Understanding the Laureate™ 1/8 DIN Panel Meter for Thermocouple Temperature
The Laureate™ 1/8 DIN Panel Meter for thermocouples is factory calibrated for Type J, K, T, E, N, R, and S thermocouples, with the entire span of each thermocouple type presented in a single range for exceptional accuracy and repeatability. Thermocouple type, unit of measure (°C or °F), display resolution (1° or 0.1°), and open-sensor indication direction are all selectable from the front panel or via the serial interface. Cold junction compensation is standard on every unit.
Factory-Calibrated Accuracy
All ranges for all thermocouple types are factory-calibrated, with calibration factors stored in EEPROM on the signal conditioner board. Because those calibration factors live on the board itself rather than in the main meter electronics, a thermocouple sensor and its signal conditioner board can be swapped in the field without needing to recalibrate the entire meter. Overall accuracy at 25°C is ±0.01% of full span, plus a conformity error that varies by thermocouple type — as low as ±0.03°C for Type T near 0°C to ±400°C, up to ±0.2°C for Type T below 0°C. Factory recalibration is recommended every 12 months, and all Laurel instruments are calibrated at the factory using Fluke calibrators that are themselves recalibrated yearly and certified traceable to national standards.
Cold Junction Compensation
A cold junction compensation (CJC) transistor, positioned adjacent to the two thermocouple input terminals, senses ambient temperature at the connection point and is calibrated as a system together with the signal conditioner board, with all calibration factors stored in EEPROM. This is what allows the meter to correctly interpret a thermocouple's millivolt signal regardless of the actual ambient temperature at the panel where the meter is mounted.
High-Speed Signal Conversion and Filtering
Using Concurrent Slope analog-to-digital conversion (U.S. Patent 5,262,780), the meter reads at up to 60 conversions per second at 60 Hz power, integrating the signal over a full power line cycle. This high read rate supports peak and valley capture, real-time computer interfacing, and closed-loop control applications. Three digital filtering modes are available: an unfiltered mode for true peak/valley readings and control applications; a batch average mode that averages every 16 conversions; and an adaptive moving average mode offering 8 selectable time constants from 80 ms to 9.6 seconds, which briefly switches to its shortest time constant when a significant signal change occurs, then reverts to the selected setting.
Tare, Peak/Valley, and Control Inputs
The meter supports both auto-tare (triggered by grounding an input line via an external pushbutton) and manual tare (entered via a control input pushbutton or the free Instrument Setup Software). Peak and valley values are automatically captured and can be viewed via a front-panel pushbutton, a rear-connector control signal, or transmitted as serial data. Two rear-panel control inputs (CMOS/TTL levels or dry contacts) can be configured to trigger any of 14 meter commands.
Modular Options
The meter accepts up to five isolated plug-in boards: power supply, signal conditioner, and optional relay output (2 or 4 relays, mechanical or solid-state), analog output (4-20 mA, 0-20 mA, 0-10V, or -10V to +10V), and a communications board (RS232, RS485, USB, Ethernet, or WiFi, supporting Laurel Custom ASCII or Modbus RTU/TCP). All boards are isolated from meter and power grounds. The enclosure is rated NEMA-4X (IP-65) from the front when panel mounted, in a 1/8 DIN cutout of 92mm x 45mm.
Where Is This Panel Meter Used?
This thermocouple Panel Meter is used across industries wherever accurate, real-time temperature monitoring or control is needed:
- Industrial Automation and Process Control — monitoring reactor, oven, or process temperatures with alarm or shutdown relay outputs.
- Manufacturing — metal treatment, plastics extrusion, and food production, where staying within a specified temperature band prevents defects and equipment damage.
- Laboratories — incubators, water baths, and environmental chambers requiring precise, logged temperature data.
- HVAC Systems — tracking refrigerant line, air handler, and heating element temperatures.
- Energy and Utilities — steam generation and cooling processes in power plants and utility infrastructure.
- Chemical Processing — reaction vessel and distillation column temperature monitoring where preventing runaway reactions is safety-critical.
- Metallurgy and Heat Treatment — furnace, annealing, and quenching temperature control for material property consistency.
Operation as a 4-20 mA Transmitter, Controller, or Supervisory Monitor
With the optional analog output board, this meter can serve as an isolated 4-20 mA transmitter, with the output scaled to the linearized display and tracking the meter's full read rate — useful in closed-loop and PID control applications. With the relay output boards, it can operate as a fast ON/OFF controller with individually selectable setpoint modes per relay, or as a supervisory monitor using band-deviation alarm modes with latching or non-latching relay behavior.
Thermocouple Panel Meter Frequently Asked Questions
Why does the accuracy specification differ between thermocouple types on the same meter?
Each thermocouple type has a different conformity error inherent to how linear its voltage-to-temperature curve is across its rated span — for example, Type T is rated to ±0.03°C over 0°C to +400°C but ±0.2°C from -257°C to 0°C, while Type E is rated ±0.18°C across its full range. The meter's own ±0.01% of full span accuracy is added to this conformity error, which is why the overall spec varies by thermocouple type rather than being a single number.
Can I swap the thermocouple sensor and signal conditioner board without recalibrating the whole meter?
Yes. Calibration factors are stored in EEPROM on the signal conditioner board itself, together with the CJC calibration, so a board can be swapped in the field and the meter will continue reading accurately without a full recalibration. This is specifically what the modular board architecture is designed to support.
How often does this meter need to be recalibrated?
Factory recalibration is recommended every 12 months, even though the meter is shipped factory-calibrated and traceable to national standards through Fluke calibrators recalibrated yearly.
What is the maximum lead resistance the meter can tolerate for rated accuracy?
Rated accuracy holds up to 1 kΩ of lead resistance. Runs with resistance beyond that specification can start to introduce measurement error, which matters most on very long thermocouple extension wire runs.
What's the difference between the three digital filtering modes?
Unfiltered gives true peak and valley readings, useful for control applications where response speed matters most. Batch average smooths the reading by averaging every 16 conversions. Adaptive moving average offers 8 selectable time constants (80 ms to 9.6 seconds) and automatically switches to its fastest response briefly when it detects a significant signal change, then reverts to the selected time constant — balancing stability with responsiveness.
What's the difference between auto-tare and manual tare on this meter?
Auto-tare is triggered by grounding a designated input line via an external pushbutton. Manual tare is entered directly via a control input pushbutton or through Laurel's free Instrument Setup Software, letting the operator set a specific tare value rather than capturing it automatically from a live signal.
How does the meter indicate an open (broken) thermocouple sensor?
The display flashes at full-scale when an open sensor is detected, and the direction of that indication (upscale or downscale) is selectable from the front panel, letting the failure mode match whatever behavior downstream alarm or control logic expects.
What is the meter's input resistance and input current, and why does that matter for thermocouple measurement?
The meter has an input resistance of 1 GΩ and input current of just 100 pA, which is important because thermocouples are low-level millivolt signal sources — very high input impedance and very low input current mean the meter draws negligible current from the thermocouple circuit, avoiding self-loading errors that a lower-impedance instrument could introduce.
What overvoltage protection does the input have?
The signal input is protected up to 125 Vac, which helps prevent damage to the sensitive input circuitry from an accidental miswiring event or a transient on the sensor line.
Can this meter be networked with other Laureate meters on the same communications line?
Yes. Up to 30 Laureate Panel Meters and/or LT Transmitters can be configured for RS485 and daisy-chained using Laurel's Ethernet-to-RS485 converter board for LAN integration, with Modbus RTU (serial) or Modbus TCP (Ethernet/WiFi) available as communication protocols.
Thermocouple Panel Meter Questions From the Field
Why is my thermocouple reading fluctuating or showing values that don't correspond to any real temperature?
This is frequently traced to a cold junction compensation problem rather than a bad sensor — if the reference junction temperature isn't being measured or applied correctly, the displayed reading can drift or jump in ways that don't track the actual process. Confirming that CJC is functioning correctly, and checking that the input terminal area isn't exposed to an unusual draft or heat source that would throw off the CJC transistor's own reading, is a standard first troubleshooting step.
I configured the wrong thermocouple type by mistake — how far off will my readings be?
The error is usually significant and systematic rather than a minor rounding difference — using one thermocouple type's reference coefficients to interpret a different type's signal has been documented to introduce a consistent multi-degree error across most of the operating range. Reconfiguring the meter to the correct thermocouple type (rather than trying to apply a manual correction factor) is the correct fix.
How do I find a break in a long thermocouple extension wire run?
A commonly used field technique is the progressive short-circuit method: with the thermocouple still connected to the meter, temporarily short the two wires together at successive points along the run. If the meter reads local ambient temperature at a given point, the wiring is intact up to there — the break is between that point and the next one where the reading no longer changes.
Does it matter whether a thermocouple probe is grounded or ungrounded?
Yes. A grounded probe, where the measuring junction is electrically connected to the metal sheath, responds faster to temperature changes but is more prone to picking up electrical noise and ground loops. An ungrounded probe isolates the junction electrically, trading a small amount of response speed for better noise immunity — worth considering in electrically noisy installations near VFDs or switching equipment.
Can extension wire other than the matched thermocouple type be used to run the signal to the meter?
No — it needs to be the correct matched extension wire for the thermocouple type in use (for example, chromel/alumel extension wire for Type K), not generic copper wire, since standard wire introduces its own thermoelectric voltage at the connection points and will distort the reading.
Why does my thermocouple reading change depending on which panel or building the meter is installed in?
This is generally traced to whether the meter's cold junction reference is being read dynamically from an actual sensor at the input terminals versus assumed as a fixed value — a fixed reference works fine on a bench at one ambient temperature but produces a location-dependent error once installed in a different panel at a different ambient temperature. This meter's CJC transistor, positioned directly at the input terminals and calibrated as a system with the signal conditioner board, is specifically designed to avoid this class of problem by tracking actual local ambient temperature.
Why does my reading become noisy or unstable, specifically in a particular part of the facility, but not elsewhere?
Localized noise is commonly traced to nearby switching equipment, VFDs, or heavy power wiring routed close to the thermocouple signal cable in that specific area. Since thermocouple signals are only tens of microvolts per degree, they're especially susceptible to induced noise; rerouting the signal cable away from power wiring, or switching to a longer filter time constant on the meter, are the standard remedies for a location-specific noise issue.
My readings look plausible but seem to be consistently offset by a fixed amount from a reference thermometer — what should I check first?
A consistent fixed offset (rather than random noise or drift) points toward either a thermocouple type mismatch, a CJC calibration issue, or simply the two instruments being calibrated against slightly different reference standards. Verifying the thermocouple type setting first, then comparing both instruments against a common calibration reference rather than against each other, isolates which is actually correct.





















Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.




