LT DIN Rail Digital Transmitters with Serial Data Communication and Analog outputs for Time of Single or Accumulated Event Applications

LT DIN Rail Digital Transmitters with Serial Data Communication and Analog outputs for Time of Single or Accumulated Event Applications

Price: $334.00
  • P/NLT60FR
- +

Features

  • Times single or cumulative events from 1 µs to 999,999 hrs
  • Transmits single event time or accumulated time of all events
  • Timing resolution to 0.2 µs
  • Timing from 0.2 µs to 999,999 hrs
  • Selectable HH.MM.SS clock format or 6-digit H, M or S decimal format
  • Inputs from NPN or PNP proximity switches, contact closures, digital logic, magnetic pickups down to 12 mV, or AC inputs up to 250 Vac
  • Triggers on positive or negative pulse edges
  • 4-20 mA, 0-20 mA, 0-10V or -10V to +10V transmitter output, (isolated)
  • Analog output resolution 0.0015%  of span, accuracy ±0.02%  of span
  • RS232 or RS485 serial data, Modbus or Laurel ASCII protocol (isolated)
  • Dual 120 mA solid state relays for alarm or control (isolated)
  • 5V, 10V, 12V, or 24V dc transducer excitation output (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)

The Laureate™ LT Series DIN rail analog transmitter with serial data communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy with Input frequencies from 0.005 Hz to 1 MHz. The LT Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

accumulated time

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and RS232/RS485 output transmitter for stopwatch puts out (isolated) analog and serial data signals whose values track the time of single events which produce start and stop pulses, or the accu­mulated time of multiple events. It can also time the width of a single pulse. The highest resolution is 0.2 µs, making the transmitter ideal for fast events. The longest timing interval is 999,999 hrs. For long events, the analog output is updated continuously during timing. There are two primary timing modes:

  • A-A Stopwatch Mode. Time is measured between a start pulse and a stop pulse, both on Channel A, from either the positive or negative edges.
  • A-B Stopwatch Mode. Time is measured between a start pulse on Channel A (positive or negative edge) and a stop pulse on Channel B (positive or negative edge). This mode allows inputs from different sources. In addition, the A and B inputs can be tied together to start the stopwatch with one polarity and stop it with the other polarity.

Event time (Item #1) is measured by counting 5.5 MHz clock pulses from a calibrated quartz crystal. The stopwatch output is updated during timing at a rate controlled by a gate time, up to 25/sec. Time is reset to zero when the next start pulse occurs. Accumulated time from multiple events up to 999,999 hours (Item #2) is also tracked.

The dual-channel signal conditioner used for pulse detection accepts inputs from proximity switches with PNP or NPN output, TTL or CMOS logic, magnetic pickups, contact closures, and other signals from 12 mV to 250 Vac. Jumper selections provide optimum operation for different sensor types and noise conditions. A built-in 5V, 10V, 12V, or 24V dc excitation supply can power proximity switches and other sensors, and eliminate the need for an external power supply.

Exceptional Accuracy and Stability. Laureate transmitters determine frequency by taking the inverse of period as measured with a calibrated quartz crystal time base. This results in extremely accurate and stable 6-digit internal readings (±999,999 counts), which are then processed in software. The analog output is generated by an ultra-linear 16-bit (65,536 step) digital-to-analog converter (DAC) for 0.02% output accuracy. The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

Standard Features of Laureate LT Transmitters Include:

  • Serial communications output, (isolated), RS232 or RS485 (half or full duplex), jumper selectable. Three protocols are user selectable: Modbus RTU, Modbus ASCII, or Laurel ASCII. Modbus operation is fully compliant with Modbus Over Serial Line Specification V1.0 (2002). The Laurel ASCII protocol is simpler than the Modbus protocol and is recommended when all devices are Laureates.
  • 4-20 mA, 0-10V or -10V to +10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 0.0015% resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. Note that Ethernet data I/O is provided by Laurel's LTE series transmitters.
  • Dual-channel pulse inputs for voltage signals, NPN or PNP proximity switches, contact closures, magnetic pickups or flow meters.
  • Dual solid state relays, (isolated), for alarm or control. Rated 120 mA at 130 Vac or 170 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA,  or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Removable screw terminal connections of Laurel transmitters

LT series DIN rail Transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LT series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an RS232 or RS485 serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LT Transmitters to a Local Area Network (LAN)

Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for LT Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

4-20 mA Current & Serial Data Output Transmitter for Time of Single or Accumulated Events

Pulse Input
Types AC, pulses from NPN, PNP transistors, contact closures, magnetic pickups.
Signal Ground Common ground for channels A & B.
Minimum Signal Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V).
Maximum Signal 250 Vac
Maximum Frequency 1 MHz, 30 kHz, 250 Hz (selectable).
Contact Debounce 0, 3, 50 ms (selectable).
Time Base Accuracy Quartz crystal calibrated to ±2 ppm.
Span Tempco ±1 ppm/°C (typ)
Long-term Drift ±5 ppm/year
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Stopwatch Operation
Timing Modes:  
  With CH A only + to + edge, or - to - edge.
  With CH A tied to CH B + to - edge, or - to + edge.
  With CH A and CH B + edge of A to + edge of B, + edge of A to - edge of B, - edge to A to - edge of B, - edge of A to - edge of B
Timing Interval 1 µs to 999,999 hrs
Timing Resolution 0.2 µs to 1 hr
Selectable Decimal Time 999999 H, M or S format with decimal point
Selectable Clock Time HH.MM.SS format
Output Update Rate Programmable gate time from 10 ms to 199.99 s + 30 ms
Analog Output (standard)
Output Levels 4-20 mA, 0-20 mA, 0-10 Vdc, -10 to +10Vdc (user selectable)
Compliance at 20 mA 10V (0-500Ω load)
Compliance at 10V 2 mA (5 kΩ or higher load)
Output Resolution 16 bits (65,536 steps)
Output Accuracy ±0.02% of output span
Output Update Rate Programmed gate time + 30 ms + 0-2 signal periods
Output Isolation 250V rms working, 2.3 kV rms per 1 minute test
Serial Data Output (standard)
Signal Types RS232 or RS485 (half or full duplex), jumper selectable
Data Rates 300, 600, 1200, 2400, 4800, 9600, 19200 baud
Output Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Protocols Modbus RTU, Modbus ASCII, Custom ASCII
Modbus Compliance Modbus over Serial Line Specification V1.0 (2002)
RS232/485 Connector Screw terminals for easy daisy chaining
Digital Addresses 247 for Modbus, 31 for Custom ASCII
Dual Relay Output (standard)
Relay Type Two solid state relays, SPST, normally open, Form A
Load Rating 120 mA at 140 Vac or 180 Vdc
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Power Input
Standard Power 85-264 Vac or 90-300 Vdc
Low Power Option 10-48 Vdc or 12-32 Vac
Power Frequency DC or 47-63 Hz
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Power Consumption at 24V 1.5W typical, 3W with max excitation output
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Cooling Required Mount transmitters with ventilation holes at top and bottom. Leave 6 mm (1/4") between transmitters, or force air with a fan.
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options available using Laurel's free Instrument Setup Software.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

Laureate LT transmitter pinout

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

Stopwatch Mode
Stopwatch mode of Laureate digital stopwatch transmitter The stopwatch mode is used to time single events between start and stop pulses on the same channel. Duration of a single wave shape can be measured by tying the A and B channels together.
Timing Process Dynamics
Timing process dynamics by Laureate digital stopwatch transmitter The start and stop pulses used for timing can be generated by the dual relays in a Laureate panel meter, counter, or transmitter. For instance, the start and stop pulse edges can be created as temperature passes two alarm setpoints, or as temperature cycles in a hysteresis control mode.
Replacing an Oscilloscope with a Laureate Meter or Transmitter
Replacing an oscilloscope by a Laureate digital stopwatch transmitter An oscilloscope is great for viewing and timing pulses in a lab. However, in fixed installations where digital timing accuracy and control outputs are required, a low-cost Laureate time interval meter or transmitter will be the instrument of choice. Resolution to 0.2 µs is feasible.
Instrumenting a Pulsed Laser System
Instrumenting a laser system using Laureate timers and stopwatches Some of the many possibilities in instrumenting a pulsed laser system with Laureate dual-channel counters and transmitters: elapsed time, number of pulses, pulse width, pulse separation, duty cycle, and pulse rep rate.

 

 

CAL-Digital

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LT60FR
Price as Configured: $334.00

Click on the Option Board Links for More Product Information

Base Item
$164.00
Main Board
$0.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
Part Number as Configured:
LT60FR
Price as Configured:
$334.00
Quantity:
- +
Extended Price:
$334.00

Understanding the Laureate™ LT Series DIN Rail Transmitter for Time of Single or Accumulated Events

The Laureate™ LT Series DIN rail transmitter for stopwatch applications puts out isolated analog and serial data signals whose values track the time of single events (start and stop pulses), or the accumulated time of multiple events. It can also time the width of a single pulse. The highest resolution is 0.2 µs, making the transmitter ideal for fast events. The longest timing interval is 999,999 hrs. For long events, the analog output is updated continuously during timing.

A-A and A-B Stopwatch Modes

In A-A Stopwatch Mode, time is measured between a start pulse and a stop pulse, both on Channel A, from either positive or negative edges. In A-B Stopwatch Mode, time is measured between a start pulse on Channel A (positive or negative edge) and a stop pulse on Channel B (positive or negative edge), allowing inputs from different sources. Channel A and B inputs can also be tied together to start the stopwatch with one polarity and stop it with the other. Four edge combinations are supported: + edge of A to + edge of B, + edge of A to - edge of B, - edge of A to + edge of B, and - edge of A to - edge of B.

Timing Specifications

Event time (Item #1) is measured by counting 5.5 MHz clock pulses from a calibrated quartz crystal. The stopwatch output is updated during timing at a rate controlled by gate time, up to 25/sec; time resets to zero when the next start pulse occurs. Accumulated time from multiple events up to 999,999 hours (Item #2) is also tracked. Timing interval ranges 1 µs to 999,999 hrs, with resolution from 0.2 µs to 1 hr. Time base accuracy is calibrated to ±2 ppm, with span tempco of ±1 ppm/°C typical and long-term drift of ±5 ppm/year. Format is selectable as decimal (H, M, or S with decimal point) or HH.MM.SS clock time. Output update rate is programmable gate time from 10 ms to 199.99 s, plus 30 ms.

Real-World Applications

  • Stopwatch Mode — times single events between start and stop pulses on the same channel; duration of a single wave shape can be measured by tying Channels A and B together.
  • Timing Process Dynamics — start and stop pulses can be generated by the dual relays in a Laureate panel meter, counter, or transmitter, such as pulse edges created as temperature passes two alarm setpoints, or as temperature cycles in a hysteresis control mode.
  • Replacing an Oscilloscope — in fixed installations requiring digital timing accuracy and control outputs, a low-cost time interval meter or transmitter is the instrument of choice, with resolution to 0.2 µs feasible.
  • Instrumenting a Pulsed Laser System — dual-channel counters and transmitters can measure elapsed time, number of pulses, pulse width, pulse separation, duty cycle, and pulse repetition rate.

Factory-Calibrated Accuracy

All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM, enabling field replacement of signal conditioner boards without necessitating recalibration of the transmitter. Factory recalibration is recommended annually.

Where Time of Single or Accumulated Event DIN Rail Transmitters Are Used

  • Process Cycle Timing — start/stop pulse timing from relay-generated alarm setpoints.
  • Laser & Pulsed System Instrumentation — pulse width, separation, and repetition rate monitoring.
  • Fixed-Installation Oscilloscope Replacement — continuous digital timing with 4-20 mA and relay outputs.
  • Test Stand Event Timing — cycle time and dwell time measurement for automated test sequences.
  • Machine Cycle Duration Monitoring — single-event and accumulated-event time tracking.
  • Multi-Point RS485 Timing Networks — daisy-chained transmitters reporting to a central controller.
  • OEM Precision Timing Instrumentation — DIN rail integration into existing control panels.

Time of Single or Accumulated Event DIN Rail Transmitter Frequently Asked Questions

Why does tying Channel A and Channel B together specifically allow measuring the duration of a single wave shape?

Documented description specifically explains this configuration starts the stopwatch with one signal polarity and stops it with the other polarity — since a single pulse or wave shape inherently transitions from one polarity state to the other and back, tying A and B together lets the transmitter treat the rising and falling edges of that same single waveform as the start and stop triggers, directly measuring that waveform's own duration.

Why does the documented long-term drift figure (±5 ppm/year) matter separately from the ±2 ppm time base accuracy figure?

These are documented as addressing different timescales of the same underlying quartz crystal reference — the ±2 ppm figure describes the crystal's accuracy at calibration, while the separately documented ±5 ppm/year long-term drift describes how much that accuracy is expected to shift over the course of a year; both figures are relevant together for understanding total expected timing error at some point after the original factory calibration.

Does using relay-generated start/stop pulses (as in the documented "Timing Process Dynamics" application) introduce any additional timing delay beyond the transmitter's own resolution?

The page documents this application without detailing the relay's own switching time as a separate error contributor — since the transmitter's own resolution is documented down to 0.2 µs, while relay switching times are documented elsewhere in the LT Series specifications (such as the "typical response time of only 17 ms" cited for solid state relays on a related product) as being on a millisecond timescale, any relay-introduced delay would be expected to be far larger than the transmitter's own timing resolution for applications with millisecond-or-longer event durations.

Can decimal time format and HH.MM.SS clock format both be viewed simultaneously, or is one selected exclusively?

Documented specification lists these as "Selectable Decimal Time" and "Selectable Clock Time" as two distinct format options — the documented phrasing "selectable" for each is consistent with the display format being a configuration choice between the two, rather than both formats being simultaneously available on the display at once.

Does the documented "no error contributed" framing apply to timing accuracy the same way it's used for quadrature position accuracy on a related LT Series transmitter?

This specific stopwatch/timer page does not use that same "no error contributed" phrasing — its documented accuracy is instead expressed through the specific ±2 ppm time base accuracy, ±1 ppm/°C span tempco, and ±5 ppm/year drift figures; these should be treated as this product's own specific documented accuracy figures rather than assuming the zero-contributed-error framing used on a different LT Series product applies here as well.

Does measuring pulse width (as mentioned in the documented laser system application) use A-A mode, A-B mode, or the tied-together A/B configuration?

The page documents pulse width as one of several measurable parameters in the laser system application without specifying which exact stopwatch mode configuration is used for that particular measurement — based on the documented general description elsewhere on the page (that duration of a single wave shape is measured by tying A and B together), the tied-together configuration is the one specifically documented as suited to measuring a single pulse's own width.

Does the documented minimum signal range table (nine ranges from ±12 mV to +1.25/+2.1V) apply the same way to both the start pulse and stop pulse inputs?

Documented specification lists these nine minimum signal ranges under the general Pulse Input section covering the dual-channel signal conditioner as a whole, without separately distinguishing a start-pulse range from a stop-pulse range — since Channels A and B share the same documented signal conditioner hardware and specifications, the same set of nine selectable minimum signal ranges is consistent with applying to whichever channel is functioning as the start or stop input in a given configuration.

Why does contact debounce (0, 3, or 50 ms selectable) matter specifically for stopwatch timing, given the transmitter's much finer 0.2 µs resolution?

Documented specification lists contact debounce as a separate, coarser-timescale filtering option specifically intended for mechanical contact closures, which can produce brief electrical "bounce" (rapid multiple transitions) when physically opening or closing — selecting an appropriate debounce setting is consistent with preventing that bounce from being misread as multiple separate start or stop events, which is a genuinely different concern from the transmitter's own much finer underlying timing resolution used once a clean, debounced edge is detected.

Does the documented 25/sec maximum stopwatch output update rate limit how short an event this transmitter can actually time?

No — the 25/sec figure describes how often the display and output are documented as refreshing during an ongoing timing measurement, distinct from the separately documented minimum timing interval of 1 µs and resolution down to 0.2 µs, which describe the transmitter's ability to measure a single short event's duration; a very short single event is still captured and reported at its own precise duration, even though continuous updates during a longer event are documented as limited to 25 times per second.

Can the four documented A-B edge combinations (+/+, +/-, -/+, -/-) all be used interchangeably for the same physical measurement, or does the choice matter?

The choice matters and depends on the actual signal characteristics of the specific start and stop sources being used — since Channel A and Channel B can originate from genuinely different sensor types with different documented polarity conventions, selecting the edge combination that matches the actual rising or falling transition each specific sensor produces for the "start" and "stop" events is necessary for the transmitter to trigger at the intended physical moment, rather than the four combinations being freely interchangeable for a given real-world setup.

Q-Switched Laser Pulse Timing Questions From the Field

What is a Q-switch, and why does it specifically enable short, high-intensity laser pulses rather than continuous output?

Documented explanation specifically describes Q-switching as a technique where a Q-switch element is held in an "off" or low-Q state while the laser gain medium continues to be pumped, storing energy without emitting a beam; when the Q-switch is then rapidly switched to allow emission, the accumulated stored energy is documented as being released in a short, intense pulse rather than continuously, which is the underlying mechanism behind Q-switched pulse generation.

Is there a documented typical range for Q-switched pulse repetition rate and pulse width in real systems?

Yes — documented industrial and research examples specifically cite pulse repetition rates ranging from single-shot up to tens of MHz depending on the system, with one documented custom photoacoustic laser specifically citing a 10 Hz to 10 kHz repetition rate control range and a 40-150 ns pulse width control range; another documented example cites an 8 kHz repetition rate with a 50 microsecond designated pulse width for a laser marking application.

Do Q-switched laser pulses always consist of one clean single pulse, or can secondary pulses appear within a single Q-switch emission period?

Not always a single clean pulse — documented technical description specifically identifies "relaxation oscillation" as a phenomenon that can produce a primary emission pulse along with multiple secondary emission pulses of substantially lower intensity occurring over the same emission period, particularly documented as arising when a longer emission period (such as the cited 50 microsecond example) is set on certain Q-switched laser configurations.

Does increasing a Q-switched laser's pulse repetition rate always increase its average output power proportionally?

No — documented experimental results specifically show output power increasing with repetition rate only up to a point, then leveling off; one documented example cites output power becoming comparable between 150 kHz and 200 kHz repetition rates, attributed specifically to switching losses saturating the achievable output power at higher repetition rates, rather than power continuing to scale proportionally with rate indefinitely.

Does pulse width in a Q-switched laser typically stay constant regardless of pump power, or does it vary?

Documented experimental data specifically shows pulse width varying with both pump power and repetition rate in real systems — one documented example cites pulse width ranging from about 19.3 ns down to 5.1 ns across different absorbed pump power and repetition rate combinations, with documented analysis noting no simple, direct relationship between pulse width and pump power alone at a fixed repetition rate.

Does pulse energy increase or decrease as Q-switched repetition rate increases, for a fixed pump power?

Documented experimental data specifically shows pulse energy decreasing as repetition rate increases — one documented example cites maximum pulse energy of 194 µJ at a 20 kHz repetition rate dropping to 22 µJ at 200 kHz repetition rate under comparable pump conditions, reflecting that the same total available pump energy is documented as being divided among a larger number of pulses per second at higher repetition rates.

Is pulse-to-pulse stability (variation in pulse strength) a documented practical concern in real Q-switched laser systems?

Yes, though documented results suggest it can be well-controlled — one documented high-repetition-rate Q-switched laser example specifically reports pulse strength stable to within about 6.2% variation across its pulse train, without evidence of piezoelectric ringing effects, illustrating that while pulse-to-pulse variation is a genuine documented characteristic worth measuring, well-engineered systems can achieve relatively tight, quantified stability.

Does the choice of pulse width setting in a Q-switched marking laser affect the practical outcome on the material being processed?

Yes — documented application description specifically notes that using a particular pulse laser beam configuration (with its associated relaxation-oscillation pulse structure at a given designated pulse width) allows relatively shallow marking lines to be produced on a material, directly tying the documented pulse timing characteristics to a genuine, practical processing outcome rather than being a purely academic timing parameter.