LT DIN Rail Digital Transmitters with Serial Data Communication and Analog Outputs for Pulse to 4-20 mA Converter for Frequency, Rate, or Period Applications

LT DIN Rail Digital Transmitters with Serial Data Communication and Analog Outputs for Pulse to 4-20 mA Converter for Frequency, Rate, or Period Applications

Price: $334.00
  • P/NLT60FR
- +

Features

  • Inputs from NPN or PNP proximity switches, contact closures, digital logic, magnetic pickups down to 12 mV, or AC inputs up to 250 Vac
  • Arithmetic functions A+B, A-B, AxB, A/B, A/B-1 (draw)
  • Two independently field-scalable pulse input channels from 0.005 Hz to 1 MHz
  • Line frequency measurement to 60.0000 in a few line cycles
  • 4-20 mA, 0-20 mA, 0-10V or -10V to +10V transmitter output, (isolated)
  • Analog output resolution 0.0015%  of span, accuracy ±0.02%  of span
  • RS232 or RS485 serial data, Modbus or Laurel ASCII protocol (isolated)
  • Dual 120 mA solid state relays for alarm or control (isolated)
  • 5V, 10V, 12V, or 24V dc transducer excitation output (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)
  • Optional - Extended main board for simultaneous rate and total, custom curve linearization, arithmetic functions A+B, A-B, AxB, A/B, A/B-1

The Laureate™ LT Series DIN rail analog transmitter with serial data communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy with Input frequencies from 0.005 Hz to 1 MHz. The LT Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and RS232/RS485 output transmitter for pulse to 4-20 mA converter accepts two independently scalable input channels from a wide range of pulse sources, such as NPN or PNP proximity switches, contact closures, digital logic, magnetic pickups down to 12 mV, or AC voltages to 250 Vac. Input frequencies can range from 0.005 Hz to 1 MHz. Applications include the measurement of AC line frequency, RPM, speed or rate from proximity switch inputs, and flow rate from turbine flow meter inputs.

  • With a Standard main board, the transmitter output can be scaled to track frequency in Hz, rate (such as gallons per minute), or period (inverse of frequency). Square root extraction is standard.

  • With an Extended main board, the transmitter output can track rate or totalized rate (such as gallons) whether the transducer output is linear, requires square root extraction, or requires custom curve linearization. The latter can be provided by a curvilinear spline fit with up to 180 data points. The transmitter can also count up to a preset total or down from a preset total to zero. Such applications typically make use of optional dual solid state relays, which are available as options. External reset of totals is via a special three-position screw terminal connector. The two input channels A & B can also be combined arithmetically to provide an analog transmitter output that tracks A+B (e.g., sum of two flows), A-B (e.g., difference of two flows), AxB (e.g., horsepower as product of force and RPM), A/B (ratio of two flow), and A/B-1 (draw or relative elongation of material between rollers).

Exceptional Accuracy and Stability. Laureate transmitters determine frequency by taking the inverse of period as measured with a calibrated quartz crystal time base. This results in extremely accurate and stable 6-digit internal readings (±999,999 counts), which are then processed in software. The analog output is generated by an ultra-linear 16-bit (65,536 step) digital-to-analog converter (DAC) for 0.02% output accuracy. The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

Standard Features of Laureate LT transmitters Include:

  • Serial communications output, (isolated), RS232 or RS485 (half or full duplex), jumper selectable. Three protocols are user selectable: Modbus RTU, Modbus ASCII, or Laurel ASCII. Modbus operation is fully compliant with Modbus Over Serial Line Specification V1.0 (2002). The Laurel ASCII protocol is simpler than the Modbus protocol and is recommended when all devices are Laureates.
  • 4-20 mA, 0-10V or -10V to +10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 0.0015% resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. Note that Ethernet data I/O is provided by Laurel's LTE Series transmitters.
  • Dual-channel pulse inputs for voltage signals, NPN or PNP proximity switches, contact closures, magnetic pickups or flow meters.
  • Dual solid state relays, (isolated), for alarm or control. Rated 120 mA at 130 Vac or 170 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Removable screw terminal connections of Laurel transmitters

LT series DIN rail Transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LT series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an RS232 or RS485 serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LT Transmitters to a Local Area Network (LAN)

Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for LT Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

4-20 mA Transmitter for Frequency, Rate or Period Input

Signal Input
Types AC, pulses from NPN, PNP transistors, contact closures, magnetic pickups
Grounding Common ground for channels A & B
Chan. A Frequency 0.005 Hz to 1 MHz
Chan. B Frequency 0.005 Hz to 250 kHz
Minimum Signal Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V)
Maximum Signal 250 Vac
Noise Filter 1 MHz, 30 kHz, 250 Hz (selectable)
Contact Debounce 0, 3, 50 ms (selectable)
Time Base Accuracy Quartz crystal calibrated to ±2 ppm
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Analog Output (standard)
Output Levels 4-20 mA, 0-20 mA, 0-10 Vdc, -10 to +10Vdc (user selectable)
Compliance, 4-20 mA 10V (0-500Ω load)
Compliance, 0-10V 2 mA (5 kΩ load)
Output Resolution 16 bits (65,536 steps)
Output Accuracy 0.02% of output span
Output Update Rate Programmed gate time + 30 ms + 0-2 signal periods
Output Isolation 250V rms working, 2.3 kV rms per 1 minute test
Serial Data Output (standard)
Signal Types RS232 or RS485 (half or full duplex), jumper selectable
Data Rates 300, 600, 1200, 2400, 4800, 9600, 19200 baud
Output Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Protocols Modbus RTU, Modbus ASCII, Custom ASCII
Modbus Compliance Modbus over Serial Line Specification V1.0 (2002)
Ethernet Connector RJ45 jack for 10/100Base-T Ethernet cable
RS232/485 Connector Screw terminals for easy daisy chaining
Digital Addresses 247 for Modbus, 31 for Custom ASCII
Dual Relay Output (standard)
Relay Type Two solid state relays, SPST, normally open, Form A
Load Rating 120 mA at 140 Vac or 180 Vdc
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Power Input
Standard Power 85-264 Vac or 90-300 Vdc
Low Power Option 10-48 Vdc or 12-32 Vac
Power Frequency DC or 47-63 Hz
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Power Consumption at 24V 1.5W typical, 3W with max excitation output
Environmental
Operating Temperature -40°C to70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Cooling Required Mount transmitters with ventilation holes at top and bottom. Leave 6 mm (1/4") between transmitters, or force air with a fan.
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options available using Laurel's free Instrument Setup Software.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

Laureate LT transmitter pinout

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

AC Line Frequency
AC Line Frequency Meter or Transmitter Laureate frequency meters and transmitters accept line voltages up to 250 Vac and output line frequency to 6-digit accuracy (50.0000 or 60.0000) in a few line cycles. Fast low frequency response is achieved by timing the period and taking its inverse.
RPM and Speed
Tachometer and Pulse Input Rate Meter or Transmitter Laureate frequency meters and transmitters can sense low-level signals from magnetic pickups or the NPN or PNP transistor output of active sensors. These can be powered directly by the meter or transmitter. Output in RPM or units of speed is achieved by mathematical scaling.
Simultaneneous Flow Rate and Total
Simultaneous Rate Meter and Totalizer Mode of Frequency Meter or Transmitter Laureate frequency meters and transmitters are compatible with all flow meters which generate pulses at a frequency proportional to flow rate. The Extended version can display scaled rate or total for the same input at the push of a button, and alarm from both the rate and total.
Combining Two Rates
Dual Pulse Input Rate Meter or Transmitter to Combine Two Rates Extended Laureate frequency meters and transmitters offer A+B, A-B and A/B arithmetic functions. A+B allows two input flows to be summed for total flow, while A-B allows outflow to be subtracted from inflow for net flow. Flow ratios aid in the proper mixing of ingredients.
Custom Curve Linearization
Custom Curve Linearization for Laureate Rate Meters and Transmitters Extended Laureate frequency meters and transmitters can linearize the output turbine flow meters, which tend to be nonlinear on the low end. Linearizing improves the dynamic range and accuracy of turbine flow meters.
System-level Capabilities
Laureate dual-channel rate meters and transmitters can independently scale, display and alarm two pulse input channels. All signal or alarm data can further be transmitted via RS232 or RS485, including peak readings and arithmetic combinations of the two rates. System Interfaces of Laureate Frequency Meters, Rate Meters and Transmitters

 

 

CAL-Digital

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LT60FR
Price as Configured: $334.00

Click on the Option Board Links for More Product Information

Base Item
$164.00
Main Board
$0.00
With Standard Main Board: Scalable to ±999,999 for frequency, rate, square root of rate, up or down total, period, A-to-B time interval.
With Extended Main Board: Above, plus rate and total simultaneously, ratio (A/B), draw (A/B-1), other arithmetic functions (AxB, A+B, A-B), phase angle, stopwatch, up/down counting, batching operation, linearization of nonlinear inputs.
$55.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
Part Number as Configured:
LT60FR
Price as Configured:
$334.00
Quantity:
- +
Extended Price:
$334.00

Understanding the Laureate™ LT Series DIN Rail Transmitter for Frequency, Rate, or Period

The Laureate™ LT Series DIN rail transmitter for pulse to 4-20 mA conversion accepts two independently scalable input channels from a wide range of pulse sources, such as NPN or PNP proximity switches, contact closures, digital logic, magnetic pickups down to 12 mV, or AC voltages to 250 Vac. Input frequencies range from 0.005 Hz to 1 MHz on Channel A, and 0.005 Hz to 250 kHz on Channel B. Applications include AC line frequency measurement, RPM, speed or rate from proximity switch inputs, and flow rate from turbine flow meter inputs.

Standard vs. Extended Main Board

With a Standard main board, the output can be scaled to track frequency in Hz, rate (such as gallons per minute), or period (inverse of frequency); square root extraction is standard. With an Extended main board, the output can track rate or totalized rate whether the transducer output is linear, requires square root extraction, or requires custom curve linearization (via curvilinear spline fit with up to 180 data points). The Extended board can also count up to a preset total or down from a preset total to zero — external reset of totals is via a special three-position screw terminal connector — and can combine Channels A and B arithmetically: A+B (sum of two flows), A-B (difference of two flows), AxB (horsepower as product of force and RPM), A/B (ratio of two flows), and A/B-1 (draw, or relative elongation of material between rollers).

Accuracy, Stability, and Update Rate

Frequency is determined by taking the inverse of period as measured with a calibrated quartz crystal time base (±2 ppm), producing extremely accurate and stable 6-digit internal readings (±999,999 counts). The analog output is generated by an ultra-linear 16-bit (65,536 step) DAC for 0.02% output accuracy. Output update rate is programmed gate time plus 30 ms plus 0-2 signal periods — for a 60 Hz signal, update rate is 20 per second, ideal for alarm and control.

Signal Specifications

Nine minimum-signal ranges span -12 to +12 mV up to +1.25 to +2.1V; maximum signal is 250 Vac. Noise filter is selectable at 1 MHz, 30 kHz, or 250 Hz; contact debounce is selectable at 0, 3, or 50 ms.

Real-World Applications

  • AC Line Frequency — accepts line voltages up to 250 Vac, outputs line frequency to 6-digit accuracy (50.0000 or 60.0000) in a few line cycles; fast low-frequency response is achieved by timing the period and taking its inverse.
  • RPM and Speed — senses low-level signals from magnetic pickups or NPN/PNP active sensors, which can be powered directly by the transmitter; output in RPM or speed units via mathematical scaling.
  • Simultaneous Flow Rate and Total — compatible with any flow meter generating pulses at a frequency proportional to flow rate; the Extended version displays scaled rate or total for the same input at the push of a button, alarming from both.
  • Combining Two Rates — A+B sums two input flows for total flow; A-B subtracts outflow from inflow for net flow; ratios aid proper ingredient mixing.
  • Custom Curve Linearization — linearizes turbine flow meter output, which tends to be nonlinear at the low end, improving dynamic range and accuracy.

Factory-Calibrated Accuracy

All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM, enabling field replacement of signal conditioner boards without necessitating recalibration of the transmitter. Factory recalibration is recommended annually.

Where Frequency, Rate & Period DIN Rail Transmitters Are Used

  • Turbine & Paddlewheel Flow Metering — pulse-to-4-20 mA conversion for flow rate and totalization.
  • Motor & Conveyor Speed Monitoring — RPM retransmission from magnetic pickups or proximity switches.
  • Power Generation Frequency Monitoring — precision line frequency measurement and alarming.
  • Ingredient Mixing & Ratio Control — dual-channel arithmetic for blend ratio monitoring.
  • Net Flow Metering — A-B subtraction for inflow/outflow balance applications.
  • Multi-Point RS485 Rate Networks — daisy-chained transmitters reporting to a central controller.
  • OEM Pulse Signal Conditioning — DIN rail integration into existing control panels.

Frequency, Rate & Period DIN Rail Transmitter Frequently Asked Questions

Why does Channel A support frequencies up to 1 MHz while Channel B is limited to 250 kHz?

Documented specification lists these as genuinely different maximum frequencies for the two channels rather than a shared figure — this asymmetry means an application specifically needing to measure a very high-frequency signal (up to 1 MHz) should route that signal to Channel A, while Channel B remains suited to signals within its own documented, lower maximum frequency.

Does square root extraction being "standard" on the Standard main board mean it's always active, or is it a selectable option?

Documented phrasing specifically describes square root extraction as a standard capability of the Standard board, distinguishing it from features requiring the Extended board — this indicates it's available as a configuration option on the Standard board (relevant for signals like differential-pressure-based flow that need square root correction), rather than a feature the Standard board lacks entirely.

Why does external reset of totals require a special three-position screw terminal connector rather than a standard two-wire contact input?

Documented specification specifically calls out a three-position connector for this function, though the page doesn't detail the specific reason for the third terminal — this is presented as the documented physical interface for triggering an external total reset, distinct from the transmitter's other two general-purpose control inputs described elsewhere in its documentation.

Does the A/B-1 (draw) function apply only to material tension between rollers, or can it be used for other ratio-based measurements?

Documented example specifically illustrates A/B-1 with material draw (relative elongation between rollers) as one application, alongside A/B itself being documented separately as flow ratio for ingredient mixing — since A/B-1 is mathematically just A/B with 1 subtracted, it's a general-purpose relative-difference-ratio calculation that the documented example applies to draw specifically, though the underlying math could suit other applications needing a similar relative-ratio figure.

Why does the output update rate formula include "0-2 signal periods" as a variable component rather than being a fixed number?

Documented formula (gate time + 30 ms + 0-2 signal periods) specifically reflects that after the programmed gate time and fixed 30 ms overhead elapse, the transmitter may need to wait for the current signal period to complete before returning a valid reading — since where in its cycle the input signal happens to be when the gate time ends is not fixed, this documented uncertainty range accounts for that variability.

Does custom curve linearization for a nonlinear turbine flow meter need to be re-entered if the meter's own K-factor calibration changes?

The documented custom curve linearization process specifically uses up to 180 user-entered data points reflecting the actual relationship between the transducer's raw output and true flow at the time those points were established — since a changed K-factor would shift that underlying relationship, the documented linearization data would need to be re-derived and re-entered to remain accurate, rather than automatically adjusting to a new calibration on its own.

Can the AxB arithmetic function (documented example: horsepower from force and RPM) be reconfigured for a different multiplicative relationship?

Documented framing presents horsepower as one specific example application of the general AxB multiplication function — since AxB is a general-purpose multiplication of whatever is measured on Channel A by whatever is measured on Channel B, the same documented function supports any other application genuinely needing the product of two rate or frequency inputs, not solely the horsepower example given.

Does peak reading transmission (mentioned under system-level capabilities) apply to both Channel A and Channel B independently, or only to a combined arithmetic result?

Documented system-level description specifically states that peak readings, along with arithmetic combinations of the two rates, can be transmitted via RS232 or RS485 — this is presented as available for the individually scaled and displayed channels as well as for their combined arithmetic result, rather than being limited to only one or the other.

Does selecting a shorter noise filter setting (250 Hz) instead of 1 MHz improve accuracy, or only noise rejection?

Documented specification lists these three filter settings (1 MHz, 30 kHz, 250 Hz) as noise filter options rather than accuracy settings — the correct choice depends on matching the filter bandwidth to the actual signal frequency being measured; selecting too narrow a filter (such as 250 Hz) for a genuinely higher-frequency input signal would filter out the signal itself along with the noise, so the filter setting should be chosen based on the input signal's own documented frequency, not simply toward the narrowest setting for "better" noise rejection.

Does the transmitter's ±2 ppm crystal time base accuracy apply equally to both the frequency measurement and the totalized count in Extended mode?

Documented specification lists the ±2 ppm figure specifically under Time Base Accuracy, describing the fundamental timing reference the transmitter uses to derive frequency, rate, and period readings — since totalization in Extended mode is built from the same underlying pulse-counting and timing mechanism, the same time base accuracy figure applies to both the instantaneous frequency/rate reading and the accumulated total, rather than the total being governed by a separate, undocumented accuracy specification.

Turbine Flow Meter Straight-Run Piping Questions From the Field

Why do turbine flow meters specifically require a documented minimum length of straight pipe upstream and downstream?

Documented guidance specifically explains that turbine meters must be installed to minimize measurement errors caused by turbulence, since the meter's accuracy depends on a properly developed flow profile reaching the rotor; elbows, valves, reducers, and other pipe fittings upstream are documented as distorting that flow profile, which is why a defined length of undisturbed straight pipe is specified before the meter.

What are the commonly documented turbine flow meter straight-run figures, and do they differ between upstream and downstream?

Yes — documented industry guidance commonly cites roughly 15-20 pipe diameters upstream (sometimes including the length of an upstream strainer) and around 5 pipe diameters downstream as general figures for turbine meters, with upstream requirements documented as consistently longer than downstream, since upstream disturbances have a greater documented effect on the flow profile actually reaching the meter's rotor.

Does having two elbows upstream of a turbine meter require more straight run than a single elbow, and does the specific arrangement matter?

Yes — documented guidance specifically distinguishes these cases: two elbows in the same plane are documented as requiring a smaller increase in straight run than two elbows in different planes, with the latter specifically cited as potentially requiring as much as 50 pipe diameters upstream, since out-of-plane elbows are documented as inducing a more complex, harder-to-settle swirling flow pattern.

Can a flow conditioner or straightening device reduce the documented straight-run requirement for a turbine meter installation?

Yes — documented guidance specifically describes flow conditioners (such as tube-bundle or perforated-plate types) as capable of reducing required upstream straight run, with one documented example specifically citing a reduction from around 10 pipe diameters down to roughly 5 when a conditioner is used, at the cost of some additional documented pressure loss through the conditioner itself.

Why is a strainer commonly recommended upstream of a turbine flow meter specifically, beyond general flow profile concerns?

Documented guidance specifically identifies this as protection for the meter's moving parts — a turbine meter's rotor and bearings are documented as vulnerable to damage from contaminants entrained in the flow, so an upstream strainer is specifically recommended to capture debris before it reaches the rotor, independent of the separate straight-run requirement addressing flow profile distortion.

Why is maintaining adequate downstream backpressure specifically important for a turbine flow meter, beyond simply having enough straight run?

Documented guidance specifically explains that insufficient downstream backpressure can cause cavitation or flashing at the meter, both distinct from straight-run flow-profile issues — one documented rule of thumb specifically recommends downstream pressure of at least 1.25 times the fluid's vapor pressure, plus twice the pressure drop through the meter itself, specifically to prevent these damaging conditions.

Does even a small amount of entrained air or gas in the flow stream meaningfully affect turbine meter accuracy?

Yes — documented guidance specifically cites entrained air or gas above roughly 100 mg/l as capable of producing a documented measurement error, with the meter tending to read high in the presence of entrained gas; larger quantities of entrained air are further documented as capable of causing physical damage to the rotor itself, beyond just an accuracy error.

Do all flow meter technologies share the same straight-run sensitivity as turbine meters, or do some technologies avoid this requirement?

No — documented comparison specifically notes that Coriolis flow meters measure mass flow and density directly and are documented as insensitive to fluid flow profile, requiring no upstream or downstream straight pipe run at all, in clear contrast to turbine meters' documented sensitivity to flow profile distortion; positive displacement meters are similarly documented as measuring volume mechanically and being unaffected by flow profile, though with different documented maintenance tradeoffs from their moving parts.