.
Is a single universal dry circuit test specification used across all industries, or do requirements vary?
Documented sources show meaningful variation — while ASTM standards and MIL-STD-1344 are commonly cited references, documented guidance specifically notes that MIL-STD-1344 Method 3002.1 is intended for connector manufacturer qualification testing rather than completed cable/harness assembly production testing, and that assemblers wanting to comply typically rely on the connector manufacturer's own confirmation of compliance rather than independently re-running the qualification test — confirming which specific standard and test parameters actually apply to a given testing context is worth verifying rather than assuming one universal specification covers all cases.

























Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.


