LT DIN Rail Analog Transmitters with Serial Data Communication and Analog outputs for RTD Temperature Applications P/N LT20P392F

LT DIN Rail Analog Transmitters with Serial Data Communication and Analog outputs for RTD Temperature Applications

Price: $334.00
  • P/NLT20P392F
- +

Features

  • Reads 100Ω platinum, 10Ω copper & 120Ω nickel RTDs
  • Accuracy ±0.01%  of reading ±0.04°C (±0.07°F)
  • 2, 3 or 4-wire connection with lead resistance compensation
  • Selectable 1°, 0.1°, or 0.01° resolution, °C, °F, K or R
  • All input ranges are user selectable and factory calibrated
  • Up to 60 conversions per second, Ideal for peak or valley capture
  • 4-20 mA, 0-20 mA, 0-10V or -10V to +10V transmitter output, (isolated)
  • Analog output resolution 0.0015%  of span, accuracy ±0.02%  of span
  • RS232 or RS485 serial data, Modbus or Laurel ASCII protocol (isolated)
  • Dual 120 mA solid state relays for alarm or control (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Operating temperature from -40°C to 70°C (-40°F to 158°F)

The Laureate™ LT Series DIN rail analog transmitter with serial data communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy of 0.01% of reading ± 2 counts, with high read rates at up to 60 or 50 conversions per second. The LT Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and RS232/RS485 output transmitter for RTD temperature offers the same high performance, signal conditioning and programmable features as Laureate digital panel meters, counters & timers provides six voltage input ranges and four current input ranges, all factory calibrated and jumper selectable. The temperature transmitter provides a linearized, highly accurate, stable and repeatable transmitter output for 100 ohm platinum, 10 ohm copper and 120 ohm nickel RTDs. Pt100 platinum RTDs can have a DIN alpha of 0.00385 or ANSI alpha of 0.00392. The RTD type and temperature range, specified in °C or °F, are user-selectable. The temperature range can be as wide as the entire span of the RTD type or as narrow as 150 counts (such as 15.0°), limited only by considerations of electrical noise and digital filtering time constants.

RTD connections can be via 2, 3 or 4 wires. With 3 or 4-wire connections, the transmitter automatically compensates for changes in lead resistance to the sensor. With 2-wire connection, the transmitter can measure and then subtract the lead wire resistance.

All RTD types are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Calibration data is stored in EEPROM on the signal conditioner board. This allows signal conditioner boards and ranges to be changed in the field with no need for recalibration. Typical accuracy for a Pt100 is better than ±0.04°C (±0.07°F) ±0.01%  of reading. Open sensor indication is standard and may be set up to indicate either upscale or downscale. RTD excitation is provided by the transmitter. RTD connections can be of the 2-, 3- or 4-wire type. With 3- and 4-wire connections, the transmitter automatically compensates for changes in lead resistance to the sensor. Instrument Setup Software allows user calibration for RTDs whose resistance is different than nominal at 0°C.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

High read rate of up to 50 or 60 conversions per second, the Laureate™ LT Series transmitter uses Concurrent Slope (US Pat. 5,262,780) analog-to-digital conversion to integrate signals over a full power line cycle (50 Hz or 60 Hz). This read rate enables peak and valley capture, real-time computer interfacing, and control applications. Peak and valley values are automatically captured and can be viewed using Laurel’s free Instrument Setup Software (compatible with Windows PCs) or transmitted as serial data.

Standard Features of Laureate LT Transmitters Include:

  • Serial communications output, (isolated), RS232 or RS485 (half or full duplex), jumper selectable. Three protocols are user selectable: Modbus RTU, Modbus ASCII, or Laurel ASCII. Modbus operation is fully compliant with Modbus Over Serial Line Specification V1.0 (2002). The Laurel ASCII protocol is simpler than the Modbus protocol and is recommended when all devices are Laureates.
  • 4-20 mA, 0-10V or -10V to +10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 0.0015% resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. Note that Ethernet data I/O is provided by Laurel's LTE series transmitters.
  • Dual solid state relays, (isolated), for alarm or control. Rated 120 mA at 130 Vac or 170 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA,  or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.

  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Two tare functions: auto-tare and manual tare. In auto-tare, an input line is grounded by an external pushbutton. This causes the current weight, which is normally the empty weight of the container to be stored in memory as an offset. In manual tare, the tare value can be entered manually via a control input pushbutton or using Laurel's free Instrument Setup Software.

Peak and valley values are automatically captured. These may be displayed via  Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

Removable screw terminal connections of Laurel transmitters

LT series DIN rail Transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LT series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an RS232 or RS485 serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LT Transmitters to a Local Area Network (LAN)

Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Alternatively, Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration.

Flexible Communication Options for LT Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laurel network with Ethernet-to-analog converter board

Ethernet & 4-20 mA Output Thermocouple Temperature Transmitter

RTD Metal Alpha R at 0°C R at top
of range
Excitation
Current
Range Max Error
Platinum
Pt100
0.003850 (DIN) 100Ω 390.48Ω
at 850°C
196 µA -202°C to +850°C
-331°F to +1562°F
±0.03°C ±0.01% of rdg
±0.05°F ±0.01% of rdg
Platinum
Pt100
0.003902 (ANSI) 100Ω 394.36Ω
at 850°C
196 µA -202°C to +850°C
-331°F to +1168°F
±0.04°C ±0.01% of rdg
±0.07°F ±0.01% of rdg
Nickel
Ni120
0.00672 120Ω 380.31Ω
at 260°C
196 µA -80°C to +260°C
-112°F to +500°F
±0.05°C ±0.01% of rdg
±0.09°F ±0.01% of rdg
Copper
Cu10
0.00427 9.035Ω 19.116Ω
at 260°C
5.0 mA -100°C to +260°C
-148°F to +500°F
±0.05°C ±0.01% of rdg
±0.09°F ±0.01% of rdg
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
RTD Input
Calibration, Pt 100 DIN IEC 751 (IPTS-68)
Calibration, Pt 100 ANSI NIST Monograph 126
Configuration 2, 3 or 4-wire connection
Excitation current 0.2 mA
Max error at 25°C, Pt100 ±0.04°C (±0.07°F) ±0.01% of reading
Span tempco ±0.003% of reading/°C
Zero tempco ±0.03 deg/deg
Sensor lead resistance 2-wire: 10 mdeg/Ω/deg up to 10Ω;
tempco per conductor 3 & 4-wire: 10 μdeg/Ω/deg up to 100Ω
Over-voltage protection 125 Vac
Open sensor indication 0 mA or > 20 mA output, selectable
Provision for user calibration Multiplier of RTD resistance plus offset in degrees
Analog Output (standard)
Output Levels 4-20 mA, 0-20 mA, 0-10 Vdc, -10 to +10Vdc (user selectable)
Compliance, 4-20 mA 10V (0-500Ω load)
Compliance, 0-10V 2 mA (5 kΩ load)
Output Resolution 16 bits (65,536 steps)
Output Accuracy ±0.02% of output span
Output Isolation 250V rms working, 2.3 kV rms per 1 minute test
Step response time 50 ms
Serial Data Output (standard)
Signal Types RS232 or RS485 (half or full duplex), jumper selectable
Data Rates 300, 600, 1200, 2400, 4800, 9600, 19200 baud
Output Isolation 250V rms working, 2.3 kV rms per 1 min test
Serial Protocols Modbus RTU, Modbus ASCII, Custom ASCII
Modbus Compliance Modbus over Serial Line Specification V1.0 (2002)
RS232/485 Connector  Screw terminals for easy daisy chaining 
Digital Addresses 247 for Modbus, 31 for Custom ASCII
Dual Relay Output (standard)
Relay Type Two solid state relays, SPST, normally open, Form A
Load Rating 120 mA at 140 Vac or 180 Vdc
Power Input
Standard Power 85-264 Vac or 90-300 Vdc
Low Power Option 10-48 Vdc or 12-32 Vac
Power Frequency DC or 47-63 Hz
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Power Consumption at 24V 1.5W typical
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Cooling Required Mount transmitters with ventilation holes at top and bottom. Leave 6 mm (1/4") between transmitters, or force air with a fan.
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options available using Laurel's free Instrument Setup Software.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

Laureate LT transmitter pinout

RTD hookup can be via 2, 3 or 4 wires to the J5 connector. The transmitter applies an excitation current of 196 µA (Pt 100 and Ni 120) or 5 mA (Cu 10).

4-wire hookup of RTD signal conditioner board In 4-wire hookup, different pairs of leads are used to apply the excitation current and sense the voltage drop across the RTD, so that the IR drop across the excitation leads is not a factor.
3-wire hookup of RTD signal conditioner board In 3-wire hookup, the transmitter senses the combined voltage drop across the RTD plus two excitation leads. It also senses the voltage drop across one excitation lead, and then subtracts twice this voltage from the combined total. This technique effectively subtracts all lead resistance and compensates for ambient temperature changes if the two excitation leads are identical.
2-wire hookup of RTD signal conditioner board In 2-wire hookup, the transmitter senses the combined voltage drop across the RTD and both lead wires. The voltage drop across the lead wires can be measured by shorting out the RTD during transmitter setup, and this voltage is then automatically subtracted from the combined total. However, changing resistance of the lead wires due to ambient temperature changes will not be compensated.

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

 

CAL-Analog

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LT20P392F
Price as Configured: $334.00

Click on the Option Board Links for More Product Information

Base Item
$164.00
Main Board
$0.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
$81.00
$81.00
$81.00
$81.00
$81.00
$81.00
Note: The same signal conditioner board can be user configured for all RTD types listed and °C or °F, as well as for resistance measurement
Note: All ranges are factory calibrated and user selectable
$81.00
Part Number as Configured:
LT20P392F
Price as Configured:
$334.00
Quantity:
- +
Extended Price:
$334.00

Understanding the Laureate™ LT Series DIN Rail Transmitter for RTD Temperature

The Laureate™ LT Series DIN rail transmitter for RTD temperature provides a linearized, highly accurate, stable, and repeatable output for 100 ohm platinum, 10 ohm copper, and 120 ohm nickel RTDs. Pt100 platinum RTDs can have a DIN alpha of 0.003850 or ANSI alpha of 0.003902. The RTD type and temperature range, specified in °C or °F, are user-selectable. The temperature range can be as wide as the entire span of the RTD type or as narrow as 150 counts (such as 15.0°), limited only by considerations of electrical noise and digital filtering time constants.

RTD Type Specifications

Platinum Pt100 (DIN, alpha 0.003850): 100Ω at 0°C, 390.48Ω at 850°C, 196 µA excitation, -202°C to +850°C range, max error ±0.03°C ±0.01% of reading. Platinum Pt100 (ANSI, alpha 0.003902): 100Ω at 0°C, 394.36Ω at 850°C, 196 µA excitation, -202°C to +850°C range, max error ±0.04°C ±0.01% of reading. Nickel Ni120 (alpha 0.00672): 120Ω at 0°C, 380.31Ω at 260°C, 196 µA excitation, -80°C to +260°C range, max error ±0.05°C ±0.01% of reading. Copper Cu10 (alpha 0.00427): 9.035Ω at 0°C, 19.116Ω at 260°C, 5.0 mA excitation, -100°C to +260°C range, max error ±0.05°C ±0.01% of reading.

Calibration Standards and Input Specifications

Pt100 DIN calibration is per IEC 751 (IPTS-68); Pt100 ANSI calibration is per NIST Monograph 126. Configuration is 2, 3, or 4-wire connection. Excitation current is 0.2 mA (196 µA for Pt100/Ni120, 5 mA for Cu10). Max error at 25°C for Pt100 is ±0.04°C (±0.07°F) ±0.01% of reading. Span tempco is ±0.003% of reading/°C; zero tempco is ±0.03 deg/deg. Sensor lead resistance tempco per conductor is 10 mdeg/Ω/deg up to 10Ω for 2-wire, or 10 µdeg/Ω/deg up to 100Ω for 3- and 4-wire. Overvoltage protection is 125 Vac. Open sensor indication is selectable as 0 mA or greater than 20 mA output. A provision for user calibration allows a multiplier of RTD resistance plus an offset in degrees, for RTDs whose resistance differs from nominal at 0°C.

2, 3, and 4-Wire RTD Hookup

In 4-wire hookup, different pairs of leads apply the excitation current and sense the voltage drop across the RTD, so the IR drop across the excitation leads isn't a factor. In 3-wire hookup, the transmitter senses the combined voltage drop across the RTD plus two excitation leads, and separately senses the drop across one excitation lead, then subtracts twice this voltage from the combined total — this subtracts all lead resistance and compensates for ambient temperature changes if the two excitation leads are identical. In 2-wire hookup, the transmitter senses the combined voltage drop across the RTD and both lead wires; lead-wire voltage drop can be measured by shorting out the RTD during setup and automatically subtracted, but changing lead-wire resistance from ambient temperature changes isn't compensated.

Factory-Calibrated Accuracy

All RTD types are factory-calibrated, with calibration factors stored in EEPROM on the signal conditioner board, allowing boards and ranges to be changed in the field without recalibrating the transmitter. The same signal conditioner board can be user-configured for all RTD types listed, °C or °F, and resistance measurement. Factory recalibration is recommended every 12 months.

Where RTD Temperature DIN Rail Transmitters Are Used

  • Precision Process Temperature Monitoring — high-accuracy Pt100 measurement for critical process control.
  • Long Cable Run Installations — 3- or 4-wire connection compensating for lead resistance.
  • Multi-Metal RTD Retrofit — a single transmitter field-configurable across Pt100, Ni120, and Cu10 sensors.
  • 4-20 mA Temperature Retransmission — linearized RTD output for PLC/SCADA integration.
  • HVAC & Building Automation — RTD-based temperature monitoring across distributed points.
  • Multi-Point RS485 Temperature Networks — daisy-chained transmitters reporting to a central controller.
  • OEM Precision Temperature Instrumentation — DIN rail integration into existing control panels.

RTD Temperature DIN Rail Transmitter Frequently Asked Questions

Why does the Cu10 RTD use a 5 mA excitation current while Pt100 and Ni120 use only 196 µA?

Documented specification specifically lists Cu10's nominal resistance as only 9.035Ω at 0°C, far lower than Pt100's or Ni120's 100Ω/120Ω — since a lower-resistance sensor produces a proportionally smaller voltage signal for a given excitation current, the documented higher 5 mA excitation current for Cu10 is what's needed to develop a comparably measurable voltage drop across that much smaller resistance.

Why does Pt100 DIN (alpha 0.003850) have a documented tighter max error (±0.03°C) than Pt100 ANSI (alpha 0.003902) at ±0.04°C, given both are Pt100 sensors?

These are documented as genuinely distinct calibration standards, not simply two names for the same curve — Pt100 DIN is calibrated per IEC 751 (IPTS-68) while Pt100 ANSI is calibrated per NIST Monograph 126, and the specific documented max error figure for each reflects how closely the transmitter's linearization can track that particular standard's defined resistance-versus-temperature curve, rather than being inherently identical for any "Pt100" sensor regardless of standard.

Why does the documented lead-resistance tempco allow up to 100Ω for 3- and 4-wire connections, but only up to 10Ω for 2-wire?

Documented specification ties this specifically to how much of the lead resistance error each wiring method can actually compensate — since 3- and 4-wire configurations are documented as actively compensating for lead resistance (either by subtracting it directly or by using separate excitation and sense leads), they can tolerate a much higher lead resistance before that compensation's own limits are reached, whereas 2-wire's more limited compensation approach is documented as reliable only up to a lower 10Ω lead resistance.

Does the 2-wire hookup's method of subtracting lead-wire resistance (measured by shorting the RTD during setup) provide the same ongoing accuracy as 3- or 4-wire methods?

No — documented description specifically notes that while the initial lead-wire voltage drop can be measured and subtracted during 2-wire setup, subsequent changes in lead-wire resistance due to ambient temperature changes are not compensated by this method; 3- and 4-wire hookups are documented as actively compensating for lead resistance on an ongoing basis, which 2-wire's one-time subtraction at setup does not replicate.

Does the "provision for user calibration" (multiplier plus offset) apply automatically, or does it need to be deliberately configured for an RTD that's out of nominal tolerance?

It's documented as a deliberate configuration option specifically intended for RTDs whose resistance differs from nominal at 0°C — rather than the transmitter automatically detecting and correcting for such deviation, this user calibration provision (a multiplier applied to RTD resistance plus a degree offset) is a feature the user applies specifically when a particular sensor's actual characteristics diverge from the standard nominal curve.

Does open sensor indication behave the same way regardless of which RTD type (Pt100, Ni120, Cu10) is configured?

Documented specification lists open sensor indication (0 mA or greater than 20 mA, selectable) as a general specification under RTD Input, without separately distinguishing behavior by RTD type — this points to the open-sensor detection and output response being a shared mechanism across the supported RTD types rather than one specifically tailored per sensor type.

Can the temperature span be set narrower than the RTD's full documented range, and does doing so affect accuracy?

Yes, span can be set as narrow as 150 counts (such as 15.0°) — documented guidance specifically notes this narrow-span limit is governed by considerations of electrical noise and digital filtering time constants, meaning an appropriately narrow span is supported by design, though very narrow spans push closer to the practical noise floor these considerations describe.

Does changing which analog output level (4-20 mA versus 0-10V, for example) is selected affect the transmitter's RTD measurement accuracy?

No — documented specification lists RTD input accuracy figures (max error, span tempco, zero tempco) separately from the analog output section's own documented output accuracy (±0.02% of output span); the analog output stage converts whatever accurately-measured temperature value the RTD input stage has already determined, so selecting a different output level type doesn't change the underlying input measurement accuracy.

Does the transmitter's step response time of 50 ms apply to the RTD input measurement itself, or specifically to the analog output?

Documented specification lists the 50 ms step response time specifically under Analog Output — this describes how quickly the analog output signal settles to a new value following a step change in the underlying measurement, rather than describing the RTD input conversion timing itself, which is governed separately by the transmitter's documented Concurrent Slope™ conversion process.

Can the same physical transmitter be field-reconfigured between Pt100, Ni120, and Cu10 RTD types, or does changing metal type require different hardware?

Documented note specifically states the same signal conditioner board can be user-configured for all RTD types listed, °C or °F, and resistance measurement — this indicates configuration-level flexibility on shared hardware rather than requiring a separate physical transmitter model purchased for each individual RTD metal type.

Does field-replacing the signal conditioner board require recalibrating the transmitter afterward?

No — documented description specifically states calibration factors are stored in EEPROM on the signal conditioner board itself, which allows boards and ranges to be changed in the field with no need for recalibration; the calibration data travels with the board rather than being separately stored in the main transmitter unit requiring re-entry after a board swap.

Does the 0.2 mA figure listed under "Excitation current" in the RTD Input table apply to all four RTD types, or is it a general figure distinct from the per-type table?

These are documented in two separate places with different figures — the per-type table specifically lists 196 µA for Pt100 and Ni120, and 5.0 mA for Cu10, while the general RTD Input specification section separately lists "Excitation current: 0.2 mA," which is consistent with the rounded 196 µA figure used for the platinum and nickel types rather than applying uniformly to Cu10's documented 5 mA excitation.

RTD Self-Heating Error Questions From the Field

What specifically causes "self-heating error" in an RTD measurement?

Documented explanation specifically describes this as a direct consequence of passing excitation current through the RTD's own resistive element — since current flowing through any resistor generates heat (proportional to current squared times resistance), that self-generated heat raises the RTD's own temperature slightly above the actual temperature of whatever it's measuring, causing the sensor to read artificially high.

Is there a documented standard limit for how much self-heating error is considered acceptable?

Yes — documented guidance specifically cites IEC 60751 as stating that self-heating should not exceed 25% of the sensor's own tolerance specification, giving a concrete, standards-based benchmark for how much of a sensor's overall accuracy budget self-heating is permitted to consume, rather than leaving the acceptable error level undefined.

Why do smaller, thin-film RTD elements tend to be more susceptible to self-heating error than larger, wire-wound elements?

Documented comparison specifically cites typical self-heating coefficients around 2.5 mW/°C for small thin-film elements versus around 65 mW/°C for larger wire-wound elements — since a lower self-heating coefficient means a given amount of dissipated power produces a larger temperature rise, smaller thin-film elements are documented as generally more sensitive to a given excitation current's self-heating effect than bulkier wire-wound construction.

Does simply lowering excitation current always improve measurement accuracy overall, or is there a genuine tradeoff?

There's a genuine documented tradeoff — while lower excitation current reduces self-heating error, documented guidance specifically warns that setting current too low increases the risk that small voltage changes across the RTD translate into disproportionately large apparent temperature swings, since the signal-to-noise ratio of the measurement degrades; excitation current must be chosen to balance both self-heating and adequate signal resolution.

Can a documented worked example show roughly how much temperature error a specific self-heating scenario produces?

Yes — one documented reference design specifically calculates that with 1 mA excitation at maximum RTD resistance, power dissipation stays under 0.4 mW, keeping self-heating measurement error to less than 0.01°C for a typical thin-film element in that scenario — illustrating that appropriately chosen excitation current can keep self-heating error to a genuinely small, well-characterized fraction of a degree.

Does the RTD's physical mounting and surrounding environment affect how much self-heating error actually results from a given excitation current?

Yes — documented analysis specifically notes that the magnitude of self-heating error depends on how well the generated heat can be dissipated, which in turn depends on how the RTD is constructed and mounted; the same excitation current can produce different amounts of actual self-heating error depending on installation-specific thermal dissipation conditions, not the excitation current alone.

Are there documented alternative techniques beyond simply using low excitation current to reduce self-heating error?

Yes — documented approaches specifically include using narrow current pulses rather than continuous excitation, timing the measurement to complete faster than the RTD element's thermal rise time, which limits cumulative heat buildup during the brief measurement window even if the pulsed current magnitude itself isn't minimized to the same degree a continuous low-current approach would require.

Does 4-wire ratiometric measurement design specifically help address self-heating error, or only lead-resistance error?

Documented reference designs specifically address both concerns together in the same 4-wire ratiometric circuit — while the 4-wire topology itself is primarily aimed at eliminating lead-resistance error, documented design guidance for such circuits separately and specifically selects an excitation current magnitude chosen to keep self-heating error within a defined small figure, showing that a well-designed 4-wire measurement circuit deliberately manages both error sources together rather than the wiring topology alone solving the self-heating concern.