The Laureate™ load cell, strain gauge and microvolt meter is a 5-digit panel meter with exceptionally high accuracy and stability for use with low-level signals.
Meter accuracy in ratiometric bridge operation is 0.01% of reading ± 2 counts. Custom curve linearization, which is available with the optional Extended main board, can extend the working range of load cells. Custom curve linearization also allows greater accuracy to be achieved with lower cost, less linear load cells.
A built-in isolated excitation supply can provide up to 120 mA of current at 10V to power four 350-ohm load cells in parallel. The meter operates in a ratiometric mode to eliminate errors due to supply variations. When excitation sense inputs are used in 6-wire connection, the meter compensates for variation in resistance of the transducer leads, thereby allowing long cable runs.
High read rates at up to 60 or 50 conversions per second while integrating the signal over a full power cycle are provided by Concurrent Slope (US Pat 5,262,780) analog-to-digital conversion. High read rates are ideal for peak or valley capture, real-time computer interface, and control.
Peak and valley values are automatically captured. These may be displayed via a front panel pushbutton command or control signal at the rear connector, or be transmitted as serial data.
Digital filtering is selectable for electrically noisy environments, including a batch averaging filter and an adaptive moving average filter which provides a choice of 8 time constants from 80 ms to 9.6 s. When a significant change in signal level occurs, that filter adapts by briefly switching to the shortest time to follow the change, then reverts back to the selected time constant. In a selectable Auto filter mode, the filter time constant is automatically selected based on detected signal noise.
Scale and offset can be provided by either of two user-selectable methods:
Auto-tare is standard for weighing applications to subtract the weight of an empty container and may be controlled by an external pushbutton contact closure or a logic signal. Additional capabilities for weighing applications are provided by the Laureate weight & scale meter.
Designed for system use. Optional plug-in boards for communications and control include Ethernet, WiFi and serial communication boards, dual or quad relay boards, and an isolated analog output board. Laureates may be powered from 85-264 Vac or optionally from 12-32 Vac or 10-48 Vdc. The display is available with red or green LEDs. The 1/8 DIN case meets NEMA 4X (IP65) specifications from the front when panel mounted. Any setup functions and front panel keys can be locked out for simplified usage and security. A built-in isolated 5, 10, or 24 Vdc excitation supply can power transducers and eliminate the need for an external power supply. All power and signal connections are via UL / VDE / CSA rated screw clamp plugs.
In 4-wire connection, the excitation and sense lines are tied together. The meter can make ratiometric corrections for supply voltage variations, but not compensate for variations in lead resistance. This connection is often used with short cable runs. | |
In 6-wire connection, the sense lines are separate from the excitation lines, thereby eliminating effects due to variations in lead resistance. This allows long cable runs in outdoor environments with temperature extremes. | |
For large scales, up to four 350 ohm load cells can be powered by a single Laureate, whose excitation output is rated 120 mA at 10V. The excitation and sense points of the four bridges are connected in parallel. The load cell outputs will be averaged if the load cells have the same sensititivity in mV/V. |
Seat belt tensile strength testing using Laureate digital panel meters.
The three meters display Force, Displacement and Speed.
Input Range mV | Resolution | Input Resistance | Error at 25°C |
---|---|---|---|
±20.000 | 1 µV | 1 GΩ | ±0.01% of full scale ±2 counts |
±50.000 | 1 µV | ||
±100.00 | 10 µV | ||
±250.00 | 10 µV | ||
±500.00 | 10 µV |
Full-Scale Input, mV | Zero Adjust | Span Adjust | Input Resistance | Error at 25°C |
---|---|---|---|---|
±20.00 | 0 to ±99,999 | 1 GΩ | ±0.01% of reading ±2 counts |
|
±50.00 | -99,999 | |||
±100.00 | to | |||
±250.00 | +99,999 | |||
±500.00 |
Accuracy | |
---|---|
Error at 25°C | ±0.01% of reading ± 2 counts in ratiometric mode, ±0.01% of full scale ± 2 counts in absolute mode. |
Span tempco | 0.0015% of reading/°C |
Zero tempco | 0.1 µV/°C |
Load cell linearization | Programmable using Extended meter version |
Noise Rejection | |
CMR, DC to 60 Hz | 130 dB |
NMR at 50/60 Hz | 90 dB with min filtering |
A-to-D Conversion | |
Technique | Concurrent Slope™ (Pat 5,262,780) |
A-to-D rate | 60/s at 60 Hz, 50/s at 50 Hz |
Output update rate | 56/s at 60 Hz, 47/s at 50 Hz |
Display update rate | 3.5/s at 60 Hz, 3/s at 50 Hz |
Display | |
Readout | 5 LED digits, 7-segment, 14.2 mm (.56"), red or green |
Range | -99999 to 99999 or -99990 to 99990 (count by 10) |
Indicators | Minus sign, 2 red LED lamps |
Power Supply Boards (one required) | |
Voltage, standard | 85-264 Vac or 90-300 Vdc |
Voltage, optional | 12-32 Vac or 10-48 Vdc |
Frequency | DC or 47-63 Hz |
Power consumption (typ. with four 350Ω load cells at 10V) | 2.4W @ 120 Vac, 2.7W @ 240 Vac, 2.5W @ 10 Vdc, 2.6W @ 20 Vdc, 2.75W @ 30 Vdc, 3.0W @ 40 Vdc, 3.35W @ 48 Vdc |
Power Isolation | 250V rms working, 2.3 kV rms per 1 min test |
Excitation Output (standard) | |
5 Vdc | 5 Vdc ± 5%, 100 mA |
10 Vdc | 10 Vdc ± 5%, 120 mA |
24 Vdc | 24 Vdc ± 5%, 50 mA |
Output Isolation | 50 Vdc to meter ground |
Ratiometric operation | Automatically compensates for changes in excitation level. |
Analog Output Board (one optional) | |
Output levels | 4-20 mA, 0-20 mA, 0-10V, -10 to +10V (jumper selectable) |
Current compliance | 20 mA at 10V ( > 5 kΩ load) |
Voltage compliance | 12V at 20 mA (< 600 Ω load) |
Scaling | Zero and full scale adjustable from -99999 to +99999 |
Resolution | 16 bits (0.0015% of full scale) |
Isolation | 250V rms working, 2.3 kV rms per 1 min test |
Relay Output Boards (one optional) | |
Dual magnetic relays | 2 Form C, 8A at 250 Vac or 24 Vdc, 0.3A at 250 Vdc, resistive load |
Quad magnetic relays | 4 Form A (NO), 8A at 250 Vac or 24 Vdc, 0.3A at 250 Vdc, resistive load |
Dual solid state relays | 2 Form A (NO), 120 mA at 140 Vac or 180 Vdc, resistive load |
Quad solid state relays | 4 Form A (NO), 120 mA at 140 Vac or 180 Vdc, resistive load |
Relay commons | Isolated commons for dual relays or each pair of quad relays |
Relay isolation | 250V rms working, 2.3 kV rms per 1 minute test |
Relay latching modes | Latching or non-latching |
Relay active modes | Active on or off, active high or low |
Hysteresis modes | QA passband mode, split hysteresis, span hysteresis |
Communication Boards (one optional) | |
Board selections | RS232, RS485 with dual RJ11 connectors, RS485 with dual RJ45 connectors, USB, Ethernet, USB-to-RS485 gateway, Ethernet-to-RS485 gateway, WiFi with built-in antenna plus USB & RS485, WiFi with external antenna plus USB & RS485 |
Protocols | Laurel Custom ASCII (serial), Modbus RTU (serial), Modbus TCP (Ethernet or WiFi) |
Digital addresses | 247 (Modbus), 31 (Laurel ASCII), |
Isolation | 250V rms working, 2.3 kV rms per 1 min test |
Signal Connections | |
Environmental | |
Operating temperature | 0°C to 55°C standard, -40°C to 70°C with -X option |
Storage temperature. | -40°C to 85°C |
Relative humidity | 95% at 40°C, non-condensing |
Protection | NEMA-4X (IP-65) when panel mounted |
Auto-Tare To read the net weight of an object, the empty container is first weighed, and an external button is pushed to zero out the display. The meter will then read net weight when an object is added to the empty container. The tare value is stored in memory for subsequent readings. | |
Determining Volume Using Load Cells An easy way to determine volume of an irregularly shaped tank with no need for linearizing is to weigh the tank using load cells. The meter will automatically tare out the weight of the empty tank and then scale the load cell signals to units of volume, such as liters or gallons. | |
Capture for Tensile Strength of Wire Peak readings are automatically captured at rates up to 60 per second, while the display updates at a legible 3.5 per second. The peak reading can be recalled at the push of a button or be always displayed. It can also be transmitted to a computer via RS232, RS485, USB or Ethernet. |