LTE DIN Rail Digital Transmitters with Ethernet Communication and Analog Outputs for Pulse to 4-20 mA Converter for Frequency, Rate, or Period Applications

LTE DIN Rail Digital Transmitters with Ethernet Communication and Analog Outputs for Pulse to 4-20 mA Converter for Frequency, Rate, or Period Applications

Price: $558.00
  • P/NLTE60FR
- +

Features

  • Inputs from NPN or PNP proximity switches, contact closures, digital logic magnetic pickups down to 12 mV, or AC inputs up to 250 Vac
  • Arithmetic functions A+B, A-B, AxB, A/B, A/B-1 (draw)
  • Two independently field-scalable pulse input channels from 0.005 Hz to 1 MHz
  • Line frequency measurement to 60.0000 in a few line cycles
  • 4-20 mA, 0-20 mA, 0-10V or -10V to +10V transmitter output, (isolated)
  • Analog output resolution 0.0015%  of span, accuracy ±0.02%  of span
  • Ethernet data I/O, Modbus TCP
  • Dual 120 mA solid state relays for alarm or control (isolated)
  • 5V, 10V, 12V, or 24V dc transducer excitation output (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)
    Optional - Extended main board for simultaneous rate and total, custom curve linearization, arithmetic functions A+B, A-B, AxB, A/B, A/B-1

The Laureate™ LTE Series DIN rail analog transmitter with ethernet communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy with Input frequencies from 0.005 Hz to 1 MHz. The LTE Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and ethernet output transmitter for pulse to 4-20 mA converter accepts two independently scalable input channels from a wide range of pulse sources, such as NPN or PNP proximity switches, contact closures, digital logic, magnetic pickups down to 12 mV, or AC voltages to 250 Vac. Input frequencies can range from 0.005 Hz to 1 MHz. Applications include AC line frequency, RPM or speed from proximity switch inputs, and flow from turbine flow meter inputs.

  • With a Standard main board, the transmitter output can be scaled to track frequency in Hz, rate (such as gallons per minute), or period (inverse of frequency). Square root extraction is standard.

  • With an Extended main board, the transmitter output can track rate or totalized rate (such as gallons) whether the transducer output is linear, requires square root extraction, or requires custom curve linearization. The latter can be provided by a curvilinear spline fit with up to 180 data points. The transmitter can also count up to a preset total or down from a preset total to zero. Such applications typically make use of optional dual solid state relays, which are available as options. External reset of totals is via a special three-position screw terminal connector. The two input channels A & B can also be combined arithmetically to provide an analog transmitter output that tracks A+B (e.g., sum of two flows), A-B (e.g., difference of two flows), AxB (e.g., horsepower as product of force and RPM), A/B (ratio of two flow), and A/B-1 (draw or relative elongation of material between rollers).

Exceptional Accuracy and Stability. Laureate transmitters determine frequency by taking the inverse of period as measured with a calibrated quartz crystal time base. This results in extremely accurate and stable 6-digit internal readings (±999,999 counts), which are then processed in software. The analog output is generated by an ultra-linear 16-bit (65,536 step) digital-to-analog converter (DAC) for 0.02% output accuracy. The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

Standard Features of Laureate LTE Transmitters Include:

  • Ethernet I/O, (isolated). The supported protocols are Modbus RTU and ASCII, which are tunneled via Modbus TCP. Note that RS232 or RS485 data I/O is provided by Laurel's LT Series transmitters.
  • 4-20 mA, 0-20 mA or 0-10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 16-bit (0.0015 ) resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. The supply can drive 20 mA into a 500 ohm (or lower) load for 10V compliance, or 10V into a 5K ohm (or higher) load for 2 mA compliance.
  • Dual-channel pulse inputs for voltage signals, NPN or PNP proximity switches, contact closures, magnetic pickups or flow meters.
  • Dual solid state relays, (isolated). Available for local alarm or control. Rated 120 mA at 130 Vac or 180 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 s. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Removable screw terminal connections of Laurel transmitters

LTE series DIN rail transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LTE series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an ethernet serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LTE Transmitters to a Local Area Network (LAN)

Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration. 

Flexible Communication Options for LTE Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laureate Ethernet network by Laurel Electronics

Laureate™ Ethernet & 4-20 mA Transmitter for Frequency, Rate or Period

Signal Input
Types AC, pulses from NPN, PNP transistors, contact closures, magnetic pickups
Grounding Common ground for channels A & B
Channel A Frequency 0.005 Hz to 1 MHz
Channel B Frequency 0.005 Hz to 250 kHz
Minimum Signal Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V)
Maximum Signal 250 Vac
Noise Filter 1 MHz, 30 kHz, 250 Hz (selectable)
Contact Debounce 0, 3, 50 ms (selectable)
Time Base Accuracy Quartz crystal calibrated to ±2 ppm
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Analog Output (standard)
Output Levels 4-20 mA and 0-10 Vdc (selectable)
Compliance, 4-20 mA 10V (0-500Ω load)
Compliance, 0-10V 2 mA (5 kΩ load)
Output Resolution 16 bits (65,536 steps)
Output Accuracy 0.02% of output span
Output Update Rate Programmed gate time + 30 ms + 0-2 signal periods
Output Isolation 250V rms working, 2.3 kV rms per 1 minute test
Ethernet Data I/O (standard)
Type 10/100 Base-T Ethernet per IEEE 802.3
Data Rates 300, 600, 1200, 2400, 4800, 9600, 19200 baud
Output Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Protocol Modbus TCP
Modbus Compliance Modbus over Serial Line Specification V1.0 (2002)
Digital Addresses 247
Dual Relay Output (standard)
Relay Type Two solid state relays, SPST, normally open, Form A
Load Rating 120 mA at 140 Vac or 180 Vdc
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Power Input
Standard Power 85-264 Vac or 90-300 Vdc
Low Power Option 10-48 Vdc or 12-32 Vac
Power Frequency DC or 47-63 Hz
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Power Consumption 2.5W typical at 24V, 4W with max excitation output
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Cooling Required Mount transmitters with ventilation holes at top and bottom. Leave 6 mm (1/4") between transmitters, or force air with a fan.
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options available using Laurel's free Instrument Setup Software.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

LTE Ethernet Transmitter Pinout

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

AC Line Frequency
AC Line Frequency Meter or Transmitter Laureate frequency meters and transmitters accept line voltages up to 250 Vac and output line frequency to 6-digit accuracy (50.0000 or 60.0000) in a few line cycles. Fast low frequency response is achieved by timing the period and taking its inverse.
RPM and Speed
Tachometer and Pulse Input Rate Meter or Transmitter Laureate frequency meters and transmitters accept line voltages up to 250 Vac and output line frequency to 6-digit accuracy (50.0000 or 60.0000) in a few line cycles. Fast low frequency response is achieved by timing the period and taking its inverse.
Simultaneneous Flow Rate and Total
Simultaneous Rate Meter and Totalizer Mode of Frequency Meter or Transmitter Laureate frequency meters and transmitters are compatible with all flow meters which generate pulses at a frequency proportional to flow rate. The Extended version can display scaled rate or total for the same input at the push of a button, and alarm from both the rate and total.
Combining Two Rates
Dual Pulse Input Rate Meter or Transmitter to Combine Two Rates Extended Laureate frequency meters and transmitters offer A+B, A-B and A/B arithmetic functions. A+B allows two input flows to be summed for total flow, while A-B allows outflow to be subtracted from inflow for net flow. Flow ratios aid in the proper mixing of ingredients.
Custom Curve Linearization
Custom Curve Linearization for Laureate Rate Meters and Transmitters Extended Laureate frequency meters and transmitters can linearize the output turbine flow meters, which tend to be nonlinear on the low end. Linearizing improves the dynamic range and accuracy of turbine flow meters.
System-level Capabilities
Laureate dual-channel rate meters and transmitters can independently scale, display and alarm two pulse input channels. All signal or alarm data can further be transmitted via RS232 or RS485, including peak readings and arithmetic combinations of the two rates. System Interfaces of Laureate Frequency Meters, Rate Meters and Transmitters

 

 

CAL-Digital

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LTE60FR
Price as Configured: $558.00

Click on the Option Board Links for More Product Information

Base Item
$388.00
Main Board
$0.00
With Standard Main Board: Scalable to ±999,999 for frequency, rate, square root of rate, up or down total, period, A-to-B time interval.
With Extended Main Board: Above, plus rate and total simultaneously, ratio (A/B), draw (A/B-1), other arithmetic functions (AxB, A+B, A-B), phase angle, stopwatch, up/down counting, batching operation, linearization of nonlinear inputs.
$55.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
Part Number as Configured:
LTE60FR
Price as Configured:
$558.00
Quantity:
- +
Extended Price:
$558.00

Understanding the Laureate™ LTE Series DIN Rail Transmitter for Frequency, Rate & Period Input

The Laureate™ LTE Series DIN rail transmitter for frequency, rate, or period accepts two independently scalable input channels from a wide range of pulse sources, such as NPN or PNP proximity switches, contact closures, digital logic, magnetic pickups down to 12 mV, or AC voltages to 250 Vac. Channel A accepts frequencies from 0.005 Hz to 1 MHz; Channel B accepts 0.005 Hz to 250 kHz. Applications include AC line frequency, RPM or speed from proximity switch inputs, and flow from turbine flow meter inputs.

Standard vs. Extended Main Board

With the Standard main board, the transmitter output can be scaled to track frequency in Hz, rate (such as gallons per minute), or period (inverse of frequency); square root extraction is standard. With the Extended main board, the output can track rate or totalized rate whether the transducer output is linear, requires square root extraction, or requires custom curve linearization (via curvilinear spline fit with up to 180 data points). The Extended board also supports counting up to a preset total or down from a preset total to zero, with external reset via a special three-position screw terminal connector, and arithmetic combination of the two input channels: A+B, A-B, AxB, A/B, and A/B-1 (draw).

Signal Specifications

Nine minimum signal ranges span -12 to +12 mV up to +1.25 to +2.1V. Maximum signal is 250 Vac. Noise filter is selectable at 1 MHz, 30 kHz, or 250 Hz; contact debounce is selectable at 0, 3, or 50 ms. Time base accuracy is a quartz crystal calibrated to ±2 ppm. Output update rate is programmed gate time + 30 ms + 0-2 signal periods — for a 60 Hz signal, this is 20 per second.

AC Line Frequency Measurement

Laureate frequency meters and transmitters accept line voltages up to 250 Vac and output line frequency to 6-digit accuracy (50.0000 or 60.0000) in a few line cycles. Fast low-frequency response is achieved by timing the period and taking its inverse.

Ethernet Data I/O

Standard Ethernet Data I/O is 10/100 Base-T per IEEE 802.3, isolated to 250V rms working / 2.3 kV rms per 1 minute test, with Modbus TCP at digital address 247. Analog output levels are 4-20 mA and 0-10 Vdc (selectable), 16-bit resolution, 0.02% of output span accuracy.

Where LTE Frequency, Rate & Period Transmitters Are Used

  • Networked Generator & Grid-Tie Frequency Monitoring — 6-digit AC line frequency accuracy over Ethernet.
  • Networked Turbine Flow Measurement — rate/total flow with custom curve linearization.
  • Tachometer & RPM Monitoring Networks — proximity switch speed sensing with Modbus TCP output.
  • Multi-Flow Ratio & Blending Control — A/B ratio and A+B/A-B combined-flow monitoring.
  • Multi-Point Networked Rate Monitoring — several transmitters on one Modbus TCP network.
  • OEM Networked Frequency/Rate Instrumentation — DIN rail integration into Ethernet-based control panels.

LTE Frequency, Rate & Period Transmitter Frequently Asked Questions

Why does Channel A support a higher maximum frequency (1 MHz) than Channel B (250 kHz)?

Documented specification lists these as separate, distinct maximum frequencies for the two channels, without detailing the internal circuit reason for the difference — this is consistent with Channel A being documented as the primary, higher-bandwidth channel intended for the widest range of pulse sources, while Channel B's somewhat lower documented ceiling still comfortably covers common secondary-channel uses like a second flow input or a direction/inhibit signal in combined dual-channel applications.

Why does the documented output update rate specifically work out to 20 per second for a 60 Hz signal?

Documented formula specifically defines update rate as programmed gate time plus 30 ms plus 0-2 signal periods — at 60 Hz, each signal period is about 16.7 ms, so with a minimal gate time setting, the total documented delay (30 ms plus up to two 16.7 ms periods) works out to roughly 50 ms, consistent with the documented 20-per-second update rate cited specifically for a 60 Hz signal example.

Does achieving the documented 6-digit AC line frequency accuracy (50.0000 or 60.0000) require multiple complete line cycles to accumulate, or is it available faster?

Documented description specifically states this 6-digit accuracy is achieved "in a few line cycles," attributing the fast response specifically to timing the signal period and taking its inverse, rather than requiring a long fixed counting window; this inverse-period technique is documented as the specific mechanism that allows full 6-digit accuracy without needing to wait through many cycles.

Can the A/B-1 (draw) arithmetic function produce a negative output value, and what would that represent?

Documented description specifically identifies A/B-1 as representing "draw or relative elongation of material between rollers" — mathematically, if Channel A's rate is slower than Channel B's rate, the A/B ratio would fall below 1, making A/B-1 negative; this is consistent with representing relative compression or shrinkage between rollers rather than elongation, the opposite condition from the documented example use case.

Does selecting a narrower noise filter setting (such as 250 Hz) limit which of the two documented frequency channels can be used?

The documented noise filter options (1 MHz, 30 kHz, 250 Hz) are listed once under general Signal Input specifications without being tied to one specific channel — selecting a filter setting narrower than the actual signal frequency being measured on either channel risks attenuating that genuine signal, so the appropriate filter setting should be matched to the actual signal frequency on whichever channel it's applied to, rather than the filter setting itself restricting which channel can be used.

Does this LTE Frequency/Rate transmitter's documented Modbus TCP-only protocol limit compatibility compared to the RS232/RS485 LT Series frequency variant?

Yes — this page documents Modbus TCP specifically as the supported Ethernet Data I/O protocol at digital address 247, while the LT Series serial variant is documented elsewhere as separately supporting Modbus RTU/ASCII and Laurel Custom ASCII; a control system needing a protocol other than Modbus TCP would need to reference the LT Series serial variant rather than this LTE Ethernet variant.

Does the documented external reset for totalizing require a specific three-position screw terminal, or can any control input be used?

Documented description specifically states external reset of totals is "via a special three-position screw terminal connector," identifying this as a specific, dedicated hardware connection point rather than a generic control input function; this indicates the reset mechanism is documented as tied to that particular physical terminal arrangement rather than being freely assignable to any of the transmitter's other general-purpose control inputs.

Does custom curve linearization on this Extended-board LTE transmitter apply only to flow rate, or can it also correct the totalized value?

Documented description specifically frames custom curve linearization as correcting nonlinearity in "turbine flow meters, which tend to be nonlinear on the low end," improving dynamic range and accuracy — since the Extended board is documented as separately capable of displaying "rate or total for the same input," and both are derived from the same underlying linearized signal processing, the documented linearization capability is consistent with improving accuracy for both the rate and total derived from that corrected signal.

Does the documented ±2 ppm time base accuracy apply equally to frequency, rate, and period measurement modes?

Yes — documented specification lists the ±2 ppm quartz crystal time base accuracy once, under the general Signal Input section, without listing separate time base accuracy figures for frequency, rate, or period modes specifically; since all three modes are documented as derived from the same underlying crystal-timed period measurement, this single accuracy figure is consistent with applying across all three documented measurement modes.

Does power consumption increase specifically when using the maximum transducer excitation output, and by how much per the documented specification?

Yes — documented specification lists power consumption as "2.5W typical at 24V, 4W with max excitation output," directly quantifying the additional power draw when the transducer excitation output is set to its maximum documented level; this represents a genuine, specified increase in overall power consumption tied directly to how much excitation current is being supplied to an external transducer.

Power Grid Frequency Monitoring Questions From the Field

Why is grid frequency deviation treated as such a serious, actively managed parameter rather than a minor measurement curiosity?

Documented analysis specifically explains that grid frequency reflects the real-time balance between electricity supply and demand across the entire interconnected system — documented grid codes specifically prescribe acceptable frequency deviation limits, and system operators are documented as required to take corrective action to stay within them, since frequency is one of the few parameters that signals system-wide imbalance in real time.

Is there a documented typical magnitude of frequency deviation considered acceptable before automatic protective action is triggered?

Yes — one documented patent-level source specifically cites frequency deviations on the order of ±0.5 Hz as within acceptable limits for generating equipment, while deviations beyond that, even lasting only minutes, are documented as capable of causing significant damage to generating equipment; documented grid-level analysis separately notes automatic controls are designed to keep deviations to within roughly 100 mHz under normal system-wide operation.

Does a large frequency deviation event genuinely risk cascading, system-wide failures, or is this an overstated concern?

It's documented as a genuine, serious risk — documented analysis specifically describes mechanisms such as machine protections or under-frequency load shedding (ULFS) as automatically disconnecting critical network elements when frequency deviates significantly, with documented acknowledgment that this protective disconnection can itself potentially cascade into further failures and ultimately blackouts if not properly managed.

Is measuring the time between zero-crossings of the AC waveform documented as an accurate way to derive grid frequency, or does this method have known limitations?

Documented analysis specifically flags a common limitation with this approach — techniques assuming the waveform is a pure sine wave and using zero-crossing timing as a direct frequency proxy are documented as introducing their own error, since real grid waveforms aren't perfectly sinusoidal; documented critique specifically notes several prior approaches share this same simplifying-assumption weakness.

Does the increasing presence of renewable energy generation on the grid documented as changing how frequency deviation behaves or is managed?

Yes — documented analysis specifically identifies the gradual substitution of conventional electromechanical generation with inverter-based renewable sources as a genuine new concern for grid synchronization performance, specifically because renewable inverter interfaces are documented as lacking the natural physical inertia that conventional rotating generators provide, which traditionally helped dampen frequency swings from power imbalances.

Do grid frequency fluctuations follow a simple, predictable statistical pattern, or is their distribution documented as more complex?

Documented statistical analysis specifically finds grid frequency fluctuations deviate from a simple normal (Gaussian) distribution, particularly in the tails, being better characterized by other statistical models across several real-world grids studied (Continental Europe, Nordic, Great Britain, Japan, and North American Eastern Interconnection); this documented non-Gaussian behavior reflects that grid frequency fluctuation is a more statistically complex phenomenon than a simple bell-curve model would suggest.

Is frequency measurement genuinely node-dependent across a large interconnected grid, or is grid frequency essentially the same value everywhere at a given instant?

Documented analysis specifically notes that frequency measurements are node-dependent, reflecting the superposition of two distinct phenomena at any given measurement point: system-wide frequency changes from overall supply-demand imbalance, and separate inter-area frequency oscillations arising from weak dynamic coupling between different regions of the grid; this documented nuance means a single frequency reading at one location doesn't necessarily represent the instantaneous frequency everywhere else on the interconnected system.

Is locally generated power (such as from a private or backup generator) documented as requiring the same kind of frequency monitoring as utility grid power?

Yes — documented patent description specifically identifies locally generated AC power, including standby/emergency generation and cogeneration sources like private hydroelectric or waste-heat steam generators, as genuine, specifically documented applications for frequency monitoring circuitry, particularly relevant when such local generation is tied into the broader utility grid and must stay synchronized with it.