Panel Meters for Serial Input 6-Digit and Remote Display Applications

Panel Meters for Serial Input 6-Digit and Remote Display Applications

Price: $325.00
  • P/NL60001
- +

Features

  • Accepts RS232, RS485 or USB streaming data
  • Accuracy of 0.01% of reading ± 2 counts
  • Able to extract readings embedded in ASCII data strings
  • Displays 6-digit numeric readings with any decimal point
  • Up to 60 conversions per second, Ideal for peak or valley capture
  • Digital span adjust from 0 to ±99,999, zero adjust from -99,999 to +99,999
  • Front panel scalable to ±99,999 for use with current shunts
  • 1/8 DIN size with bright red or green 0.56" (14.2mm), high LED digits
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)
  • Wide choice of Plug-in-Play options:
    - 2 or 4 relays, mechanical or solid state, for alarm or control (isolated)
    - 1 or 2 Analog output, 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V (isolated)
    - Communications: Ethernet, WiFi, USB, RS232, RS485 (isolated)


Laureate™ 1/8 DIN Panel Meters Remote Displays

(or Serial Input Panel Meters) are slave displays which accept RS232, RS485 or USB data from computers, programmable controllers, Laureate instruments, or other devices with a streaming serial data output. They can also provide relay closures and an analog output based on the received readings. They blend in with 1/8 DIN Laureate panel meters, counters and timers to provide a numeric 6-digit display from -999,999 to +999,999.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated panel meters. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Streaming serial data can be generated at rates up to 9600 baud by a weighing scale or other instrument. In particular, it can be generated by Laureate panel meters, counter/timer or transmitter. A Remote Display can duplicate the reading of that instrument, or it can be used to display additional data items which can be transmitted serially, but which cannot be shown simultaneously on a single display. For example, while a Laureate counter can only display one selected parameter at a time (such as rate A), it can transmit multiple parameters (such as rate A, rate B, ratio A/B and peak). The Remote Display can be set up to display any serially transmitted item, and an indicator light shows which item has been chosen for display.

Setup of the Remote Display uses Laurel’s free Instrument Setup Software. Selectable modes allow easy interface to other Laureate instruments and extract 1, 2 or 3 data items. For other instruments, readings can be extracted from streaming ASCII strings that contain multiple data values and non-numeric characters, such as Start and Stop characters. Any number of characters between the Start character and the data can be masked Off. Up to 8 display characters (including sign and decimal point) can be masked On. Any number of characters between the last displayed character and the Stop character can be masked Off.

An optional (isolated) analog output board allows the Remote Display to serve as a highly accurate digital-to-analog converter and transmitter.

An optional relay output board with 2 or 4 relays can add remote alarm or control capability. The relays can be 8A contact relays or opto-(isolated) 120 mA AC/DC solid state relays. The relays can respond to the transmitted values or to any of 8 serially transmitted control characters. The control characters can be generated by Laureate panel meters, counter or timer, thereby assuring that the local and remote alarm points are identical.

Display
Readout 6 LED digits, 7-segment, 14.2 mm (.56"), red or green.
Range -999,999 to +999,999
Indicators Four LED lamps
* Range not ETL certified.
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Power
Voltage, standard 85-264 Vac or 90-300 Vdc
Voltage, optional 12-32 Vac or 10-48 Vdc
Frequency DC or 47-63 Hz
Power  consumption (typical, base meter) 1.2W @ 120 Vac, 1.5W @ 240 Vac, 1.3W @ 10 Vdc, 1.4W @ 20 Vdc, 1.55W @ 30 Vdc, 1.8W @ 40 Vdc, 2.15W @ 48 Vdc
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Interface (one required)
Board selections RS232, USB, RS485 (dual RJ11), RS485 (dual RJ45),
ASCII data rates 300 to 9600 baud
Isolation 250V rms working, 2.3 kV rms per 1 min test
Analog Output (optional)
Output levels 4-20 mA, 0-20 mA, 0-10V, -10 to +10V
Current compliance 2 mA at 10V ( > 5 kΩ load)
Voltage compliance 12V at 20 mA (< 600Ω load)
Scaling Zero and full scale adjustable from -99999 to +99999
Resolution 16 bits (0.0015% of full scale)
Isolation 250V rms working, 2.3 kV rms per 1 min test
Relay Outputs (optional)
Dual magnetic relays 2 Form C, 10A max, 440Vac or 125Vdc max, 2500VA or 300W
Quad magnetic relays 4 Form A (NO), 10A max, 440Vac or 125Vdc max, 2500VA or 300W
Dual solid state relays 2 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Quad solid state relays 4 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Relay commons Isolated commons for dual relays or each pair of quad relays
Relay isolation 250V rms working, 2.3 kV rms per 1 minute test
Relay latching modes Latching or non-latching
Relay active modes Active on or off, active high or low
Hysteresis modes QA passband mode, split hysteresis, span hysteresis
Environmental
Operating temperature -40°C to 70°C (-40°F to 158°F)
Storage temperature. -40°C to 85°C (-40°F to 185°F)
Relative humidity 95% at 40°C, non-condensing
Protection NEMA-4X (IP-65) when panel mounted
Mechanical
Enclosure 1/8 DIN, high impact plastic, UL 94V-0, color: black
Mounting 1/8 DIN panel cutout required: 3.622" x 1.772" (92 mm x 45 mm).
Dimensions 4.68" x 2.45" x 5.64" (119 mm x 62 mm x 143 mm) (W x H x D)
Maximum panel thickness 4.5 mm (0.18")
Tightening Torque - Connectors Screw terminal connectors: 5 lb-in (0.56 Nm)
Tightening Torque - Pawls Digital Panel Meter Case Pawls: 5 lb-in (0.56 Nm)
Weight of base meter 210 g (7.4 oz) typical (DPM, counter, timer, 6-digit remote display)
Weight of option boards 30 g (1.0 oz) typical per board (analog output, relay output, communications)
General
Programming Methods Four front panel buttons or via Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options include using the front panel buttons, the free Instrument Setup Software, or a hardware jumper.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Laureate™ 1/8 DIN Case For Laureate Digital Panel Meters, Counters, Timers & Remote Displays

Laurel panel meter case

Key Features

  • Meets 1/8 DIN Standard.
  • Installs from front of panel.
  • Short depth behind the panel: only 4" (102 mm) plus connectors.
  • Understated 0.157" (4 mm) thick bezel.
  • Meets NEMA 4X (IP-65) for high-pressure wawshdon when panel mounted.
  • Screw clamps connectors meet VDE / IEC / UL / CSA safety standards.
  • Rugged GE Lexan® housing material.
  • Safety certified per EN 61010-1.
CE CertifiedRoHs CertifiedETL Certified

Dimensions

Mechanical specifications of Laureate digital panel meters and electronic counters

Maximum panel thickness: 4.5 mm (0.18")
Weight of base meter:
210 g (7.4 oz) typical (DPM, counter, timer, 6-digit remote display)
Weight of option boards: 30 g (1.0 oz) typical per board (analog output, relay output, communications)
Tightening Torque - Connectors: Screw terminal connectors: 5 lb-in (0.56 Nm)
Tightening Torque - Pawls: Digital Panel Meter Case Pawls: 5 lb-in (0.56 Nm)
Dimensioned CAD assembly drawings in EPRT, STEP, x_t. dwg, pdf file formats: Laureate-meter-case.zip (zipping prevents browser from opening CAD files as text files).

Panel Mounting

Mechanical Panel Mounting GuideSlide the meter into a 45 x 92 mm 1/8 DIN panel cutout. Ensure that the provided gasket is in place between the front of the panel and the back of the meter bezel. The meter is secured by two pawls, each held by a screw, as illustrated. Turning each screw counterclockwise extends the pawl outward from the case and behind the panel. Turning each screw clockwise further tightens it against the panel to secure the meter.

Turning each screw counterclockwise loosens the pawl and retracts it into its well. This position allows installed meter to be removed from their panel, or new meters to be installed in a panel. Do not remove the screws from their pawls. Doing so would cause the screw and pawl to fall off and likely get lost. Do not overtighten so as not to damage the plastic parts.

 

Duty Cycle & Pulse Width Modulation (PWM) Modes
Remote digital displays addressed on a single RS485 serial data line
Monitoring Laser OperationPM and Speed
digital indicator The display of a Laureate counter can be augmented by up to three slave displays to show up to four parameters simultaneously : Items #1, #2, #3 and peak rate (if rate was one of the selected items). All four parameters can also be displayed at a remote location.

 

 

CAL-Digital

Certificate of Calibration

$65.00

DLS-XLOG2

XLog2 Data logging Software

$495.00

IPC

Splashproof Cover

$55.00

CON01

CON01 Connector

$75.00

CBL01

RS232 Cable for Meters

$35.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL05

USB Data Cable for Meters

$47.00

CBL06

USB-to-RS485 Adapter Cable

$47.00

CBL07

USB Programming & Data Cable

$47.00

CBL08

RS485 Splitter Cable

$33.00

CBL6

6-foot Power Cable

$41.00

CBL12

12-foot Power Cable

$47.00

Modular Design for Maximum Flexibility at Minimum Cost

All boards are isolated from meter and power grounds. Optional Plug-in-Play boards for communications and control include Ethernet, WiFi, serial communication boardsdual or quad relay boards, and an analog output board. Laureates may be powered from 85-264 Vac or optionally from 12-32 Vac or 10-48 Vdc. The display is available with bright red or green 0.56" (14.2mm) high LED digits. The 1/8 DIN case meets NEMA 4X (IP65) specifications from the front when panel mounted. Any setup functions and front panel keys can be locked out for simplified usage and security. A built-in 5, 10, 12, or 24 Vdc excitation supply can power transducers, eliminating the need for an external power supply. All power and signal connections are via UL / VDE / CSA rated screw clamp plugs.

The Laureate™ Series features modular design with up to 7 isolated plug-in boards, applicable to all Laureate 1/8 DIN Panel Meters.

Schematic for Digital Panel Meter

Modular Hardware

The design of the Laureate™ Series is modular for maximum flexibility at minimum cost. All boards are isolated from meter and power grounds. The base configuration for panel meters or counter consists of a main module (with computer and plug-in display boards), a power supply board, and a signal conditioner board. Optional plug-in-play boards include an isolated setpoint controller board, an isolated analog output board, and an isolated digital interface board. Modular design and a choice of plug-in options allow the Laureate to be customized for a broad range of applications from simple monitoring to control and computer interface. There can be up to five plug-in boards in a 1/8 DIN Laureate.

Dual Board sets

Connecting Laureate Panel Meters to a Local Area Network (LAN)

Up to 30 Laureate Panel Meters and/or LT Transmitters can be configured for RS485 and daisy-chained to an LT Transmitter using Laurel’s High Speed Ethernet-to-RS485 converter board for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for Panel Meters

Laureate Panel Meters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

Ordering Guide
Part Number as Configured: L60001
Price as Configured: $325.00

Click on the Option Board Links for More Product Information

Base Item
$169.00
Display Color
$0.00
$0.00
Power (Isolated) $75.00
$115.00
Relay Output (Isolated)
$0.00
$101.00
$70.00
$128.00
$96.00
Analog Output (Isolated)
$0.00
$115.00
$172.00
Data Interface (Isolated) $81.00
$109.00
$135.00
$81.00
$135.00
$237.00
$259.00
$182.00
$204.00
$171.00
$193.00
$204.00
$226.00
Part Number as Configured:
L60001
Price as Configured:
$325.00
Quantity:
- +
Extended Price:
$325.00

Understanding the Laureate™ 1/8 DIN Panel Meters Remote Displays for Serial Input

The Laureate™ 1/8 DIN Panel Meters Remote Displays (or Serial Input Panel Meters) are slave displays that accept RS232, RS485, or USB data from computers, programmable controllers, Laureate instruments, or other devices with a streaming serial data output. They can also provide relay closures and an analog output based on the received readings, and blend in with 1/8 DIN Laureate panel meters, counters, and timers to provide a numeric 6-digit display from -999,999 to +999,999. Accuracy is 0.01% of reading ± 2 counts.

Streaming Data and Multi-Parameter Display

Streaming serial data can be generated at rates up to 9600 baud by a weighing scale or other instrument, particularly Laureate panel meters, counter/timers, or transmitters. A Remote Display can duplicate the reading of that instrument, or display additional data items transmitted serially that can't be shown simultaneously on a single display — for example, while a Laureate counter can only display one selected parameter at a time (such as rate A), it can transmit multiple parameters (rate A, rate B, ratio A/B, and peak). An indicator light shows which item has been chosen for display.

Character Masking for Non-Laureate Sources

Selectable modes ease interface to other Laureate instruments, extracting 1, 2, or 3 data items directly. For other instruments, readings are extracted from streaming ASCII strings containing multiple data values and non-numeric characters, such as Start and Stop characters. Any number of characters between the Start character and the data can be masked Off, up to 8 display characters (including sign and decimal point) can be masked On, and any number of characters between the last displayed character and the Stop character can be masked Off.

Analog Output and Relay Options

An optional isolated analog output board lets the Remote Display serve as a highly accurate digital-to-analog converter and transmitter. An optional relay output board with 2 or 4 relays adds remote alarm or control capability, using 8A contact relays or opto-isolated 120 mA AC/DC solid state relays. Relays can respond either to the transmitted values themselves or to any of 8 serially transmitted control characters — control characters generated by the source Laureate panel meter, counter, or timer specifically ensure the local and remote alarm points stay identical.

Real-World Application

A Laureate counter's display can be augmented by up to three slave displays to show up to four parameters simultaneously: Items #1, #2, #3, and peak rate (if rate was one of the selected items). All four parameters can also be displayed at a remote location.

Factory-Calibrated Accuracy

All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM. Factory recalibration is recommended annually.

Where Serial Input Remote Display Panel Meters Are Used

  • Multi-Parameter Process Displays — showing rate, total, and peak simultaneously from a single source counter that can only display one item locally at a time.
  • Remote/Hazardous-Area Monitoring — duplicating a reading at a safe distance from the primary sensor or process.
  • Weigh Scale & Third-Party Instrument Readout — displaying data from non-Laureate serial-output devices via character masking.
  • Networked Multi-Point Monitoring — up to 30 Laureate meters/displays daisy-chained on a single RS485 line.
  • Synchronized Local/Remote Alarming — control-character-driven relays ensuring a remote alarm exactly matches the source instrument's alarm condition.
  • Retrofit Digital-to-Analog Conversion — converting a serial data stream into a 4-20 mA or 0-10V signal for legacy analog equipment.
  • Large-Facility Centralized Monitoring — Ethernet or WiFi-networked remote displays reporting to a central control room.

Serial Input Remote Display Panel Meter Frequently Asked Questions

Why would a Laureate counter transmit more parameters serially than it can display on its own front panel at once?

The front panel display is documented as being limited to showing one selected parameter at a time (such as rate A), but the underlying instrument is still continuously computing multiple values (rate A, rate B, ratio A/B, peak) — transmitting all of them serially, even though only one is shown locally, is what allows one or more remote displays to show the additional parameters that don't fit on the source instrument's own single display.

What's the practical difference between masking characters "Off" versus masking characters "On" when extracting data from a non-Laureate ASCII string?

Masking a character "Off" tells the Remote Display to ignore that character entirely (such as Start/Stop delimiters or irrelevant text in the string), while masking up to 8 characters "On" specifically identifies which characters (including sign and decimal point) actually constitute the numeric reading to be displayed — together, this lets the meter extract just the meaningful number from a data string that also contains other non-numeric content.

Why would relays respond to serial control characters instead of simply comparing the transmitted value against a locally configured setpoint?

Documented rationale specifically ties this to keeping local and remote alarm points identical — if the remote display independently evaluated the transmitted value against its own separately configured setpoint, a configuration mismatch between the source instrument and the remote display could cause the two alarm points to diverge; using control characters generated directly by the source instrument means the remote relay fires exactly when the source instrument itself determines an alarm condition exists, eliminating that mismatch risk.

Can the Remote Display's analog output board work with a serial data source that isn't a Laureate instrument?

Yes, in principle — since the analog output board operates on whatever numeric value the Remote Display has extracted and is currently showing (via either direct Laureate interfacing or the character-masking extraction method for other instruments), a properly configured character-masking setup for a non-Laureate serial source would let the analog output board convert that extracted value the same way it would a Laureate-sourced reading.

Does duplicating a Laureate meter's reading on a Remote Display require any different setup than extracting one specific item from a multi-item stream?

Documented selectable modes specifically distinguish these use cases — one mode is oriented toward straightforward interfacing with other Laureate instruments to extract 1, 2, or 3 data items directly (a simpler setup, since the data format is already known and structured), while the character-masking approach is specifically documented for extracting readings from streaming ASCII strings from other, non-Laureate instrument types with less predictable formatting.

If I add a fourth or fifth slave display to a single source counter, will all of them show the same limit of four total parameters?

The documented example specifically describes up to three slave displays augmenting one source counter's own display to show up to four parameters total (Items #1, #2, #3, and peak) — this reflects the number of distinct parameters the source counter is documented as tracking and transmitting in that configuration, not a hard limit on the number of physical remote displays that could theoretically be added if more distinct data items were being transmitted.

Does the Remote Display need its own signal conditioner board the way a standard Laureate panel meter does?

No — since the Remote Display's entire input mechanism is documented as accepting streaming serial data (RS232, RS485, or USB) rather than a direct analog or pulse signal from a transducer, it doesn't require the kind of signal conditioner board (VF, FR, etc.) that standard Laureate meters use to interface with physical sensors; its core function is receiving and displaying already-digitized data.

Can the indicator light showing which item is selected for display be changed remotely, or does it require physically pressing a button on the Remote Display?

Documented setup guidance specifically ties item selection to configuration done via Instrument Setup Software or front panel programming — while the specific mechanism for changing which item is actively displayed during normal operation isn't detailed beyond the indicator light showing the current selection, the underlying configuration of what items are available to select is done through the standard setup process.

Does the Remote Display's 0.01% of reading ± 2 counts accuracy spec depend on how accurate the original source instrument's measurement is?

Yes — since the Remote Display is receiving an already-digitized numeric value over the serial link rather than independently measuring a physical signal, its documented accuracy specifically describes how faithfully it reproduces the transmitted value on its own display, not the accuracy of whatever measurement the source instrument originally made; the overall system accuracy is the combination of the source instrument's own measurement accuracy plus the Remote Display's reproduction accuracy.

Can the Remote Display accept data from more than one source instrument at the same time, or is it limited to a single serial data stream?

The documented operating principle describes the Remote Display as a slave unit receiving a single streaming serial data connection (RS232, RS485, or USB) from one source at a time — while an RS485 multi-drop bus can carry addressed data from multiple source instruments to multiple displays on the same physical line, an individual Remote Display is documented as extracting and showing data from its configured source connection, not simultaneously merging separate readings from multiple unrelated sources into one display.

RS485 Multi-Drop Networking & Termination Questions From the Field

Why does an RS485 multi-drop network need termination resistors at all if the devices are already electrically connected?

Documented technical guidance specifically explains that termination resistors match the network's characteristic impedance (typically 120 ohms for twisted-pair cable) to prevent signal reflections — without proper termination, signals traveling down the bus can reflect back from the unterminated ends and interfere with the original signal, corrupting data and causing communication errors.

Why do RS485 networks specifically need bias resistors in addition to termination resistors?

Documented explanations specifically distinguish these as addressing different problems — termination prevents signal reflections, while bias resistors address the "tri-state" or floating condition that occurs when no device on the bus is actively transmitting; without biasing, this floating state can be misread by receivers as noise or phantom start bits, documented as potentially causing a receiver to output the wrong logic state or oscillate.

If my master device already has internal bias built in, do I still need to add external bias resistors?

Not necessarily, and documented guidance specifically warns against assuming you always do — checking whether the master device already has internal bias before adding external bias resistors matters because over-biasing (stacking additional bias on top of a device that already provides it) can reduce the signal voltage margin, potentially causing its own communication problems rather than solving one.

Is the A/B (or D0/D1) wire labeling for RS485 devices standardized across manufacturers?

No — documented guidance specifically flags this as inconsistent across vendors, noting the Modbus specification itself technically uses D0 (inverting) and D1 (non-inverting) designations, while many device manufacturers instead label these terminals A and B, but without a universally consistent mapping between the two naming conventions; swapping the two wires is documented as the single most common RS485 wiring mistake when communication fails after initial wiring.

How many devices can realistically be connected on a single RS485 multi-drop line before needing a repeater?

Documented standard practice specifically cites a maximum of 32 devices on a single RS485 line without repeaters, extendable to as many as 256 devices when repeaters are added — this specification is a genuine hard limit tied to the electrical loading characteristics of standard RS485 transceivers on a shared bus, not just a recommended guideline.

Does a device address conflict on an RS485 network cause a visible error message, or does it fail silently?

Documented troubleshooting guidance specifically flags address conflicts as a real, distinct failure mode worth checking for specifically — while exact symptoms can vary by implementation, an unresponsive or intermittently misbehaving slave device is a documented symptom associated with address conflicts, meaning this often needs to be actively investigated and ruled out rather than always producing an obvious, self-explanatory error.

Can standard Cat5e/Cat6 Ethernet cable be used for an RS485 network instead of purpose-built RS485 cable?

Yes, documented guidance specifically confirms this works well for most short-to-medium industrial runs, since Cat5e/Cat6 is twisted-pair cable with an impedance (100Ω) reasonably close to the RS485 standard (120Ω) — the documented caveat is that the A and B signals must use a single matched twisted pair within the cable (not split across different color pairs), or the noise-canceling benefit of the twist is lost.

In a master/slave RS485 network, can two slave devices ever transmit data at the same time?

No, by design — documented protocol behavior specifically describes RS485 master/slave communication as strictly sequential: the master sends a request and then all devices listen, only the specifically addressed slave responds, and then all devices return to listening again, with half-duplex RS485 operation ensuring no two devices transmit simultaneously and avoiding data collisions on the shared bus.