Understanding the Laureate™ 1/8 DIN Panel Meters for True RMS AC Voltage and Current
The Laureate™ 1/8 DIN Panel Meters for True RMS are 60,000-count panel meters with exceptional performance at half the price of a benchtop meter. They can measure true RMS voltage or current of AC signals with considerable waveform distortion, such as rectifier outputs or waveforms chopped by an SCR, TRIAC, or transistor circuit.
Fast Response and Crest Factor
True RMS readings arrive 0-16.7 ms after completion of one input signal cycle, allowing anomalies to be detected and alarmed before they become expensive problems, with fast on/off control achievable via optional dual or quad relays. Accuracy is 0.03% of full scale with the meter's standard 1 megohm input resistance, signals from DC to 5 kHz, and signal amplitudes down to 0.5% of full scale. Crest factor (Vp/Vrms) is 3.0 at full-scale Vrms of 20,000 counts and a Vp of 60,000 counts. The meter begins flashing overrange at 132% of full scale Vrms (at which point available crest factor is 2.27), and meaningful readings with rated resolution continue up to 212% of full scale Vrms for sinusoidal signals (crest factor 1.414). For safety, maximum RMS input should never exceed 600V or 5A.
AC or DC Coupling
AC or DC coupling is jumper selectable — AC coupling suits applications like measuring ripple on a DC power supply. Multiple integral cycles are averaged for signals above 50/60 Hz; a single cycle is captured for signals from 3 Hz to 50/60 Hz; below 3 Hz and at DC, capture rate is every 333 ms.
Use With Current Transformers
High common-mode rejection allows stable readings with current shunts located on the high side of the line. A built-in 5A input capability lets 5A current transformer outputs be applied directly, with no stepdown transformer needed, and readings easily scaled for the transformer ratio via front-panel pushbuttons.
Real-World Applications
- Synchronizing Two Motor Generators — two Laureate AC RMS voltmeters (or one multiplexed via external toggle switch) measure the two voltages; a Laureate dual-channel counter measures both frequencies to six-figure accuracy; a Laureate phase meter measures phase angle — together confirming voltages, frequencies, and waveshapes are matched before paralleling.
- Instrumenting an AC Line — AC loads like motors only operate reliably within specified voltage and frequency tolerances; Laureate meters and counters provide low-cost, accurate instrumentation and alarming, including true RMS voltage for non-sinusoidal waveshapes like UPS square waves.
- Measuring Voltages Over 600V — an external step-down voltage transformer brings the signal under 600V for the meter's 600.0V range, also providing electrical isolation; for example, a 2:1 stepdown for 1000V AC measurement, scaled with a factor of 0.2.
Factory-Calibrated Accuracy
All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM that can be scaled via software to accommodate external shunts, enabling field replacement of the signal conditioner board without recalibrating the meter. Factory recalibration is recommended annually.
Where True RMS AC Panel Meters Are Used
- Motor & VFD Monitoring — accurate voltage/current readings despite PWM switching harmonics that would mislead an average-responding meter.
- UPS & Backup Power Systems — true RMS measurement of square-wave or stepped-waveform inverter outputs during normal and backup operation.
- Power Generation & Paralleling — synchronizing generators before connecting them to a shared bus or the grid.
- SCR/TRIAC-Controlled Heating & Lighting — accurate RMS readout of phase-chopped, non-sinusoidal control waveforms.
- Substations & Switchgear — line voltage and current monitoring with current transformer inputs, without a separate stepdown transformer.
- Rectifier & DC Power Supply Ripple Testing — AC-coupled measurement of ripple riding on a DC bus.
- Test & Calibration Benches — a low-cost alternative to benchtop true RMS meters for panel-mounted, continuous monitoring rather than spot-checks.
True RMS AC Panel Meter Frequently Asked Questions
What actually happens to the meter's reading once a signal exceeds 132% of full scale?
At 132% of full scale Vrms, the meter begins flashing to indicate overrange — at that point the available crest factor has dropped to 2.27 (from the rated 3.0 at full scale), meaning the meter's headroom for peak excursions above the RMS value is shrinking as the signal approaches its true limits.
Can the meter still give a meaningful reading above the 132% overrange flash point?
Yes, to a further point — for sinusoidal signals specifically, meaningful readings with rated resolution continue up to 212% of full scale Vrms, where available crest factor is 1.414 (the theoretical minimum for a pure sine wave). Beyond that point, there's no more crest factor margin left to accommodate any peak above the RMS value.
Why does AC coupling matter specifically for measuring ripple on a DC supply?
AC coupling removes the DC component of a signal before RMS calculation, isolating just the AC ripple riding on top of the DC bus — without AC coupling, the large DC offset would dominate the reading and the small ripple component of interest would be effectively invisible.
Does the meter's capture rate stay constant regardless of the input signal's frequency?
No — capture rate is tied to the signal itself: multiple integral cycles are averaged above 50/60 Hz, a single cycle is captured for signals between 3 Hz and 50/60 Hz, and below 3 Hz (down to DC) the capture rate becomes a fixed 333 ms interval rather than being tied to signal cycles at all.
Why can a 5A current transformer be connected directly to this meter without a stepdown transformer?
The meter's built-in 5A input range is specifically designed to accept the standard secondary output of common current transformers directly — since 5A is a widely used CT secondary rating, this avoids the extra cost, complexity, and potential error of an intermediate stepdown stage that would otherwise be needed to bring the signal into range.
What's actually different between using a current shunt versus a current transformer with this meter?
A shunt produces a voltage output proportional to current and is read on one of the meter's voltage ranges (such as 200.00 mV), while a current transformer produces a scaled-down current output read directly on the meter's current ranges — the meter's high common-mode rejection specifically supports shunts placed on the high side of the line, where common-mode voltage relative to ground can otherwise be a significant source of error.
Does measuring a distorted, non-sinusoidal waveform reduce the maximum voltage the meter can safely handle?
The 600V/5A safety maximum is an absolute limit regardless of waveform shape, but a distorted waveform's higher crest factor can cause the meter to reach its overrange flash point (132% of the RMS full-scale setting) at a lower RMS reading than a clean sine wave would, since more of the available "headroom" is being consumed by higher peak excursions relative to the RMS value.
Is phase angle measurement between two AC signals handled by this True RMS meter, or does that require a separate instrument?
A separate instrument — the documented motor generator synchronization application specifically pairs this True RMS voltmeter with a separate Laureate phase angle meter and a separate dual-channel frequency counter, each measuring a different parameter (voltage, phase, frequency) needed to confirm safe generator paralleling.
Does the meter's 1 megohm input resistance ever become a limiting factor for accurate voltage measurement?
For most voltage-sensing applications, 1 megohm is high enough to avoid meaningfully loading the source being measured, but a documented factory-special version with 10 megohm input resistance is available for applications needing even less loading — that higher-impedance version trades off maximum frequency, dropping from 5 kHz to 1 kHz on three of the voltage ranges.
If I need to measure a signal with content above 5 kHz, is this meter still a good fit?
Not for accurate measurement above that point — the meter's rated accuracy band of 0.03% of full scale specifically applies to signals from DC to 5 kHz, so applications needing accurate RMS measurement of significant harmonic content above 5 kHz would be pushing past this meter's documented bandwidth and should consider an instrument rated for that higher frequency range instead.
Crest Factor & True RMS Measurement Questions From the Field
Why do two meters both labeled "true RMS" sometimes give noticeably different readings on the same distorted waveform?
This is documented as a specification, not a defect: every true RMS instrument has its own rated maximum crest factor within which its accuracy specification actually holds, and documented examples show conventional RMS converters can carry roughly 4% error at a crest factor of 4 — a waveform with a crest factor exceeding a given meter's rated limit (such as certain PWM signals, which can exceed crest factor 30) will produce results that diverge from a meter better suited to that specific waveform.
Does a true RMS meter's specified sine-wave accuracy automatically apply to any waveform it measures?
No — this is documented as a common and consequential misconception. A meter's headline accuracy specification is typically established using a sine wave, and the actual shape of a real-world input signal can dramatically affect measurement accuracy in practice, independent of whether the meter is technically "true RMS" capable.
Why do crest-factor-related measurement errors on distorted waveforms almost always cause a reading that's too low rather than too high?
This is a specifically documented, predictable pattern: crest factor errors are described as consistently reducing the reading rather than inflating it, because instruments have limited bandwidth and can't fully capture the higher-frequency harmonic energy present in high-crest-factor signals — measuring only the lower harmonics while missing the higher-frequency components produces a systematically low RMS result.
Is crest factor error unique to voltage/current meters, or does it affect power (wattmeter) measurements too?
Documented analysis specifically extends this limitation to wattmeters as well — the ratio of RMS to average current varies significantly and unpredictably from waveform to waveform depending on duty cycle, and every wattmeter design carries its own specific maximum allowable crest factor; exceeding that rated value is documented as typically producing a reading that's low, sometimes severely so.
What two distinct measurement challenges does a high crest factor create for an instrument, beyond just "accuracy"?
Documented technical guidance identifies two separate challenges: first, an input-range problem, where a signal with a high peak but low RMS value forces an instrument to simultaneously handle a large peak (risking overload) alongside a much smaller RMS value (risking poor resolution); second, a bandwidth problem, since high-crest-factor signals inherently carry more energy in higher-frequency harmonic content that a limited-bandwidth instrument may not fully capture.
Are engineered solutions to extend a meter's crest factor limitation without sacrificing low-signal accuracy actually straightforward?
No — this is documented as a genuine engineering tradeoff, not a simple fix: methods that shrink an incoming signal to fit a device's dynamic range typically sacrifice accuracy at lower voltages (which contribute the most to an accurate RMS calculation), while analog gain-switching approaches introduce their own settling-time and accuracy problems at the switch point, and using two separate RMS converters (one for high signals, one for low) to avoid these issues adds real cost and complexity.
Does crest factor limitation get worse specifically for the kinds of PWM signals common in modern motor drives and switching power supplies?
Yes, and this is specifically and quantifiably documented — PWM signals used in motor controllers and switching-mode power supplies can carry crest factors greater than 30, far beyond what conventional commercially available RMS meters are typically rated to handle accurately, making this a genuinely common real-world limitation rather than an edge case confined to unusual signals.
Can a meter's crest factor rating be trusted at any point across its measurement range, or only near full scale?
Documented guidance specifically notes this isn't uniform across the range — a true RMS instrument's ability to handle higher crest factors is generally best when the waveform's RMS voltage sits in the middle of the meter's selected range setting, meaning the same meter measuring a high-crest-factor signal near the bottom or top of a given range may not achieve the same crest-factor headroom quoted for mid-range conditions.
The crest factor (Vp / Vrms) is 3.0 at full-scale Vrms of 20,000 counts and a Vp of 60,000 counts. The panel meters start to flash overrange at 132% of full scale Vrms, at which point the available crest factor is 60,000 / 26,400 = 2.27. Meaningful readings with rated resolution continue to be obtained up to 212% of full scale Vrms (42,433 counts) for sinusoidal signals, at which point the available crest factor is 1.414. For safety reasons, the maximum RMS input signal should never exceed 600V or 5A. ETL certification is for a maximum voltage of 300 Vrms.






















Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.