LTE DIN Rail Digital Transmitter with Ethernet Communication and Analog Outputs for Time of Periodic Event Applications

LTE DIN Rail Digital Transmitter with Ethernet Communication and Analog Outputs for Time of Periodic Event Applications

Price: $558.00
  • P/NLTE60FR
- +

Features

  • Times periodic events with width from 1 µs to 199.999 s
  • Transmits average time of periodic events with width from 1 µs to 199.999 s
  •  Resolution to 0.2 µs, rep rated to 250 kHz
  • Inputs from NPN or PNP proximity switches, contact closures, digital logic, or magnetic pickups down to 12 mV
  • Trigger on positive or negative pulse edges
  • 4-20 mA, 0-20 mA, 0-10V or -10V to +10V transmitter output, (isolated)
  • Analog output resolution 0.0015%  of span, accuracy ±0.02%  of span
  • Ethernet data I/O, Modbus TCP
  • Dual 120 mA solid state relays for alarm or control (isolated)
  • 5V, 10V, 12V, or 24V dc transducer excitation output (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)

The Laureate™ LTE Series DIN rail analog transmitter with ethernet communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy with Input frequencies from 0.005 Hz to 1 MHz. The LTE Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

Time Interval Mode for Time Delay

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and ethernet output transmitter for time interval can transmit pulse width or time delay between individual pulses to a resolution of 0.2 µs for periodic events. It can also transmit average pulse width or average time delay between multiple pulses.

Time interval is measured between inputs on channels A and B. Timing starts when a pulse is applied to Channel A (selectable positive or negative edge), and ends when a pulse is applied to Channel B (selectable positive or negative edge). In case of a single pulsed signal, the A and B inputs can be tied together. A positive or negative slope may be selected to start timing, and the opposite slope must be selected to stop timing. Timing is achieved by counting 5.5 MHz clock pulses. Multiple integral time intervals are averaged over a gate time which is selectable from 10 ms to 199.99 s and also controls the maximum output rate.

The dual-channel signal conditioner used for pulse detection accepts inputs from proximity switches with PNP or NPN output, TTL or CMOS logic, magnetic pickups, contact closures, and other signals from 12 mV to 250 Vac. Jumper selections provide optimum operation for different sensor types and noise conditions. A built-in 5V, 10V or 24V dc excitation supply can power proximity switches and other sensors, and eliminate the need for an external power supply.

Exceptional Accuracy and Stability. Laureate transmitters determine frequency by taking the inverse of period as measured with a calibrated quartz crystal time base. This results in extremely accurate and stable 6-digit internal readings (±999,999 counts), which are then processed in software. The analog output is generated by an ultra-linear 16-bit (65,536 step) digital-to-analog converter (DAC) for 0.02% output accuracy. The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

Standard Features of Laureate LTE Transmitters Include:

  • Ethernet I/O, (isolated). The supported protocols are Modbus RTU and ASCII, which are tunneled via Modbus TCP. Note that RS232 or RS485 data I/O is provided by Laurel's LT Series transmitters.
  • 4-20 mA, 0-20 mA or 0-10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 16-bit (0.0015 ) resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. The supply can drive 20 mA into a 500 ohm (or lower) load for 10V compliance, or 10V into a 5K ohm (or higher) load for 2 mA compliance.
  • Dual-channel pulse inputs for voltage signals, NPN or PNP proximity switches, contact closures, magnetic pickups or flow meters.
  • Dual solid state relays, (isolated). Available for local alarm or control. Rated 120 mA at 130 Vac or 180 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 s. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Removable screw terminal connections of Laurel transmitters

LTE series DIN rail transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LTE series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an ethernet serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LTE Transmitters to a Local Area Network (LAN)

Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration. 

Flexible Communication Options for LTE Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laureate Ethernet network by Laurel Electronics

Laureate™ Ethernet & 4-20 mA Transmitter of Time Interval of Periodic Events

Pulse Input
Types AC, pulses from NPN, PNP transistors, contact closures, magnetic
Signal Ground pickups
Minimum Signal Common ground for channels A & B.
Maximum Signal Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V).
Maximum Frequency 250 Vac
Contact Debounce 1 MHz, 30 kHz, 250 Hz (selectable).
Time Base Accuracy 0, 3, 50 ms (selectable).
Span Tempco Quartz crystal calibrated to ±2 ppm.
Long-term Drift ±1 ppm/°C (typ), ±5 ppm/year
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Time Interval Mode
Timing Start Channel A pulse, + or - edges
Timing Stop Channel B pulse, + or - edges
Update Rate Gate time + 30 ms + 0-2 time intervals
Gate Time Selectable 10 ms to 199.99 s
Time to Zero Output Selectable 10 ms to 199.99 s
Resolution
0 - 199.999 s 1 ms
0 - 99.9999 s 100 µs
0 - 9.99999 s 10 µs
0 - .999999 s 1 µs
0 - .099999 s 0.2 µs (after averaging)
Analog Output (standard)
Output Levels 4-20 mA and 0-10 Vdc (selectable)
Compliance, 4-20 mA 10V (0-500Ω load)
Compliance, 0-10V 2 mA (5 kΩ load)
Output Resolution 16 bits (65,536 steps)
Output Accuracy ±0.02% of output span
Output Update Rate Programmed gate time + 30 ms + 0-2 signal periods
Output Isolation 250V rms working, 2.3 kV rms per 1 minute test
Ethernet Data I/O (standard)
Type 10/100 Base-T Ethernet per IEEE 802.3
Data Rates 300, 600, 1200, 2400, 4800, 9600, 19200 baud
Output Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Protocol Modbus TCP
Modbus Compliance Modbus over Serial Line Specification V1.0 (2002)
Digital Addresses 247
Dual Relay Output (standard)
Relay Type Two solid state relays, SPST, normally open, Form A
Load Rating 120 mA at 140 Vac or 180 Vdc
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Power Input
Standard Power 85-264 Vac or 90-300 Vdc
Low Power Option 10-48 Vdc or 12-32 Vac
Power Frequency DC or 47-63 Hz
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Power Consumption 2.5W typical at 24V, 4.0W with max excitation output
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Cooling Required Mount transmitters with ventilation holes at top and bottom. Leave 6 mm (1/4") between transmitters, or force air with a fan.
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options available using Laurel's free Instrument Setup Software.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

LTE Ethernet Transmitter Pinout

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

Time Interval Mode for Time Delay
Time Interval Mode for Time Delay For periodic pulses applied to A and B channels, time delays can be measured down to 0.2 µs resolution from the rising or falling edge of A to the rising or falling edge of B (selectable).
Time Interval Mode for Pulse Width
Time Interval Mode for Pulse Width The width of periodic pulses (t1 or t2) can be measured by tying the A and B channels together. As for time delay, readings are averaged over a user-selectable gate time.
Timing Process Dynamics with two Meters or Transmitters
Timing Process Dynamics with a Panel Meter and Time Interval Meter The start and stop pulses used for timing can be generated by the dual relay board in a Laureate panel meter, counter or transmitter. For instance, the start and stop pulse edges can be created as temperature passes two alarm setpoints, or temperature cycles in a hysteresis control mode.
Replacing an Oscilloscope with a Laureate Time Interval Meter or Transmitter
Replacing an Oscilloscope with a Laureate Time Interval Meter An oscilloscope is great for viewing and timing pulses in a lab. However, in fixed installations where digital timing accuracy and control outputs are required, a low-cost Laureate time interval meter or transmitter will be the instrument of choice. Resolution to 0.2 µs is feasible.
Instrumenting a Pulsed Laser System
Instrumenting a laser system using Laureate timers and stopwatches
Some of the many possibilities in instrumenting a pulsed laser system with Laureate dual-channel counters and transmitters: elapsed time, number of pulses, pulse width, pulse separation, duty cycle, and pulse rep rate.

 

 

CAL-Digital

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LTE60FR
Price as Configured: $558.00

Click on the Option Board Links for More Product Information

Base Item
$388.00
Main Board
$0.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
Part Number as Configured:
LTE60FR
Price as Configured:
$558.00
Quantity:
- +
Extended Price:
$558.00

What Is the LTE DIN Rail Digital Transmitter with Ethernet Communication and
Analog Outputs for Time of Periodic Events?

Some processes don't care how fast something is moving — they care how long a signal stays high, or how much time passes between one pulse and the next. This LTE DIN Rail Transmitter measures exactly that: pulse width or time delay between individual pulses arriving on its A and B input channels, down to a resolution of 0.2 µs, and converts that timing data into an isolated 4-20 mA analog output plus Ethernet-based digital data.

What Is the LTE DIN Rail Digital Transmitter?

The LTE DIN Rail Digital Transmitter is a device used to convert the timing of periodic events into digital data that can be processed by industrial systems. Many processes depend on knowing how long an event lasts or how much time separates repeating events, and this transmitter provides a reliable interface that translates that timing into actionable information, communicating over Ethernet.

Pulse Width or Time Delay, Individual or Averaged

The transmitter can be configured to measure either the width of a single pulse or the time delay between separate pulses on its A and B channels, at a resolution of 0.2 µs. Rather than reporting every single measurement, it can also be set to transmit the average pulse width or average time delay calculated across multiple pulses, which smooths out event-to-event variation when what matters is the typical timing rather than any one occurrence.

Two-Channel (A/B) Timing Input

Time interval is measured between inputs arriving on channels A and B, rather than from a single input against an internal clock reference alone. This two-channel arrangement lets the transmitter time the gap between two genuinely separate events — for example, a part leaving one sensor and arriving at the next — rather than being limited to timing a single repeating pulse train.

DIN Rail Mounting and Panel Integration

The transmitter mounts on a standard DIN rail, the same mounting standard used across most industrial control panels, so it integrates alongside relays, breakers, and other panel-mounted equipment without custom bracketry. This standardization keeps installation and panel layout consistent with the rest of a facility's control equipment.

Ethernet Communication and Analog Output

Alongside its isolated 4-20 mA analog output, the transmitter communicates over Ethernet, allowing timing data to be read directly onto a plant network rather than requiring a separate serial-to-Ethernet gateway. The analog output lets the same reading feed legacy analog-only equipment — a PLC analog input, chart recorder, or panel meter — at the same time the digital value is available on the network.

Named Applications

  • Machine Cycle Timing — sensors placed at defined points in a machine cycle feed A and B channel inputs, and the transmitter reports the time delay between them so a repeatable cycle can be checked against its expected duration.
  • Gap or Dwell Timing on a Line — a part triggers channel A on arrival and channel B on departure from a station; the transmitter's pulse-width measurement reports how long the part actually dwelled there, useful for catching stations that are running slow.
  • Event Interval Trending — where an individual reading is noisy but the underlying timing is stable, the transmitter's average time-delay mode reports a smoothed value across multiple pulses, better suited to feeding a trend chart or a process control loop than a single raw reading.

Where Else Is This Transmitter Used?

  • Manufacturing — monitoring the timing of machine cycles and process events to optimize production efficiency and catch developing downtime before it becomes a stoppage.
  • Utilities — tracking periodic events in water and wastewater treatment processes to support regulatory compliance and operational reliability.
  • Building Automation — timing HVAC compressor cycles, fan operation, and lighting control events as part of building management and energy-efficiency programs.
  • Energy Management — timing power consumption or load events to identify peak usage windows and support load-balancing strategies.
  • Transportation and Logistics — timing vehicle engine run periods or idle intervals, and timing events like container loading and unloading, to support fleet and supply-chain efficiency.
  • Infrastructure Monitoring — logging the timing of inspection or maintenance-related events on assets such as bridges or tunnels.

Conclusion

The LTE DIN Rail Digital Transmitter for time of periodic events gives a panel builder a precise way to convert pulse-width and pulse-interval timing from two independent input channels into an isolated analog signal and Ethernet-based digital data — with 0.2 µs resolution, the option to report individual or averaged readings, and documented use in applications from machine cycle timing to utility process monitoring where knowing exactly how long something took, not just how often it happens, is the whole point.

LTE Time of Periodic Events Transmitter Frequently Asked Questions

What's the difference between pulse width and time delay measurement on this transmitter?

Pulse width measures how long a single input signal stays in its active state, while time delay measures the elapsed time between a pulse arriving on one channel and a pulse arriving on the other; which mode to use depends on whether the application cares about the duration of one event or the gap between two separate events.

Why does the transmitter measure between two channels (A and B) instead of just one?

A single channel can only time a pulse against itself, such as its own on-time; using two independent channels lets the transmitter time the interval between two physically separate events — one triggering A and a later one triggering B — which single-channel timing can't capture.

When should I use averaged readings instead of individual pulse measurements?

Individual readings are useful when each specific event's timing matters on its own, such as checking whether one particular cycle ran long; averaging across multiple pulses is better suited to noisy or naturally variable signals where the underlying, typical timing is what a control loop or trend chart actually needs.

What does 0.2 µs resolution mean in practical terms?

It's the smallest timing increment the transmitter can distinguish between two measurements; for applications timing very short events or requiring fine discrimination between near-identical intervals, this resolution determines how precisely two close readings can be told apart.

Does the Ethernet interface affect measurement accuracy?

No — timing accuracy comes from the transmitter's internal signal conditioning and timing circuitry; Ethernet only changes how the already-measured reading is transmitted digitally onto the network, separate from the isolated 4-20 mA analog output.

Do I need a gateway device to put this transmitter on my network?

No — the transmitter connects directly to a LAN via its own RJ45 jack, without requiring a separate serial-to-Ethernet gateway in between.

Can this transmitter provide both an analog signal and a digital network reading at the same time?

Yes — the isolated 4-20 mA analog output and the Ethernet-based digital data are both available from the same measurement, so one can feed a PLC analog input or chart recorder while the other is polled over the network.

What kind of signal do the A and B inputs expect?

They accept periodic pulse-type digital inputs, such as those from proximity sensors, photoeyes, or contact closures marking the start and end of an event — the same general category of signal used to trigger a timing measurement in most panel-level automation.

Is this transmitter suitable for DIN rail control panels alongside other equipment?

Yes — it mounts on standard DIN rail, the same mounting format used for relays, terminal blocks, and other panel components, so it integrates into an existing panel layout without custom mounting hardware.

How is this different from a transmitter that measures rate or frequency?

A rate or frequency transmitter reports how often events occur per unit of time; this transmitter instead reports the timing characteristics of the events themselves — how long a pulse lasts or how much time separates two pulses — which is a different measurement even when both are derived from a similar pulse train.

LTE Time of Periodic Events Transmitter Questions From the Field

My timing readings look inconsistent from cycle to cycle even though the process seems stable — what should I check?

Switching from individual pulse measurement to averaged pulse width or time delay across multiple pulses is the standard way to smooth out normal event-to-event variation; if readings are still inconsistent after averaging, sensor mounting and signal quality on the triggering channel are the next things to check.

My time delay reading seems to be measuring the wrong pair of events — what's the likely cause?

Since time delay is measured specifically between channel A and channel B, confirming which physical sensor or trigger is wired to each channel is the standard first check when the reported interval doesn't match what's physically happening on the line.

My pulse width reading reads much longer or shorter than the actual event duration — what should I check?

A mismatch between the reported pulse width and the actual event usually points to the triggering sensor's signal not cleanly matching the true start and end of the event — for example, sensor bounce or a threshold set too sensitively — so checking the sensor's signal quality and trigger threshold is the standard first step.

My transmitter isn't showing up on the network — what should I check first?

Confirming the transmitter's IP configuration matches what the network expects, and that the physical Ethernet cable and switch port are functioning, are the standard first checks before suspecting a transmitter fault.

My analog output reading and my Ethernet reading don't seem to match — why?

Both outputs are derived from the same underlying measurement, so a mismatch usually points to a scaling or range configuration difference between how the analog output and the digital reading were each set up, rather than two separate measurements being taken.

My averaged reading changes very slowly even though I can see individual events happening faster than that — is this normal?

Yes — averaging across multiple pulses is designed to smooth the reading over time rather than update on every single event; if faster response to each individual event is actually what's needed, switching to individual pulse measurement rather than averaging is the appropriate setting.

Can I use this transmitter if my two events are triggered by different types of sensors on channels A and B?

Yes — the A and B channels are independent digital inputs, so different sensor types can trigger each channel as long as both provide a signal the transmitter can read as a clean pulse; verifying each sensor's output type is compatible with the transmitter's input is the standard first step during setup.

My Modbus TCP polling occasionally times out even though the transmitter appears connected — what's the likely cause?

Network congestion or too many devices polling the same transmitter simultaneously can cause intermittent timeouts; checking polling frequency from all connected clients and network traffic load is a common troubleshooting step.