LT DIN Rail Analog Transmitters with Serial Data Communication and Analog outputs for Resistance in Ohms

LT DIN Rail Analog Transmitters with Serial Data Communication and Analog outputs for Resistance in Ohms

Price: $334.00
  • P/NLT20R1
- +

Features

  • Five jumper-selectable resistance ranges of 20.000 ohm to 200.00 kohm
  • Three fixed factory-special resistance ranges of 2.0000 ohm, 2.0000 Mohm, 20.000 Mohm
  • Accuracy at 25°C ±0.01%  of reading ± 2 counts
  • 0.1 milliohm resolution on 2 ohm scale for contact resistance measurements
  • 2, 3 or 4-wire connection with lead resistance compensation
  • All input ranges are user selectable and factory calibrated
  • Up to 60 conversions per second, Ideal for peak or valley capture
  • 4-20 mA, 0-20 mA, 0-10V or -10V to +10V transmitter output, (isolated)
  • Analog output resolution 0.0015%  of span, accuracy ±0.02%  of span
  • RS232 or RS485 serial data, Modbus or Laurel ASCII protocol (isolated)
  • Dual 120 mA solid state relays for alarm or control (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)
    Optional - Extended allows up to 180 data points for custom curve linearization and a rate derived from consecutive readings.

The Laureate™ LT Series DIN rail analog transmitter with serial data communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy of 0.01% of reading ± 2 counts, with high read rates at up to 60 or 50 conversions per second. The LT Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.


The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and RS232/RS485 output transmitter for  resistance in ohms offers the same high performance, signal conditioning and programmable features as Laureate digital panel meters, counters & timers provides six voltage input ranges and four current input ranges, all factory calibrated and jumper selectable. The Resistance in Ohm transmitter is ideal for high-speed, high-accuracy resistance measurements in a production environment, such as contact resistance measurements. It is factory calibrated for five jumper selectable resistance ranges from 20.000 ohm to 200.00 kohm. Factory-special, fixed ranges of 2.0000 ohm, 2.0000 Mohm and 20.000 Mohm are also available. Accuracy is an exceptional ±0.01% of reading ± 2 counts. Resolution is one part in 20,000. In the 2 ohm range, resolution is 0.1 milliohm for contact resistance measurements.

Transmitter connections can be via 2, 3 or 4 wires. With 4-wire hookup, 2 wires are used for excitation and two separate wires are used to sense the voltage across the resistance to be measured, thereby eliminating any lead resistance effects. With 3-wire hookup, the transmitter senses the combined voltage drop across the RTD plus two excitation leads. It also senses the voltage drop across one excitation lead, and then subtracts twice this voltage from the combined total. This technique effectively subtracts the lead resistance if the excitation leads are the same.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

The optional extended Laureate computer board enhances Laureate transmitter by displaying rates derived from successive readings and enabling highly accurate custom curve linearization. For example, it can calculate liquid volume or flow rate in a horizontal cylindrical tank using levels from a 4-20 mA transmitter. Setup is straightforward: users input up to 180 data points into a spreadsheet or text file, and the computer calculates spline-fit segments, which are then downloaded to the transmitter for precise operation.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

High read rate of up to 50 or 60 conversions per second, the Laureate™ LT Series transmitter uses Concurrent Slope (US Pat. 5,262,780) analog-to-digital conversion to integrate signals over a full power line cycle (50 Hz or 60 Hz). This read rate enables peak and valley capture, real-time computer interfacing, and control applications. Peak and valley values are automatically captured and can be viewed using Laurel’s free Instrument Setup Software (compatible with Windows PCs) or transmitted as serial data.

Standard Hardware Features of Laureate LT Transmitters Include:

  • Serial communications output, (isolated), RS232 or RS485 (half or full duplex), jumper selectable. Three protocols are user selectable: Modbus RTU, Modbus ASCII, or Laurel ASCII. Modbus operation is fully compliant with Modbus Over Serial Line Specification V1.0 (2002). The Laurel ASCII protocol is simpler than the Modbus protocol and is recommended when all devices are Laureates.
  • 4-20 mA, 0-10V or -10V to +10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 0.0015% resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. Note that Ethernet data I/O is provided by Laurel's LTE series transmitters.
  • Dual solid state relays, (isolated), for alarm or control. Rated 120 mA at 130 Vac or 180 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100mA, or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Two tare functions: auto-tare and manual tare. In auto-tare, an input line is grounded by an external pushbutton. This causes the current weight, which is normally the empty weight of the container to be stored in memory as an offset. In manual tare, the tare value can be entered manually via a control input pushbutton or using Laurel's free Instrument Setup Software.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

Removable screw terminal connections of Laurel transmitters

LT series DIN rail Transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LT series Transmitters feature isolated, user-selectable analog outputs (4-20 mA, 0-20 mA, 0-10V, or -10V to +10V), an RS232 or RS485 serial data interface, and dual 120 mA solid state AC/DC relays. Most models, except those with temperature or AC RMS signal conditioners, include an isolated 5, 10, 12, or 24 Vdc transducer excitation output.

Connecting Laureate LT Transmitters to a Local Area Network (LAN)

Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for LT Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

Ethernet & 4-20 mA Output Thermocouple Temperature Transmitter

Range Ohms Resolution Accuracy Excitation Current ***
R0** 0-2.0000 Ω 0.1 mΩ ±0.01% of reading
± 2 counts
5 mA
R1* 0-20.000 Ω 1 mΩ 5 mA
R2* 0-200.00 Ω 10 mΩ 500 µA
R3* 0-2000.0 Ω 100 mΩ 50 µA
R4* 0-20000 Ω 1 Ω 5 µA
R5* 0-200.00 kΩ 10 Ω 500 nA
R6** 0-2.0000 MΩ 100 Ω 500 nA
R7** 0-20.0000 MΩ 1000 Ω 80 nA
*    Jumper-selectable, precalibrated range.
**   Factory-set fixed range.
***The applied excitation current is sensed by the meter, which operates in a ratiometric mode and automatically compensates for any changes in excitation.
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Signal Input
Input Resolution 16 bits (65,536 steps)
Input Accuracy ±0.01% of reading ± 2 counts
Update Rate, Max 50/sec at 50 Hz, 60/sec at 60 Hz
Analog Output (standard)
Output Levels 4-20 mA, 0-20 mA, 0-10 Vdc, -10 to +10Vdc (user selectable)
Compliance, 4-20 mA 10V (0-500Ω load)
Compliance, 0-10V 2 mA (5 kΩ load or higher)
Output Resolution 16 bits (65,536 steps)
Output Accuracy 0.02% of output span plus conversion accuracy
Output Isolation 250V rms working, 2.3 kV rms per 1 minute test
Serial Data Output (standard)
Signal Types RS232 or RS485 (half or full duplex), jumper selectable
Data Rates 300, 600, 1200, 2400, 4800, 9600, 19200 baud
Output Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Protocols Modbus RTU, Modbus ASCII, Custom ASCII
Modbus Compliance Modbus over Serial Line Specification V1.0 (2002)
RS232/485 Connector  Screw terminals for easy daisy chaining 
Digital Addresses 247 for Modbus, 31 for Custom ASCII
Dual Relay Output (standard)
Relay Type Two solid state relays, SPST, normally open, Form A
Load Rating 120 mA at 140 Vac or 180 Vdc
Power Input
Standard Power 85-264 Vac or 90-300 Vdc
Low Power Option 10-48 Vdc or 12-32 Vac
Power Frequency DC or 47-63 Hz
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Power Consumption at 24V 1.5W typical
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Cooling Required Mount transmitters with ventilation holes at top and bottom. Leave 6 mm (1/4") between transmitters, or force air with a fan.
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options available using Laurel's free Instrument Setup Software.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

Laureate LT transmitter pinout

Resistance Measurement with Excitation & Lead Compensation

Ohm transmitter hookup can be via 2, 3 or 4 wires to the J5 connector. The transmitter applies a fixed excitation current for each resistance range.
4-wire connection to resistance signal condidioner In 4-wire hookup, different pairs of leads are used to apply the excitation current and sense the voltage drop across the unknown resistance, so that the IR drop across the excitation leads is not a factor.
3-wire connection to resistance signal condidioner
In 3-wire hookup, the transmitter senses the combined voltage drop across the unknown resistance plus two excitation leads. It also senses the voltage drop across one excitation lead, and then subtracts twice this voltage from the combined total. This technique effectively subtracts all lead resistance and compensates for ambient temperature changes if the two excitation leads are identical.
2-wire connection to resistance signal condidioner In 2-wire hookup, the transmitter senses the combined voltage drop across the unknown resistance and both lead wires. The voltage drop across the lead wires can be measured by shorting out the resistance during transmitter setup, and this voltage is then automatically subtracted from the combined total. However, changing resistance of the lead wires due to ambient temperature changes will not be compensated.

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

QA Application with Relays in Passband Mode

Deviation mode relay operation for QA testing of resistance A deviation limit (50 mΩ in this example) is set up around both sides of a setpoint. The relay closes (or opens) when the reading falls within the deviation band, and opens (or closes) when the reading falls outside of this band. This mode sets up a passband around the setpoint and can be used for contact resistance testing in a production environment.

 

 

CAL-Analog

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LT20R1
Price as Configured: $334.00

Click on the Option Board Links for More Product Information

Base Item
$164.00
Main Board
$0.00
$33.00
Power (Isolated)
$89.00
Extended allows up to 180 data points for custom curve linearization and a rate derived from consecutive readings.
$89.00
Signal Input (Isolated)
$115.00
$81.00
$81.00
$81.00
$81.00
$81.00
$115.00
Note: All ranges are factory calibrated and user selectable
$115.00
Part Number as Configured:
LT20R1
Price as Configured:
$334.00
Quantity:
- +
Extended Price:
$334.00

Understanding the Laureate™ LT Series DIN Rail Transmitter for Resistance in Ohms

The Laureate™ LT Series DIN rail transmitter for resistance in ohms is ideal for high-speed, high-accuracy resistance measurements in a production environment, such as contact resistance measurements. It is factory calibrated for five jumper-selectable resistance ranges from 20.000 ohm to 200.00 kohm. Factory-special, fixed ranges of 2.0000 ohm, 2.0000 Mohm, and 20.000 Mohm are also available. Accuracy is ±0.01% of reading ±2 counts. Resolution is one part in 20,000; on the 2 ohm range, resolution is 0.1 milliohm for contact resistance measurements.

Resistance Range Table

R0 (factory-fixed): 0-2.0000Ω, 0.1 mΩ resolution, 5 mA excitation. R1 (jumper-selectable): 0-20.000Ω, 1 mΩ resolution, 5 mA excitation. R2: 0-200.00Ω, 10 mΩ resolution, 500 µA excitation. R3: 0-2000.0Ω, 100 mΩ resolution, 50 µA excitation. R4: 0-20000Ω, 1Ω resolution, 5 µA excitation. R5: 0-200.00kΩ, 10Ω resolution, 500 nA excitation. R6 (factory-fixed): 0-2.0000MΩ, 100Ω resolution, 500 nA excitation. R7 (factory-fixed): 0-20.0000MΩ, 1000Ω resolution, 80 nA excitation. All ranges share ±0.01% of reading ±2 counts accuracy. The applied excitation current is sensed by the transmitter, which operates in ratiometric mode and automatically compensates for any changes in excitation.

2, 3, and 4-Wire Resistance Hookup

In 4-wire hookup, different pairs of leads apply the excitation current and sense the voltage drop across the unknown resistance, so the IR drop across the excitation leads isn't a factor. In 3-wire hookup, the transmitter senses the combined voltage drop across the unknown resistance plus two excitation leads, and separately senses the drop across one excitation lead, then subtracts twice this voltage from the combined total — this subtracts all lead resistance and compensates for ambient temperature changes if the two excitation leads are identical. In 2-wire hookup, the transmitter senses the combined voltage drop across the unknown resistance and both lead wires; lead-wire voltage drop can be measured by shorting out the resistance during setup and automatically subtracted, but changing lead-wire resistance from ambient temperature changes isn't compensated.

QA Application With Relays in Passband Mode

A deviation limit (for example, 50 mΩ) is set up around both sides of a setpoint. The relay closes (or opens) when the reading falls within the deviation band, and opens (or closes) when the reading falls outside of this band. This mode sets up a passband around the setpoint and can be used for contact resistance testing in a production environment.

Factory-Calibrated Accuracy

All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM, enabling field replacement of signal conditioner boards without necessitating recalibration of the transmitter. The optional Extended computer board adds rate derived from consecutive readings and custom curve linearization using up to 180 user-entered data points. Factory recalibration is recommended annually.

Where Resistance-in-Ohms DIN Rail Transmitters Are Used

  • Production Contact Resistance Testing — QA passband relay mode for pass/fail sorting.
  • Insulation & High-Resistance Monitoring — nanoamp-level excitation for megohm-range measurement.
  • Precision Reference Resistor Verification — milliohm-resolution measurement on the 2Ω range.
  • Sensor Resistance Retransmission — 4-20 mA or serial output for resistive process sensors.
  • Long-Lead-Length Resistance Monitoring — 3- or 4-wire lead compensation for remote sensors.
  • Multi-Point RS485 Resistance Networks — daisy-chained transmitters reporting to a central controller.
  • OEM Precision Resistance Instrumentation — DIN rail integration into existing control panels.

Resistance-in-Ohms DIN Rail Transmitter Frequently Asked Questions

Why does excitation current decrease so dramatically from 5 mA on the R0/R1 ranges down to 80 nA on the R7 range?

Documented range table specifically shows excitation current dropping as full-scale resistance increases — since the voltage developed across a resistor for a given current increases with resistance, the much higher resistance values on ranges like R6 and R7 would produce excessive voltage (and excessive power dissipation) if excited with the same current used on the low-ohm ranges; the documented reduction to nanoamp-level excitation on the highest ranges keeps the developed voltage within the transmitter's measurable and safe range.

Why does the transmitter operate ratiometrically with respect to its own excitation current rather than needing a fixed, perfectly stable excitation source?

Documented specification specifically states the applied excitation current is sensed by the transmitter, which operates in ratiometric mode and automatically compensates for any changes in excitation — this means the resistance calculation is based on the actual ratio between sensed voltage and sensed excitation current at the moment of measurement, rather than assuming the excitation source is perfectly constant, which removes excitation drift as a significant error source.

Why does the 2 ohm range (R0) get a special 0.1 milliohm resolution specifically called out for contact resistance measurements?

Documented framing specifically ties this resolution to contact resistance testing — genuine electrical contact resistance in production testing is typically a very small fraction of an ohm, so a range offering 0.1 milliohm resolution is documented as specifically suited to resolving meaningful differences between good and marginal contacts, which a coarser-resolution range wouldn't be able to distinguish.

Does the QA passband relay mode's deviation band apply symmetrically above and below the setpoint, or can it be set asymmetrically?

Documented example specifically describes a deviation limit "set up around both sides of a setpoint," illustrated with a single 50 mΩ figure applied to both sides in that particular example — the documented description doesn't detail whether asymmetric upper and lower limits are separately configurable, so the specific example given represents a symmetric passband around the setpoint.

In passband mode, does the relay open when the reading is acceptable, or when it's out of tolerance?

Documented description specifically states the relay closes (or opens) when the reading falls within the deviation band, and opens (or closes) when the reading falls outside that band — this indicates the relay's specific active/inactive state relative to "in-band" versus "out-of-band" readings is itself a configurable choice (closes on pass or closes on fail), rather than being fixed to only one particular behavior.

Does choosing a factory-fixed range (R0, R6, or R7) versus a jumper-selectable range (R1-R5) affect the transmitter's documented accuracy?

No — the documented accuracy figure (±0.01% of reading ±2 counts) is listed once for all ranges in the table rather than varying between fixed and jumper-selectable ranges; the distinction between factory-fixed and jumper-selectable is specifically about how that particular range is configured into the transmitter, not about a different accuracy specification applying to one group versus the other.

Does the 3-wire lead compensation technique require the two excitation leads to be electrically identical to work correctly?

Yes — documented description specifically qualifies the 3-wire technique's effectiveness with the condition "if the two excitation leads are identical," meaning the subtraction-based compensation specifically relies on both excitation leads having matching resistance and matching response to ambient temperature changes; leads with genuinely different characteristics would reduce how completely this technique cancels lead resistance error.

Can the same physical transmitter be field-reconfigured across different resistance ranges, or does changing range require different hardware?

Documented note specifically states all ranges are factory calibrated and user selectable, and describes signal conditioner boards and ranges as changeable in the field via jumper settings, with calibration factors stored in EEPROM on the board — this points to configuration-level flexibility across the jumper-selectable ranges (R1-R5) on shared hardware, while the factory-fixed ranges (R0, R6, R7) are documented as set at the factory rather than user-jumpered.

Does the analog output's documented 16-bit resolution (0.0015% of span) mean the transmitter can resolve resistance changes finer than the resistance range's own listed resolution?

Not necessarily beyond what the input stage actually measures — the analog output's 16-bit resolution describes how finely the output signal itself can be divided across its span, but the output is documented as tracking a reading whose actual precision is set by the input measurement stage; the output resolution figure describes the output conversion's own granularity, not an independent improvement on the underlying resistance measurement's documented resolution.

Can the Extended board's custom curve linearization be applied to a resistance-based sensor whose output isn't linear with the physical quantity being measured?

Yes — documented capability describes custom curve linearization as a general Extended-board feature using up to 180 user-entered data points to create spline-fit segments, without restricting this to any specific signal type; a resistance-based sensor with a genuinely nonlinear relationship to the physical quantity it's measuring (such as certain thermistor-type sensors) is a documented candidate for this same linearization approach used elsewhere for level or flow signals.

Nanoamp High-Resistance & Insulation Measurement Questions From the Field

Why does measuring very high resistance values specifically require paying attention to surface leakage current, in a way lower-resistance measurements don't?

Documented explanation specifically identifies surface leakage current — current flowing along the outer surface of an insulator rather than through its bulk — as a component of total leakage current that becomes proportionally significant at high resistance levels; at low resistance, the bulk current dominates so completely that surface leakage is negligible by comparison, but at megohm-and-above resistance levels, surface leakage can meaningfully distort the true resistance reading if not addressed.

What is a "guard" terminal or guard technique, and specifically what problem does it solve in high-resistance measurement?

Documented explanation specifically describes the guard terminal as providing a separate current path that diverts surface leakage current away from the actual measurement circuit — by connecting the guard to a point that intercepts surface current before it reaches the measurement electronics, the instrument is documented as then reading only the genuine bulk (volume) resistance of the material being tested, rather than a reading corrupted by surface leakage.

Do high-resistance readings settle to a stable value immediately, or is there a documented practical reason to wait before recording a reading?

Documented guidance specifically notes that due to capacitance and dielectric absorption effects within the material under test, high-resistance readings can take meaningful time to stabilize — practical guidance specifically cites 60 seconds or more as a common wait time used to obtain a genuinely steady insulation resistance reading, rather than recording the initial, still-settling value.

Is there a documented reason cable and connection quality matters more for nanoamp-level measurements than for typical ohmmeter measurements?

Yes — documented guidance specifically recommends using a dedicated shielded measuring cable when measuring high resistance, with the shield itself carrying away leakage current that would otherwise corrupt the reading; at nanoamp signal levels, stray leakage paths through ordinary unshielded cabling or connectors can introduce errors that would be completely insignificant at the milliamp-to-microamp current levels used for low-resistance ranges.

Does a very low ratio between measured resistance and a reference or minimum-acceptable value always indicate the same type of underlying problem?

Documented guidance specifically describes a related diagnostic concept, the Polarization Index — a ratio of resistance readings taken at two different times during a test — where a low ratio (approaching 1.0) is documented as signaling that insulation is likely contaminated with moisture or conductive material, since genuinely sound insulation is documented as showing absorption current that decays normally over the test period rather than being dominated by steady leakage current from the start.

Can environmental contamination affecting a high-resistance measurement be distinguished from genuine, irreversible material degradation?

Documented field practice specifically describes a way to distinguish these: cleaning and thoroughly drying the material under test, then re-testing — if the resistance reading recovers after cleaning and drying, documented guidance attributes the original low reading to environmental contamination (moisture, dirt) rather than permanent damage to the material itself.

Does the specific test voltage used for a high-resistance or insulation measurement matter, or is any convenient voltage adequate?

Documented guidance specifically ties test voltage to the rated voltage of the equipment or material under test, commonly citing test voltages up to roughly twice the rated voltage for equipment rated at or below 500V, with higher-voltage equipment tested nearer to its own rated voltage — using a documented, appropriate test voltage matters specifically because it's meant to genuinely stress the insulation enough to reveal weaknesses that a lower, more casual test voltage wouldn't detect.

Does a documented minimum acceptable resistance value exist independent of what's actually being tested, or does it vary by application?

It varies by application — documented guidance specifically ties minimum acceptable insulation resistance to the voltage rating of the equipment being tested, citing one documented industry rule of thumb of twice the kV rating in megohms (for example, 10 MΩ minimum for a 5 kV-rated panel) — there's no single universal minimum resistance figure documented as applicable across every type of equipment or material.