LT DIN Rail Digital Transmitters with Serial Data Communication and Analog Outputs for Time of Periodic Event Applications

LT DIN Rail Digital Transmitters with Serial Data Communication and Analog Outputs for Time of Periodic Event Applications

Price: $334.00
  • P/NLT60FR
- +

Features

  • Times periodic events with width from 1 µs to 199.999 s
  • Transmits average time of periodic events with width from 1 µs to 199.999 s
  •  Resolution to 0.2 µs, rep rated to 250 kHz
  • Inputs from NPN or PNP proximity switches, contact closures, digital logic, or magnetic pickups down to 12 mV
  • Trigger on positive or negative pulse edges
  • 4-20 mA, 0-20 mA, 0-10V or -10V to +10V transmitter output, (isolated)
  • Analog output resolution 0.0015%  of span, accuracy ±0.02%  of span
  • RS232 or RS485 serial data, Modbus or Laurel ASCII protocol (isolated)
  • Dual 120 mA solid state relays for alarm or control (isolated)
  • 5V, 10V, 12V, or 24V dc transducer excitation output (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)

The Laureate™ LT Series DIN rail analog transmitter with serial data communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy with Input frequencies from 0.005 Hz to 1 MHz. The LT Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

Time Interval Mode for Time Delay

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and RS232/RS485 output transmitter for time interval can transmit pulse width or time delay between individual pulses to a resolution of 0.2 µs for periodic events. It can also transmit average pulse width or average time delay between multiple pulses.

Time interval is measured between inputs on channels A and B. Timing starts when a pulse is applied to Channel A (selectable positive or negative edge), and ends when a pulse is applied to Channel B (selectable positive or negative edge). In case of a single pulsed signal, the A and B inputs can be tied together. A positive or negative slope may be selected to start timing, and the opposite slope must be selected to stop timing. Timing is achieved by counting 5.5 MHz clock pulses. Multiple integral time intervals are averaged over a gate time which is selectable from 10 ms to 199.99 s and also controls the maximum output rate.

The dual-channel signal conditioner used for pulse detection accepts inputs from proximity switches with PNP or NPN output, TTL or CMOS logic, magnetic pickups, contact closures, and other signals from 12 mV to 250 Vac. Jumper selections provide optimum operation for different sensor types and noise conditions. A built-in 5V, 10V, 12V, or 24V dc excitation supply can power proximity switches and other sensors, and eliminate the need for an external power supply.

Exceptional Accuracy and Stability. Laureate transmitters determine frequency by taking the inverse of period as measured with a calibrated quartz crystal time base. This results in extremely accurate and stable 6-digit internal readings (±999,999 counts), which are then processed in software. The analog output is generated by an ultra-linear 16-bit (65,536 step) digital-to-analog converter (DAC) for 0.02% output accuracy. The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

Standard Features of Laureate LT Transmitters Include:

  • Serial communications output, (isolated), RS232 or RS485 (half or full duplex), jumper selectable. Three protocols are user selectable: Modbus RTU, Modbus ASCII, or Laurel ASCII. Modbus operation is fully compliant with Modbus Over Serial Line Specification V1.0 (2002). The Laurel ASCII protocol is simpler than the Modbus protocol and is recommended when all devices are Laureates.
  • 4-20 mA, 0-10V or -10V to +10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 0.0015% resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. Note that Ethernet data I/O is provided by Laurel's LTE series transmitters.
  • Dual-channel pulse inputs for voltage signals, NPN or PNP proximity switches, contact closures, magnetic pickups or flow meters.
  • Dual solid state relays, (isolated), for alarm or control. Rated 120 mA at 130 Vac or 170 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA,  or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Removable screw terminal connections of Laurel transmitters

LT series DIN rail Transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LT series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an RS232 or RS485 serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LT Transmitters to a Local Area Network (LAN)

Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for LT Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

4-20 mA & Serial Data Output Transmitter for Time of Periodic Events

Pulse Input
Types AC, pulses from NPN, PNP transistors, contact closures, magnetic pickups
Signal Ground Common ground for channels A & B.
Minimum Signal Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V).
Maximum Signal 250 Vac
Maximum Frequency 1 MHz, 30 kHz, 250 Hz (selectable).
Contact Debounce 0, 3, 50 ms (selectable).
Time Base Accuracy Quartz crystal calibrated to ±2 ppm.
Span Tempco ±1 ppm/°C (typ)
Long-term Drift ±5 ppm/year
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Time Interval Mode
Timing Start Channel A pulse, + or - edges
Timing Stop Channel B pulse, + or - edges
Update Rate Gate time + 30 ms + 0-2 time intervals
Gate Time Selectable 10 ms to 199.99 s
Time to Zero Output Selectable 10 ms to 199.99 s
Resolution
0 - 199.999 s 1 ms
0 - 99.9999 s 100 µs
0 - 9.99999 s 10 µs
0 - .999999 s 1 µs
0 - .099999 s 0.2 µs (after averaging)
Analog Output (standard)
Output Levels 4-20 mA, 0-20 mA, 0-10 Vdc, -10 to +10Vdc (user selectable)
Compliance at 20 mA 10V (0-500Ω load)
Compliance at 10V 2 mA (5 kΩ or higher load)
Output Resolution 16 bits (65,536 steps)
Output Accuracy ±0.02% of output span
Output Update Rate Programmed gate time + 30 ms + 0-2 signal periods
Output Isolation 250V rms working, 2.3 kV rms per 1 minute test
Serial Data Output (standard)
Signal Types RS232 or RS485 (half or full duplex), jumper selectable
Data Rates 300, 600, 1200, 2400, 4800, 9600, 19200 baud
Output Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Protocols Modbus RTU, Modbus ASCII, Custom ASCII
Modbus Compliance Modbus over Serial Line Specification V1.0 (2002)
RS232/485 Connector Screw terminals for easy daisy chaining
Digital Addresses 247 for Modbus, 31 for Custom ASCII
Dual Relay Output (standard)
Relay Type Two solid state relays, SPST, normally open, Form A
Load Rating 120 mA at 140 Vac or 180 Vdc
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Power Input
Standard Power 85-264 Vac or 90-300 Vdc
Low Power Option 10-48 Vdc or 12-32 Vac
Power Frequency DC or 47-63 Hz
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Power Consumption at 24V 1.5W typical, 3W with max excitation output
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Cooling Required Mount transmitters with ventilation holes at top and bottom. Leave 6 mm (1/4") between transmitters, or force air with a fan.
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options available using Laurel's free Instrument Setup Software.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

Laureate LT transmitter pinout

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

Applications of Time Interval Meters & Transmitters

Time Interval Mode for Time Delay
Time Interval Mode for Time Delay For periodic pulses applied to A and B channels, time delays can be measured down to 0.2 µs resolution from the rising or falling edge of A to the rising or falling edge of B (selectable).
Time Interval Mode for Pulse Width
Time Interval Mode for Pulse Width The width of periodic pulses (t1 or t2) can be measured by tying the A and B channels together. As for time delay, readings are averaged over a user-selectable gate time.
Timing Process Dynamics with two Meters or Transmitters
Timing Process Dynamics with a Panel Meter and Time Interval Meter The start and stop pulses used for timing can be generated by the dual relay board in a Laureate panel meter, counter or transmitter. For instance, the start and stop pulse edges can be created as temperature passes two alarm setpoints, or temperature cycles in a hysteresis control mode.
Replacing an Oscilloscope with a Laureate Time Interval Meter or Transmitter
Replacing an Oscilloscope with a Laureate Time Interval Meter An oscilloscope is great for viewing and timing pulses in a lab. However, in fixed installations where digital timing accuracy and control outputs are required, a low-cost Laureate time interval meter or transmitter will be the instrument of choice. Resolution to 0.2 µs is feasible.
Instrumenting a Pulsed Laser System
Instrumenting a laser system using Laureate timers and stopwatches Some of the many possibilities in instrumenting a pulsed laser system with Laureate dual-channel counters and transmitters: elapsed time, number of pulses, pulse width, pulse separation, duty cycle, and pulse rep rate.

 

 

CAL-Digital

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LT60FR
Price as Configured: $334.00

Click on the Option Board Links for More Product Information

Base Item
$164.00
Main Board
$0.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
Part Number as Configured:
LT60FR
Price as Configured:
$334.00
Quantity:
- +
Extended Price:
$334.00

Understanding the Laureate™ LT Series DIN Rail Transmitter for Time of Periodic Events

The Laureate™ LT Series DIN rail transmitter for time interval can transmit pulse width or time delay between individual pulses to a resolution of 0.2 µs for periodic events. It can also transmit average pulse width or average time delay between multiple pulses. Time interval is measured between inputs on Channels A and B: timing starts when a pulse is applied to Channel A (selectable positive or negative edge) and ends when a pulse is applied to Channel B (selectable positive or negative edge). For a single pulsed signal, the A and B inputs can be tied together, with a positive or negative slope selected to start timing and the opposite slope selected to stop timing.

Timing Mechanism and Resolution

Timing is achieved by counting 5.5 MHz clock pulses. Multiple integral time intervals are averaged over a gate time selectable from 10 ms to 199.99 s, which also controls the maximum output rate. Resolution varies by range: 1 ms for 0-199.999 s, 100 µs for 0-99.9999 s, 10 µs for 0-9.99999 s, 1 µs for 0-.999999 s, and 0.2 µs (after averaging) for 0-.099999 s. Time base accuracy is calibrated to ±2 ppm, with span tempco of ±1 ppm/°C typical and long-term drift of ±5 ppm/year. Update rate is gate time plus 30 ms plus 0-2 time intervals; Time to Zero Output is separately selectable from 10 ms to 199.99 s.

Real-World Applications

  • Time Interval Mode for Time Delay — for periodic pulses applied to A and B channels, time delays are measured down to 0.2 µs resolution from the rising or falling edge of A to the rising or falling edge of B (selectable).
  • Time Interval Mode for Pulse Width — the width of periodic pulses is measured by tying the A and B channels together, with readings averaged over a user-selectable gate time.
  • Timing Process Dynamics with Two Meters or Transmitters — start and stop pulses can be generated by the dual relay board in a Laureate panel meter, counter, or transmitter, such as pulse edges created as temperature passes two alarm setpoints, or temperature cycles in a hysteresis control mode.
  • Replacing an Oscilloscope — in fixed installations requiring digital timing accuracy and control outputs, a low-cost time interval meter or transmitter is the instrument of choice, with resolution to 0.2 µs feasible.
  • Instrumenting a Pulsed Laser System — dual-channel counters and transmitters can measure elapsed time, number of pulses, pulse width, pulse separation, duty cycle, and pulse repetition rate.

Factory-Calibrated Accuracy

All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM, enabling field replacement of signal conditioner boards without necessitating recalibration of the transmitter. Factory recalibration is recommended annually.

Where Time of Periodic Event DIN Rail Transmitters Are Used

  • Fixed-Installation Pulse Timing — oscilloscope-replacement digital timing with 4-20 mA and relay outputs.
  • Process Cycle Delay Monitoring — relay-generated start/stop pulse timing from alarm setpoints.
  • Laser & Pulsed System Instrumentation — pulse width, separation, and repetition rate averaging.
  • Rotating Machinery Event Timing — averaged period measurement for periodic mechanical events.
  • PWM & Control Signal Verification — averaged pulse width and delay monitoring in noisy environments.
  • Multi-Point RS485 Timing Networks — daisy-chained transmitters reporting to a central controller.
  • OEM Precision Timing Instrumentation — DIN rail integration into existing control panels.

Time of Periodic Events DIN Rail Transmitter Frequently Asked Questions

Why does resolution improve from 1 ms at the widest range down to 0.2 µs at the narrowest range, rather than staying constant across all ranges?

Documented resolution table specifically ties finer resolution to narrower measurement ranges — since the transmitter's internal count of 5.5 MHz clock pulses produces a fixed number of possible count values, spreading that same count resolution over a much wider maximum time range (199.999 s) necessarily yields coarser resolution than concentrating it over a much narrower range (0.099999 s), which is why the documented resolution figures scale together with the selected range.

Why does the finest documented resolution of 0.2 µs specifically require averaging, unlike the coarser resolution tiers?

Documented specification specifically notes "0.2 µs (after averaging)" only for the narrowest range, distinct from the other resolution tiers listed without that qualifier — this indicates that achieving the very finest resolution specifically depends on the documented multiple-integral-time-interval averaging process over the selected gate time, rather than being available from a single, unaveraged measurement the way the coarser resolution tiers are documented as being.

Does "Time to Zero Output" on this periodic-events transmitter serve the same function as "Time Before Zero Output" documented on other LT Series transmitters?

Both are documented with the same 10 ms to 199.99 s selectable range and appear to describe a related concept — the specific timeout duration after which the transmitter's output defaults to zero in the absence of valid signal — though this page's exact terminology is "Time to Zero Output" rather than "Time Before Zero Output"; the underlying function is consistent with the same category of timeout-to-zero behavior documented elsewhere in the LT Series.

Does tying Channels A and B together for pulse width measurement affect the transmitter's documented gate-time averaging capability?

No — documented description specifically states that for pulse width measurement (A and B tied together), "readings are averaged over a user-selectable gate time," the same underlying averaging mechanism documented for the separate time-delay measurement mode; tying the channels together changes what physical quantity is being measured, not whether gate-time averaging remains available.

Can the same physical transmitter be reconfigured between measuring time delay (A to B) and pulse width (A tied to B), or does this require different hardware?

Documented description presents both as configuration modes of the same Time Interval Mode transmitter, distinguished by whether Channels A and B are wired to separate signal sources or tied together — this is consistent with a wiring and setup configuration choice on the same physical hardware, rather than requiring separate transmitter models for time delay versus pulse width measurement.

Does averaging multiple time intervals over a longer gate time reduce genuine measurement noise, or does it also risk masking real variation between individual events?

Documented description specifically frames averaging as improving resolution (particularly enabling the finest 0.2 µs tier), which is consistent with reducing the effect of random measurement noise on the reported value — however, since the reported reading represents an average across the gate time's multiple intervals rather than any single interval, genuine event-to-event variation would similarly be smoothed into that average rather than reported individually, a general characteristic of any averaging-based measurement approach.

Does the documented ±1 ppm/°C span tempco figure apply to the time-delay reading, the pulse-width reading, or both?

Documented specification lists span tempco once under the general Pulse Input section, applying to the transmitter's underlying timing measurement as a whole, rather than listing separate tempco figures for time-delay mode versus pulse-width mode — since both modes share the same documented 5.5 MHz clock-counting mechanism, the same ±1 ppm/°C span tempco figure is consistent with applying regardless of which specific timing mode is configured.

Why does the "Instrumenting a Pulsed Laser System" application list duty cycle as one of the measurable parameters, when this page's core function is time interval rather than duty cycle measurement?

Documented description specifically lists duty cycle alongside elapsed time, pulse count, pulse width, pulse separation, and pulse repetition rate as parameters obtainable from Laureate dual-channel counters and transmitters in this application — since duty cycle is mathematically derivable from pulse width and pulse separation (both of which this time interval transmitter directly measures), documented duty cycle figures in this application context are consistent with being calculated from those two directly measured time interval quantities rather than measured as an independent parameter.

Does selecting a shorter gate time always provide a faster-updating reading, even at the cost of averaging fewer intervals?

Yes — documented specification ties output update rate directly to gate time (gate time + 30 ms + 0-2 time intervals), meaning a shorter selected gate time is consistent with a faster update rate; the documented tradeoff is that a shorter gate time also averages fewer individual time intervals together, which is the same general averaging-versus-responsiveness tradeoff documented elsewhere for gate time selection on related LT Series transmitters.

Can Time Interval Mode on this transmitter measure a time delay between two entirely unrelated periodic signals, or must A and B originate from the same underlying event source?

Documented description specifically states the mode allows "inputs from different sources" for the A-to-B time delay measurement, without requiring A and B to originate from the same underlying physical event — this is consistent with using this transmitter to measure the timing relationship between two genuinely independent periodic signals, provided both are periodic and their relative timing relationship is what's meaningful to the specific application.

Ignition Timing & Spark Pulse Interval Questions From the Field

Why is the time period between successive ignition sparks on one cylinder so much longer than the spark event itself?

Documented example specifically illustrates this with real figures — a spark event lasting around 2 ms compared to a documented 150 ms period between sparks on a single cylinder at 800 RPM in a 4-cylinder engine; the spark itself is documented as accounting for only around 2% of the actual work cycle, meaning the vast majority of the periodic interval between ignition events is not spark activity at all.

Does simply raising engine RPM during diagnostic testing change the fundamental interval being measured between spark events?

Yes — documented practice specifically describes technicians deliberately increasing engine RPM specifically to shorten the ignition cycle period, which is documented as "saturating" the waveform for better visibility on diagnostic equipment; this is a deliberate technique that directly exploits the known relationship between RPM and the inter-spark timing interval to make brief, otherwise hard-to-observe events more visible.

Why do different vehicle manufacturers require different correction factors when measuring ignition timing from a crankshaft position sensor?

Documented explanation specifically attributes this to manufacturers placing their crankshaft position reference notch (or equivalent pulse-generating feature) at different physical points relative to top dead center — this documented variation, driven partly by physical placement constraints and partly by differing engineering preferences, means a timing measurement device working across multiple engine types must document and apply a specific correction factor for each manufacturer's particular reference point placement.

Is there a documented reason ignition systems sometimes generate multiple pulses per single triggering event at low RPM, and how is this handled in timing measurement?

Yes — documented description specifically identifies that certain ignition systems generate several ignition pulses combined into a single group at low RPM to trigger one ignition process, and describes a specific retriggerable timer technique designed to suppress the extra pulses that follow the first pulse in such a group, ensuring the timing/RPM measurement circuit responds to the intended single triggering event rather than miscounting the group as multiple separate events.

Is there a documented typical accuracy figure achievable for dedicated automotive ignition timing measurement instruments?

Yes — one documented commercial ignition timing instrument specifically cites an accuracy figure of ±0.05 degrees for both steady-state and transient testing of ignition, camshaft, or injector timing, illustrating the level of precision documented as achievable and expected in dedicated automotive timing measurement equipment.

Does the correct diagnostic approach for verifying spark timing rely on absolute pulse timing alone, or does it also require comparing against a reference waveform?

Documented diagnostic guidance specifically recommends comparing a measured waveform (such as from crankshaft or camshaft position sensors) against a known-good reference waveform, typically sourced from a service manual, rather than relying solely on absolute timing values in isolation; documented practice also describes verifying that a spark event aligns correctly relative to a separately measured compression stroke signal as a specific diagnostic check for timing correctness.

Can an engine's ignition system deliberately generate pulses that don't correspond to spark plug firing, for diagnostic testing purposes?

Yes — documented equipment description specifically explains that certain diagnostic interfaces can generate a modified timing signal carrying "dwell information of a dwell period insufficient to fire a spark plug," specifically to inhibit a selected cylinder from firing during a power-balance test, while still producing a genuine, documented timing signal to the ignition module for diagnostic monitoring purposes.