Understanding the Laureate™ 1/8 DIN Panel Meter for A-to-B Time Interval of Periodic Events
The Laureate™ 1/8 DIN Panel Meter for A-to-B time interval can display pulse width or time delay between individual pulses to a resolution of 0.2 µs, or the average pulse width or average time delay across multiple periodic pulses. Unlike the single/cumulative-event stopwatch configuration, this meter is specifically built to measure and average the interval of repeating, periodic events.
How Time Interval Is Measured
Time interval is measured between inputs on Channels A and B. Timing starts when a pulse is applied to Channel A (selectable positive or negative edge) and ends when a pulse is applied to Channel B (selectable positive or negative edge). For a single pulsed signal, the A and B inputs can be tied together, with a positive or negative slope selected to start timing and the opposite slope selected to stop timing — measuring that pulse's width. Timing itself is achieved by counting 5.5 MHz clock pulses. Multiple integral time intervals are averaged over a gate time selectable from 10 ms to 199.99 s, which also controls the display update time.
Display and Resolution
Time interval can be displayed in seconds, milliseconds, or microseconds with 6-digit resolution — typically milliseconds with 1 µs resolution. Resolution varies by the actual range of the measured interval: 1 ms over 0-199.999 s, 100 µs over 0-99.9999 s, 10 µs over 0-9.99999 s, 1 µs over 0-0.999999 s, and 0.2 µs over 0-0.099999 s. For times under 100 ms, display resolution down to 0.2 µs can be achieved by applying a multiplier of 10, moving the decimal point one position, and averaging many time intervals.
Rate Based on 1/Time
Highly accurate rate can be displayed by taking the inverse of the measured time interval, with extensive arithmetic capability allowing display in engineering units such as meters/sec. This requires the Extended main board version, which adds this rate-based-on-1/time capability along with rate and total simultaneously, dynamic up-down counting, arithmetic functions between channels (A+B, A-B, A/B, AxB, A/B-1), phase angle, power factor, duty cycle, batch control, and custom curve linearization.
Maximum Signal and Overcurrent Protection
Maximum applied voltage is 600 Vac for the 20, 200, and 300V ranges, and 125 Vac for other ranges. Overcurrent protection ratios are 25x for the 2 mA range, 8x for 20 mA, 2.5x for 200 mA, and 1x for 5A.
Real-World Applications
- Time Delay Measurement — for periodic pulses applied to A and B channels, time delays can be measured down to 0.2 µs resolution from the rising or falling edge of A to the rising or falling edge of B (selectable).
- Pulse Width Measurement — the width of periodic pulses can be measured by tying A and B channels together, with readings averaged over a user-selectable gate time, same as time delay.
- Timing Process Dynamics — start and stop pulses can be generated by the dual relay board in a separate Laureate panel meter or digital counter, for example as temperature passes two alarm setpoints or cycles in a hysteresis control mode.
- Rate Based on 1/Time — a pulse or switch closure initiates timing while another stops it, with the Extended meter programmed with multipliers to display rate in engineering units like meters/sec for any time duration.
- Instrumenting a Pulsed Laser System — Laureate dual-channel counters can be used for elapsed time, pulse count, pulse width, pulse separation, duty cycle, and pulse repetition rate.
Factory-Calibrated Accuracy
The internal time base is crystal-calibrated to ±2 ppm, with span tempco of ±1 ppm/°C and long-term drift of ±5 ppm/year. All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM that can be scaled via software to accommodate external shunts, enabling field replacement of the signal conditioner board without recalibrating the meter. Factory recalibration is recommended annually.
Time Interval Panel Meter Frequently Asked Questions
How is this meter different from the single/cumulative-event stopwatch configuration?
This meter is specifically built to measure and average the time interval of repeating, periodic events, whereas the stopwatch configuration times single events (resetting on the next start pulse) or accumulates total elapsed time across multiple discrete events. Here, multiple integral time intervals from a periodic signal are automatically averaged over the gate time to produce a single, smoothed reading.
Why does display resolution change depending on the length of the measured time interval?
The meter's underlying clock counts at a fixed 5.5 MHz rate, but the display resolution shown is scaled to the actual magnitude of the interval being measured — 1 ms resolution for intervals up to 199.999 s, down to 0.2 µs resolution for intervals under 0.099999 s. This keeps the six-digit display meaningful across a very wide range of possible interval lengths rather than wasting display digits on ranges that don't need them.
How does averaging over the gate time actually improve accuracy for a periodic signal?
Rather than reporting a single measured interval (which is subject to quantization error from the counting clock, plus any jitter in the input signal itself), the meter measures multiple integral time intervals within the gate time and averages them — reducing the influence of any single noisy or jittery measurement and producing a more stable, repeatable reading, at the cost of a slower update rate for longer gate times.
Can I get finer than the standard resolution for very short time intervals?
Yes — for intervals under 100 ms, resolution down to 0.2 µs is achievable by applying a x10 multiplier, moving the decimal point one position, and averaging many time intervals, which is a documented technique for extracting additional resolution beyond the meter's baseline display scaling.
What's the difference in wiring between measuring time delay and measuring pulse width?
Time delay measurement uses A and B as genuinely separate channels — timing runs from an edge on A to an edge on B, which can come from two different physical events or sensors. Pulse width measurement instead ties A and B together electrically, so the "delay" being measured is actually between the rising and falling edge of the same single pulse.
Does this meter need the Extended counter for basic time interval averaging, or only for rate calculation?
Basic time interval and pulse width averaging work on the Standard counter. The "Rate Based on 1/Time" mode, which mathematically inverts the averaged time interval into a rate or speed reading in engineering units, specifically requires the Extended counter's additional arithmetic capability.
What determines the maximum voltage this meter's input can safely accept?
It depends on which input range is selected — the 20V, 200V, and 300V ranges are rated for up to 600 Vac maximum applied voltage, while other ranges are limited to 125 Vac. Confirming the actual expected signal voltage against the specific range selected for a given application avoids exceeding the rated input limit.
What does the overcurrent protection ratio (such as "25x for 2 mA") actually mean?
It expresses how much overcurrent, relative to the range's normal rated current, the input protection circuitry can withstand before failing — for example, the 2 mA range's protection can tolerate a fault current up to 25 times the 2 mA rating, while the 5A range's protection is rated for only 1x (its own rated current), reflecting the very different fault-tolerance margins built into each range.
Can start and stop pulses for this meter come from another instrument's relay outputs, rather than dedicated timing sensors?
Yes — a documented application uses the dual relay board in a separate Laureate panel meter or digital counter to generate the actual start and stop pulse edges, for example triggered as a monitored temperature crosses two alarm setpoints, letting this meter time (and average, if the event repeats periodically) the interval between those two process events.
Is this meter suitable for laser pulse characteristics like pulse separation and repetition rate, or only single time-delay measurements?
It's documented as suitable for a broader set of pulsed-system measurements — elapsed time, pulse count, pulse width, pulse separation, duty cycle, and pulse repetition rate are all listed as achievable using Laureate dual-channel counters in a pulsed laser instrumentation context, not just a single time-delay reading.
Periodic Time Interval Measurement Questions From Test & Measurement Engineering Sources
Why does averaging multiple periods of a repeating signal improve resolution beyond what a single measurement could achieve?
This is a well-documented technique in precision timing instrumentation: if a single measurement has a baseline resolution determined by the counting clock, averaging many periods together can improve that resolution by a factor related to the number of periods averaged — a documented example shows resolution improving by a factor of 100 (from 5 ns down to 50 ps) when 100 periods are averaged together, since the underlying quantization error is distributed across many more clock counts.
Does time interval averaging eliminate the meter's own time base error, or only reduce quantization noise?
This is a specifically documented distinction: time interval averaging does not change the time base source of measurement error, nor does it change systematic error — averaging specifically reduces the quantization (counting) error component, while the time base's own inherent accuracy is a separate factor that averaging cannot improve. A better-quality time base oscillator is what's needed to reduce that separate error source.
What is "reciprocal counting," and why is it specifically well suited to measuring low-frequency or long-period periodic signals?
Reciprocal counting is a documented technique where the gate is synchronized to the input signal itself (rather than to a fixed clock interval), so the number of signal cycles is counted exactly while the elapsed time carries the (much smaller) uncertainty — this is specifically documented as giving good resolution for low-frequency signals, where a conventional gate-synchronized-to-clock approach would only capture very few signal cycles and produce poor resolution.
Is there a tradeoff between choosing a longer gate time for better resolution and getting timely updated readings?
Yes, and this is a universally documented tradeoff in gate-time-based timing instruments: longer gate times average over more signal cycles and generally produce better resolution and stability, but at the direct cost of slower update rates, while shorter gate times give faster readings with correspondingly less averaging and detail. Choosing gate time is described as a fundamental speed-versus-accuracy tradeoff inherent to this measurement approach, not something that can be optimized away.
Why might a period/time-interval measurement actually have worse quantization error than an equivalent frequency measurement, in certain conditions?
This is a specifically documented, somewhat counterintuitive finding: for input frequencies higher than the instrument's own counted clock frequency, the ±1 count quantization error for a period-based measurement is actually larger than the corresponding error for a frequency measurement — which is why well-designed reciprocal counters are documented to automatically switch measurement mode to frequency-based rather than period-based once input frequency exceeds the internal clock frequency.
Does jitter or noise on the periodic input signal itself limit achievable measurement accuracy, separate from the instrument's own resolution?
Yes — this is specifically documented as a major limiting factor distinct from instrument resolution: measuring a single period of a signal is described as being highly affected by signal jitter and internal instrument noise, with results potentially varying by thousands of parts per million on a single-period measurement. Averaging over many periods (which this meter's gate-time averaging specifically does) is the documented way to substantially reduce this jitter-driven error, though it still doesn't necessarily match the accuracy achievable by dedicated reciprocal-counting hardware.
Can averaging over an excessively long gate time actually hide a real, meaningful change in the periodic signal's timing?
This is an inherent tradeoff of any averaging-based measurement approach — while not always framed as a "problem" in documented sources, the same averaging that reduces noise and jitter also necessarily smooths over genuine short-term variation in the underlying signal. If an application specifically needs to detect brief timing anomalies in an otherwise periodic signal, a shorter gate time (accepting somewhat less averaging benefit) preserves more sensitivity to those individual variations than a longer, heavily-averaged gate time would.























Slide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel.
The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter. 






