LT DIN Rail Digital Transmitters for Batch Controller Analog Input Applications

LT DIN Rail Digital Transmitters for Batch Controller Analog Input Applications

Price: $389.00
  • P/NLT80VF1
- +

Features

  • Provides automatic control for repetitive liquid fill operations
  • Available for turbine flow meter pulse signals from 0 Hz to 1 MHz or analog flow meter signals (4-20 mA, 0-1 mA or 0-10V)
  • ±999,999 for batch total, grand total, number of batches, or flow rate
  • Counts up from 0 to preset or down from preset to 0
  • Two or four control relays for with settable delay between cycles
  • All input ranges are user selectable and factory calibrated
  • Digital span adjustment: 0 to ±999,999; zero adjustment: -999,999 to +999,999
  • Up to 60 conversions per second, Ideal for peak or valley capture
  • Scalable to ±999,999 for use with current shunts
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Transducer excitation output, 5, 10, 12, or 24 Vdc (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • Operating temperature from -40°C to 70°C  (-40°F to 158°F)

The Laureate™ LT Series DIN rail analog transmitter with serial data communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy with Input frequencies from 0.005 Hz to 1 MHz. The LT Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and RS232/RS485 output transmitter for batch controller is a low cost, powerful and highly accurate batching controller for repetitive fill operations. It can use the Laureate V-to-F analog signal conditioner for use with 4-20 mA, 0-1 mA or 0-10V conditioned flow meter signals. Relay control can be provided by two or four 8A contact relays, or by two or four 120 mA AC/DC solid state relays. Fill operations are repeated continually with a programmable delay from 10 ms to 199.99 sec, or based on an external control input.

Three items are tracked by the batch control software. These can each be scaled to engineering units of total or flow rate and displayed by the controller's six-digit LED display: Item #1 is the current batch total, which can be set up to count up from zero to a preset limit, or down from a preset limit to zero. Item #2 can be assigned to grand total or number of batches. Item #3 is the flow rate.

Batch Control with Conditioned Flow Signals

The analog input batch controller utilizes the Laureate VF voltage-to-frequency converter signal conditioner board, which converts 4-20 mA, 0-1 mA or 0-1V conditioned flow meter signals to a frequency from 10 kHz to 110 kHz. This allows the counter controller to totalize flow, to count up to a preset value, or to count down to zero from a preset value for batch control. One of the relays is dedicated to On/Off batch control, while the other relay is available to slow down rate near the setpoint or to provide another alarm or control function based on rate or total.

The Extended Laureate computer board can display rate based on successive readings. It also allows exceptionally accurate custom curve linearization, for example to read out liquid volume or rate of flow in a horizontal cylindrical tank based on level reported by a 4-20 mA transmitter. For setup, up to 180 data points can be input into a computer spreadsheet or text file by the user. The computer then calculates spline-fit segments, which are downloaded into the transmitter.

Exceptional Accuracy and Stability. Laureate transmitters determine frequency by taking the inverse of period as measured with a calibrated quartz crystal time base. This results in extremely accurate and stable 6-digit internal readings (±999,999 counts), which are then processed in software. The analog output is generated by an ultra-linear 16-bit (65,536 step) digital-to-analog converter (DAC) for 0.02% output accuracy. The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For pulse rates of 60 Hz and above, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

Standard Features of Laureate LT Transmitters Include:

  • Serial communications output, (isolated), RS232 or RS485 (half or full duplex), jumper selectable. Three protocols are user selectable: Modbus RTU, Modbus ASCII, or Laurel ASCII. Modbus operation is fully compliant with Modbus Over Serial Line Specification V1.0 (2002). The Laurel ASCII protocol is simpler than the Modbus protocol and is recommended when all devices are Laureates.
  • 4-20 mA, 0-10V or -10V to +10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 0.0015% resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. Note that Ethernet data I/O is provided by Laurel's LTE series transmitters.
  • Dual solid state relays, (isolated), for alarm or control. Rated 120 mA at 130 Vac or 170 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA,  or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.
  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions for an update every .266 seconds.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Removable screw terminal connections of Laurel transmitters

LT series DIN rail Transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LT series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an RS232 or RS485 serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LT Transmitters to a Local Area Network (LAN)

Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for LT Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

Pulse Inputs (FR signal conditioner)
Signal Types AC, pulses from NPN, PNP transistors, contact closures,magnetic pickups.
Ch A Frequency, Max 1 MHz
Ch B Frequency, Max 250 kHz
Signal Ground Common ground for channels A & B
Minimum Signal Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V)
Maximum Signal 250 Vac
Maximum Frequency 1 MHz, 30 kHz, 250 Hz (selectable)
Conversion Technique Inverse period
Delay between batches Selectable 10 ms to 199.99 s
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Power Supply Boards (one required)
Voltage, standard 85-264 Vac or 90-300 Vdc
Voltage, optional 12-32 Vac or 10-48 Vdc
Frequency DC or 47-63 Hz
Power consumption 1.2W @ 120 Vac, 1.5W @ 240 Vac, 1.3W @ 10 Vdc, 1.4W @ 20 Vdc,
   (typical, base meter) 1.55W @ 30 Vdc, 1.8W @ 40 Vdc, 2.15W @ 48 Vdc
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Analog Output Boards (one optional)
Output Levels 4-20 mA, 0-20 mA, 0-10V, -10 to +10V (single-output option)
Current compliance 4-20 mA, 0-20 mA, 0-10V (dual-output option)
Voltage compliance 2 mA at 10V ( > 5 kΩ load)
Scaling 12V at 20 mA ( < 600 Ω load)
Resolution Zero and full scale adjustable from -99999 to +99999 16 bits (0.0015% of full scale)
Isolation 250V rms working, 2.3 kV rms per 1 min test (dual analog outputs share the same ground)
Relay Output Boards (one required for batch control)
Dual magnetic relays 2 Form C, 8A max, 440Vac or 125Vdc max, 2500VA or 300W
Quad magnetic relays 4 Form A (NO), 8A max, 440Vac or 125Vdc max, 2500VA or 300W
Dual solid state relays 2 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Quad solid state relays 4 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Relay commons Isolated commons for dual relays or each pair of quad relays
Relay isolation 250V rms working, 2.3 kV rms per 1 minute test
Relay latching modes Latching or non-latching
Relay active modes Active on or off, active high or low
Hysteresis modes QA passband mode, split hysteresis, span hysteresis
Communication Boards (one optional)
Board Selections RS232RS485 with dual RJ11 connectors, RS485 with dual RJ45 connectors, USB, High-Speed Ethernet, USB-to-RS485 gatewayHigh-Speed Ethernet-to-RS485 gateway, WiFi with built-in antenna plus USB & RS485, WiFi with external antenna plus USB & RS485
Protocols Laurel Custom ASCII (serial), Modbus RTU (serial), Modbus TCP (Ethernet or WiFi)
Digital Addresses 247 (Modbus), 31 (Laurel ASCII),
Isolation 250V rms working, 2.3 kV rms per 1 min test
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Protection NEMA-4X (IP-65) when panel mounted
Signal Connections
Signal connections to Laurel pulse input batch controller
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options include using the front panel buttons, the free Instrument Setup Software, or a hardware jumper.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

Laureate LT transmitter pinout

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

Up/Down Totalizing
Dual-channel up/down totalizing using one channel for counts and one channel for count direction Up/down totalizing is provided by a mode of the Extended totalizing transmitter where pulses are either added or subtracted on Channel A based on a direction input on Channel B. The unit can also be programmed so that counting by Channel A is inhibited by an input on Channel B.
Combining Two Totals
Dual-channel up/down totalizing, where two channels are added or subtracted. A+B, A-B, A/B and AxB arithmetic functions are available with the Extended totalizing transmitter. A+B can sum two totals, while A-B subtracts the outflow total from inflow total. A/B ratio applied to two totals helps assure the proper mixing of components.
Up or Down Counting with Preset
Two repetitive fill operations handled by a single Laureate dual-channel digital counter and totalizer A single Laureate dual-channel totalizing transmitter will handle two repetitive fill operations by counting from zero up to a preset, or down from a preset to zero. The dual relay option is required.
Machine ON Time and Utilization
Dual-channel up/down totalizing to measure machine ON time by counting power line cycles. An easy way to measure the ON time of machines is to count AC line cycles and scale the total to hours. To display machine utilization or duty cycle in percent, use the Extended totalizing transmitter. Connect Channel A to switched AC and Channel B to the AC line, and apply a 100 multiplier to the A/B ratio.
Custom Curve Linearization
Displaying rate and total in the same Laureate totalizer and rate meter or transmitter The Extended version of the Laureate dual channel totalizing transmitter can transmit scaled rate or total for the same channel at the push of a button, and alarm both the rate and total. The Extended version can also do curve linearization, thereby extending the working range and accuracy of flow transducers.

 

 

CAL-Digital

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LT80VF1
Price as Configured: $389.00

Click on the Option Board Links for More Product Information

Base Item
$164.00
Main Board
$55.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
$81.00
$81.00
Part Number as Configured:
LT80VF1
Price as Configured:
$389.00
Quantity:
- +
Extended Price:
$389.00

Understanding the Laureate™ LT Series DIN Rail Transmitter for Batch Controller Analog Input

The Laureate™ LT Series DIN rail transmitter for batch controller analog input is a low cost, powerful, and highly accurate batching controller for repetitive fill operations, using the Laureate V-to-F analog signal conditioner for 4-20 mA, 0-1 mA, or 0-10V conditioned flow meter signals. Fill operations repeat continually with a programmable delay from 10 ms to 199.99 sec, or based on an external control input.

Item #1/#2/#3 Tracking

Three items are tracked by the batch control software, each scalable to engineering units and displayed on the controller's six-digit LED display. Item #1 is the current batch total, settable to count up from zero to a preset limit, or down from a preset limit to zero. Item #2 can be assigned to grand total or number of batches. Item #3 is the flow rate.

Two-Relay Fill Control

The Laureate VF voltage-to-frequency converter signal conditioner board converts 4-20 mA, 0-1 mA, or 0-10V conditioned flow meter signals to a frequency from 10 kHz to 110 kHz, allowing the controller to totalize flow, count up to a preset value, or count down to zero from a preset value for batch control. One relay is dedicated to On/Off batch control, while the second relay is available to slow down rate near the setpoint, or to provide another alarm or control function based on rate or total.

Relay Options and Hysteresis Modes

Relay control is provided by two or four 8A contact relays, or two or four 120 mA AC/DC solid state relays. Relay latching modes are latching or non-latching; active modes are active on or off, active high or low. Hysteresis modes include QA passband mode, split hysteresis, and span hysteresis.

Custom Curve Linearization

The Extended Laureate computer board can display rate based on successive readings and allows exceptionally accurate custom curve linearization — for example, reading out liquid volume or flow rate in a horizontal cylindrical tank based on level reported by a 4-20 mA transmitter. Up to 180 data points are entered into a spreadsheet or text file; the computer calculates spline-fit segments, downloaded into the transmitter.

Real-World Batching Applications

  • Up/Down Totalizing — pulses are added or subtracted on Channel A based on a direction input on Channel B; counting can also be inhibited by a Channel B input.
  • Combining Two Totals — A+B sums two totals; A-B subtracts outflow total from inflow total; A/B ratio applied to two totals helps assure proper mixing of components.
  • Up or Down Counting with Preset — a single transmitter handles two repetitive fill operations, counting from zero up to a preset or down from a preset to zero; the dual relay option is required.
  • Machine ON Time and Utilization — count AC line cycles and scale to hours for ON time; connect Channel A to switched AC and Channel B to the AC line, applying a 100 multiplier to the A/B ratio for duty cycle percent.
  • Custom Curve Linearization — the Extended version transmits scaled rate or total for the same channel at the push of a button, alarms both, and extends the working range and accuracy of flow transducers.

Factory-Calibrated Accuracy

All signal conditioner board ranges are factory-calibrated, with calibration factors stored in EEPROM, enabling field replacement of signal conditioner boards without necessitating recalibration of the transmitter. Factory recalibration is recommended annually.

Where Batch Controller Analog Input DIN Rail Transmitters Are Used

  • Chemical & Liquid Batch Dosing — precise repetitive fill control from conditioned 4-20 mA flow signals.
  • Two-Speed Fast/Dribble Fill Systems — dedicated relays for coarse fill and setpoint-approach slowdown.
  • Multi-Component Mixing — combined-total ratio checks for accurate batch composition.
  • Dual-Station Fill Lines — two independent preset-based fill operations on one transmitter.
  • Machine Runtime & Utilization Tracking — AC-line-cycle-based ON time and duty cycle monitoring.
  • Nonlinear Tank & Sensor Linearization — custom curve correction for irregular tank geometry.
  • OEM Batch Control Instrumentation — DIN rail integration into existing control panels.

Batch Controller Analog Input DIN Rail Transmitter Frequently Asked Questions

Why is only one of the two relays specifically dedicated to On/Off batch control, while the other is left flexible?

Documented description specifically frames the first relay as fixed to On/Off batch control (starting and stopping the actual fill), while the second is documented as available for a genuinely different purpose — slowing the rate near the setpoint, or serving another alarm/control function; this reflects that on/off batch control and setpoint-approach rate reduction are two documented, functionally distinct control actions, so dedicating one relay to each allows both to operate independently and simultaneously.

Does using the second relay to "slow down rate near the setpoint" require external valve or pump hardware capable of two flow rates?

Documented description specifically frames this relay as being "available" for that purpose without detailing the specific external hardware required — since slowing rate near a setpoint inherently requires the process itself (a valve, pump, or feeder) to actually support a reduced flow state when commanded, the transmitter's documented role is providing the timely control signal, while the physical two-speed capability itself is documented as residing in the external fill equipment the relay is wired to.

Why does the VF signal conditioner convert the analog input to a 10 kHz-110 kHz frequency range specifically, rather than a lower frequency range?

Documented specification lists this specific frequency range for the V-to-F converter without further detail on the underlying design rationale — this range is consistent with the same inverse-period timing technique used throughout the LT Series, where a higher converted frequency generally supports faster, more responsive update rates for a given gate time, which is documented elsewhere as beneficial for batch control's need for fast alarm and control response.

Does this analog-input batch controller support turbine flow meter pulse inputs directly, or does it require the flow signal to already be conditioned to 4-20 mA/0-1 mA/0-10V?

Documented signal input options for this specific transmitter variant are limited to VF1 (4-20 mA), VF2 (0-1 mA), and VF3 (0-10V) — a genuinely different, separate LT Series batch controller variant is documented as using the FR pulse signal conditioner for direct turbine flow meter pulse input; this analog-input version is documented as specifically requiring an already-conditioned analog flow signal rather than accepting a raw turbine meter pulse train.

Does the delay between batches (10 ms to 199.99 s) begin counting from batch completion, or from when the next fill cycle is externally triggered?

Documented description specifically states fill operations are "repeated continually with a programmable delay... or based on an external control input," presenting these as two separate triggering approaches — in the continually-repeating mode, the documented delay is consistent with counting from batch completion before the next cycle automatically begins, while the external-control-input mode is documented as a separate, alternative triggering method not governed by the same fixed delay timer.

Can the same physical transmitter be reconfigured between VF1, VF2, and VF3 input types, or does each require separate hardware?

The documented ordering structure lists VF1, VF2, and VF3 as separate signal input board selections at time of order, rather than describing them as software-only configuration options on one shared board — this is consistent with the specific V-to-F signal conditioner board itself being matched to the intended analog input type at the hardware level, distinct from other LT Series transmitters where a single board is documented as user-configurable across multiple signal types.

Does the documented span adjustment range (0 to ±999,999) and zero adjustment range (-999,999 to +999,999) apply identically to Item #1, Item #2, and Item #3?

Documented specification lists these adjustment ranges once under the general digital span/zero adjustment specification, without separately restricting them per tracked item — since Item #1, #2, and #3 are each documented as independently scalable to engineering units, the same underlying ±999,999 adjustment capability is consistent with applying to whichever of the three items is being configured at a given time.

Can the second relay's "slow down rate near setpoint" function and its "alarm or control function based on rate or total" function be used at the same time on one transmitter?

No — documented description presents these as alternative uses for the same single relay ("available to slow down rate near the setpoint, or to provide another alarm or control function"), meaning a given transmitter configuration assigns that second relay to one of these two documented roles at a time, not both simultaneously; a second physical relay (via the quad relay option) would be needed to implement both functions concurrently.

Does the batch average filter's documented update rate of once every .266 seconds apply to Item #1, #2, and #3 readings simultaneously, or to the analog output only?

Documented specification describes the batch average filter as averaging each 16 conversions for this update rate, framed as a general signal-conditioning behavior rather than one specific to a single tracked item — since Items #1 through #3 are all derived from the same underlying conditioned signal, this filter setting is consistent with affecting the update behavior of all three tracked items together, not selectively applying to only one.

Does the transducer excitation output (5/10/12/24 Vdc) play any role in the analog batch controller's own flow signal input, or is it solely for powering external sensors?

Documented specification lists the excitation output as standard, intended to power transducers or two-wire transmitters supplying the 4-20 mA/0-1 mA/0-10V signal into the VF converter — this excitation output is documented as a separate, supporting function for powering the external flow-conditioning transducer that produces the analog input, rather than being part of the transmitter's own internal V-to-F conversion or totalizing circuitry.

Fast Fill/Dribble Fill Two-Speed Batching Questions From the Field

What specifically is the difference between "fast fill" and "dribble fill" in a two-speed batching system?

Documented terminology specifically defines fast fill as the initial, relatively high-rate filling stage used to introduce material into a container quickly, while dribble fill is the documented term for the subsequent slower fill rate used for the final portion — the switchover from fast to dribble fill, and the eventual termination of dribble fill, are documented as governed by feedback signals (typically from a scale or level sensor) rather than a fixed time or volume alone.

Why does switching to a slower fill rate near the target specifically improve batch accuracy, rather than simply shutting off the fast fill valve exactly at target?

Documented explanation specifically identifies valve reaction time and hydraulic shock (water hammer) as the underlying problem a fast-only shutoff would face — material continues to flow briefly after a valve command to close, and this documented "coast" effect is proportionally larger at high flow rates; slowing to a documented dribble rate before final cutoff reduces the amount of material delivered during that reaction-time coast, improving final accuracy.

Is there a documented typical proportion of the total batch delivered during the fast fill stage versus the dribble stage?

Yes — one documented industry guideline specifically describes the fast fill (coarse) stage delivering roughly 80-90% of total batch volume quickly, with the remaining portion delivered during the slower dribble stage; this documented split reflects the goal of using dribble fill only for the final accuracy-critical portion, not the majority of the batch, to keep overall cycle time reasonably short.

Do inaccuracies in flow-meter-based batching tend to concentrate at particular points in the batch cycle, or are they spread evenly throughout?

Documented field analysis specifically states that inaccuracies with flow-meter-based batch processes are generally concentrated at the beginning and end of the batch — this is broadly consistent with why two-speed fill control specifically targets the end-of-batch portion with a slower, more controllable dribble rate, since that documented error-prone region benefits the most from finer control.

Can a flow meter's own accuracy specification become a limiting factor specifically during the dribble fill portion of a batch?

Yes — documented guidance specifically warns that flow meter accuracy must remain consistent across a demanding turndown ratio (the ratio between the fast-fill flow rate and the much lower dribble-fill flow rate), and specifically flags that some flow meter technologies derate (lose accuracy) at very low flow rates; verifying a specific meter's accuracy at the actual dribble-fill flow rate, not just its nameplate full-scale accuracy, is documented as a genuine specification check.

Does the specific valve or actuator technology used for switching between fast and dribble fill rates genuinely matter for batch accuracy?

Yes — documented guidance specifically recommends pneumatically actuated or two-stage solenoid valves over simple gate valves specifically because of their faster, more predictable response time when switching stages; this documented recommendation reflects that a slow or inconsistent valve response undermines the precision the two-speed fill strategy is intended to provide, regardless of how accurately the flow signal itself is being measured.

Is a purely on/off two-stage valve control approach to batching accurately described as "PID control," or is that a common documented misconception?

Documented industry analysis specifically flags this as a common misunderstood terminology issue — discrete two-stage (fast/dribble) valve control is fundamentally simple on-off logic with staged setpoints, which documented analysis distinguishes from genuine continuous PID control; describing a staged on-off batching approach as "PID" is documented as a frequent but technically inaccurate mischaracterization of the actual control algorithm being used.

Beyond flow meter selection, is there a documented feedback-based method for correcting fill rate accuracy over time in feeder-based (rather than valve-based) batching systems?

Yes — documented guidance specifically describes integrating a load cell with a feeder to enable automatic correction: when the actual feeding rate deviates from the target, the system can adjust motor speed to bring output back toward the target, functioning as a documented closed-loop correction distinct from simply relying on an open-loop flow meter reading alone.