LTE DIN Rail Digital Transmitters for Batch Controller Pulse Input Applications

LTE DIN Rail Digital Transmitters for Batch Controller Pulse Input Applications

Price: $613.00
  • P/NLTE80FR
- +

Features

  • Provides automatic control for repetitive liquid fill operations
  • Available for turbine flow meter pulse signals from 0 Hz to 1 MHz or analog flow meter signals (4-20 mA, 0-1 mA or 0-10V)
  • ±999,999 for batch total, grand total, number of batches, or flow rate
  • Counts up from 0 to preset or down from preset to 0
  • Two or four control relays for with settable delay between cycles
  • Digital span adjustment: 0 to ±999,999; zero adjustment: -999,999 to +999,999
  • Up to 60 conversions per second, Ideal for peak or valley capture
  • Scalable to ±999,999 for use with current shunts
  • DIN rail mount housing, 22.5 mm wide, detachable screw-clamp connectors
  • Transducer excitation output, 5, 10, 12, or 24 Vdc (isolated)
  • Power 85-264 Vac / 90-300 Vdc or 10-48 Vdc / 12-32 Vac (isolated)
  • Operating temperature from -40°C to 70°C  (-40°F to 158°F)

The Laureate™ LTE Series DIN rail analog transmitter with ethernet communication and analog outputs for versatile connectivity.

The digitally programmable transmitter features two relays for alarm or control. The series offers exceptional accuracy with Input frequencies from 0.005 Hz to 1 MHz. The LTE Series transmitters offer the same high performance, signal conditioning, and programmable features as Laureate digital panel meters, counters, and timers.

The Laureate 4-20 mA, 0-20 mA, 0-10V or -10V to +10V and ethernet output transmitter for batch controller is a low cost, powerful and highly accurate batching controller for repetitive fill operations. It can use the Laureate FR dual channel pulse input signal conditioner for use with turbine flow meters. Relay control can be provided by two or four 8A contact relays, or by two or four 120 mA AC/DC solid state relays. Fill operations are repeated continually with a programmable delay from 10 ms to 199.99 sec, or based on an external control input.

Three items are tracked by the batch control software. These can each be scaled to engineering units of total or flow rate and displayed by the controller's six-digit LED display: Item #1 is the current batch total, which can be set up to count up from zero to a preset limit, or down from a preset limit to zero. Item #2 can be assigned to grand total or number of batches. Item #3 is the flow rate.

Batch Control with Turbine Flowmeters

The pulse-input batch controller utilizes the FR dual channel signal conditioner, which accepts pulses from turbine flow meters and most industrial transducers with a pulse output such as proximity switches with PNP or NPN output, TTL or CMOS logic, or magnetic pickup pulses down to 12 mV. The same signal is applied in parallel to the A and B input channels, which are used independently. Either channel can accept pulse rates from 0.005 Hz to 250 kHz, which exceeds the working range of turbine flow meters.

  • Channel A is used for totalizing. The measured total is scaled mathematically for control and display of volume in engineering units, such as liters.
  • Channel B is used for rate. The pulse frequency is determined by timing an integral number of periods over a specified gate time (plus 30 ms and 0-2 periods), and then taking the inverse of period. The inverse period approach allows much greater accuracy and faster update times than conventional rate meters which count signal pulses over a specified time interval. Update times can be as high as 25/sec. Rate in engineering units, such as liters per second, is obtained by multiplying the input by a scale factor.

The Extended Laureate computer board can display rate based on successive readings. It also allows exceptionally accurate custom curve linearization, for example to read out liquid volume or rate of flow in a horizontal cylindrical tank based on level reported by a 4-20 mA transmitter. For setup, up to 180 data points can be input into a computer spreadsheet or text file by the user. The computer then calculates spline-fit segments, which are downloaded into the transmitter.

Exceptional Accuracy and Stability. Laureate transmitters determine frequency by taking the inverse of period as measured with a calibrated quartz crystal time base. This results in extremely accurate and stable 6-digit internal readings (±999,999 counts), which are then processed in software. The analog output is generated by an ultra-linear 16-bit (65,536 step) digital-to-analog converter (DAC) for 0.02% output accuracy. The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For a 60 Hz signal, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

The update rate of the transmitter output is a programmed gate time + 30 ms + 0-2 signal periods. For pulse rates of 60 Hz and above, the update rate would be 20 per second. Such fast update rates are ideal for alarm and control.

All signal conditioner board ranges are factory-calibrated, with calibration factors for each range securely stored in an onboard EEPROM. These factors can be scaled via software to accommodate external shunts, enabling field replacement of signal conditioner boards without necessitating recalibration of the associated transmitter. For optimal accuracy, factory recalibration is recommended annually. All Laurel Electronics instruments undergo factory calibration using the industry-leading Fluke calibrators, which are recalibrated yearly and certified traceable to national standards, ensuring the highest level of precision and reliability.

Laureate Transmitters are easily programmed with Laurel’s free Instrument Setup Software, downloadable from our website and compatible with Windows PCs, requiring a data interface board for setup.

Standard Features of Laureate LTE Transmitters Include:

  • Ethernet I/O, (isolated). The supported protocols are Modbus RTU and ASCII, which are tunneled via Modbus TCP. Note that RS232 or RS485 data I/O is provided by Laurel's LT Series transmitters.
  • 4-20 mA, 0-10V or -10V to +10V analog transmitter output, (isolated), jumper-selectable and user scalable. All selections provide 0.0015% resolution of output span and 0.02%  output accuracy of a reading from -99,999 to +99,999 counts that is also transmitted digitally. Output isolation from signal and power grounds eliminates potential ground loop problems. Note that Ethernet data I/O is provided by Laurel's LTE series transmitters.
  • Dual solid state relays, (isolated), for alarm or control. Rated 120 mA at 130 Vac or 170 Vdc.
  • Selectable transducer excitation output, (isolated), user selectable 5V@100 mA, 10V@120 mA, 12V@100 mA,  or 24V@50 mA.
  • Power 85-264 Vac, (isolated), low-voltage 10-48 Vdc or 12-32 Vac power is optional.

Digital signal filtering modes can be selected to ensure stable readings in electrically noisy environments.
  • An unfiltered selection provides true peak and valley readings and aids in control applications.
  • A batch average filter selection averages each 16 conversions for an update every .266 seconds.
  • An adaptive moving average filter selection provides a choice of 8 time constants from 80 ms to 9.6 seconds. When a significant change in signal level occurs, the filter adapts by briefly switching to the shortest time to follow the change, then reverts back to its selected time constant. An Auto setting selects the time constant selection based on signal noise.

Peak and valley values are automatically captured. These may be displayed via Laurel's free Instrument Setup Software,  which runs on a PC under MS Windows or can be transmitted as serial data.

Two control inputs (CMOS/TTL levels, logic 0 = tied to digital ground, logic 1 = open) or dry contacts that can be set to control / activate 14 transmitter commands.

An (isolated) 5, 10, 12, or 24 Vdc excitation output is standard to power transducers or two-wire transmitters. Ratiometric operation, which automatically compensates for changes in the applied excitation, is jumper selectable for applications, such as bridges, where the signal to be measured is proportional to the excitation level.

Removable screw terminal connections of Laurel transmitters

LTE series DIN rail Transmitters & signal conditioners can be interfaced to a wide range of sensors and transducers using one of seven available plug-in signal conditioner boards. The transmitters duplicate the high performance (high accuracy, high read rate) and extensive programmable features of Laureate 1/8 DIN digital panel meters, counters and timers. They utilize the same signal conditioners boards, much of the same firmware, and Laurel's free Windows-based Instrument Setup Software. They come in a compact DIN rail mount package with detachable screw-clamp connectors for easy wiring.

The LTE series Transmitters accessible from this page include a 4-20 mA, 0-20 mA, 0-10V, or -10V to +10V analog output (isolated, user selectable), an RS232 or RS485 serial data interface (isolated, user selectable), and dual 120 mA solid state AC/DC relays (isolated). An (isolated) 5, 10, 12, or 24 Vdc transducer excitation output is included with all models other than those with a temperature or AC RMS signal conditioner.

Connecting Laureate LTE Transmitters to a Local Area Network (LAN)

Laurel LTE series Ethernet transmitters can connect directly to a LAN via an Ethernet cable. Up to 30 Laureate LT Transmitters and/or Digital Panel Meters can be configured for RS485 and daisy-chained to an LT Transmitter for seamless LAN integration. Setup for both configurations is streamlined using Laurel’s free Instrument Setup Software, which simplifies node discovery and transmitter configuration. 

Flexible Communication Options for LTE Transmitters

Laureate Transmitters can be equipped with Laurel communication boards to support various interfaces and protocols. These include serial interfaces with ASCII or Modbus RTU protocols, and Ethernet interfaces with web access, ASCII, or Modbus TCP/IP protocols, ensuring versatile connectivity for your commercial applications.

Laureate Ethernet network by Laurel Electronics

Pulse Inputs (FR signal conditioner)
Signal Types AC, pulses from NPN, PNP transistors, contact closures,magnetic pickups.
Ch A Frequency, Max 1 MHz
Ch B Frequency, Max 250 kHz
Signal Ground Common ground for channels A & B
Minimum Signal Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V)
Maximum Signal 250 Vac
Maximum Frequency 1 MHz, 30 kHz, 250 Hz (selectable)
Conversion Technique Inverse period
Delay between batches Selectable 10 ms to 199.99 s
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.
Power Supply Boards (one required)
Voltage, standard 85-264 Vac or 90-300 Vdc
Voltage, optional 12-32 Vac or 10-48 Vdc
Frequency DC or 47-63 Hz
Power consumption 1.2W @ 120 Vac, 1.5W @ 240 Vac, 1.3W @ 10 Vdc, 1.4W @ 20 Vdc,
   (typical, base meter) 1.55W @ 30 Vdc, 1.8W @ 40 Vdc, 2.15W @ 48 Vdc
Power Isolation 250V rms working, 2.3 kV rms per 1 min test
Excitation Output (standard)
5 Vdc 5 Vdc ± 5%, 100 mA (jumper selectable)
10 Vdc 10 Vdc ± 5%, 120 mA (jumper selectable)
12 Vdc 12 Vdc ± 5%, 100 mA (jumper selectable)
24 Vdc 24 Vdc ± 5%, 50 mA (jumper selectable)
Output Isolation 50 Vdc from signal ground
Analog Output Boards (one optional)
Output Levels 4-20 mA, 0-20 mA, 0-10V, -10 to +10V (single-output option)
Current compliance 4-20 mA, 0-20 mA, 0-10V (dual-output option)
Voltage compliance 2 mA at 10V ( > 5 kΩ load)
Scaling 12V at 20 mA ( < 600 Ω load)
Resolution Zero and full scale adjustable from -99999 to +99999 16 bits (0.0015% of full scale)
Isolation 250V rms working, 2.3 kV rms per 1 min test (dual analog outputs share the same ground)
Relay Output Boards (one required for batch control)
Dual magnetic relays 2 Form C, 8A max, 440Vac or 125Vdc max, 2500VA or 300W
Quad magnetic relays 4 Form A (NO), 8A max, 440Vac or 125Vdc max, 2500VA or 300W
Dual solid state relays 2 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Quad solid state relays 4 Form A (NO), AC or DC, 0V - 400V, 120Ma, 35Ohms (max at On-State)
Relay commons Isolated commons for dual relays or each pair of quad relays
Relay isolation 250V rms working, 2.3 kV rms per 1 minute test
Relay latching modes Latching or non-latching
Relay active modes Active on or off, active high or low
Hysteresis modes QA passband mode, split hysteresis, span hysteresis
Communication Boards (one optional)
Board Selections RS232RS485 with dual RJ11 connectors, RS485 with dual RJ45 connectors, USB, High-Speed Ethernet, USB-to-RS485 gatewayHigh-Speed Ethernet-to-RS485 gateway, WiFi with built-in antenna plus USB & RS485, WiFi with external antenna plus USB & RS485
Protocols Laurel Custom ASCII (serial), Modbus RTU (serial), Modbus TCP (Ethernet or WiFi)
Digital Addresses 247 (Modbus), 31 (Laurel ASCII),
Isolation 250V rms working, 2.3 kV rms per 1 min test
Environmental
Operating Temperature -40°C to 70°C (-40°F to 158°F)
Storage Temperature -40°C to 85°C (-40°F to 185°F)
Relative Humidity 95% at 40°C, non-condensing
Protection NEMA-4X (IP-65) when panel mounted
Signal Connections
Signal connections to Laurel pulse input batch controller
Mechanical
Enclosure Rugged black polycarbonate housing material
Mounting 35 mm rail per DIN EN 50022
Dimensions 129 x 104 x 22.5 mm case
Connectors Detachable screw clamp connectors meet VDE / IEC / UL / CSA standards. RJ45 jack for Ethernet
Tightening Torque Screw terminal connectors: 5 lb-in (0.56 Nm)
Weight Complete transmitter: 183 g (6.5 oz)
Replacement Case Screws
Size 6
Thread Pitch 6-19
Length 1/2"
Head Style Pan Head
Drive Style Phillips
Head Diameter 0.256-0.270
Head Height 0.087-0.097
Full/Partial Thread Full
Drive Size 2
Material Steel
Finished Black Oxide
General
Programming Utilize Laurel's free Instrument Setup Software, which runs on a PC under MS Windows. 
Security Lockout options include using the front panel buttons, the free Instrument Setup Software, or a hardware jumper.
Warranty 3 years parts & labor
Recalibration: All ranges are calibrated at the factory. Recalibration is recommended every 12 months.

Transmitter Pinout

Laureate LT transmitter pinout

 

Free Instrument Setup Software for Series 2 Laureates

Digital Panel Meter Laurel Electronics Digital Transmitters
1/8 DIN Digital Panel Meters DIN Rail Transmitters

Free Downloadable Windows-based Instrument Setup (IS) software (Data Interface Board Required) for use with our programmable Digital Panel Meters, Scale Meters, Counters, Timers, Remote Displays, and Transmitters, are an easy method to set up Laureate 1/8 DIN digital panel meters, counters, timers, remote displays, and DIN-rail transmitters, as explained in the Instrument Setup Software Manual. Laureate 1/8 DIN instruments can also be set up from the front panel, as explained in their respective Owners Manuals. Instrument Setup software is of benefit whether or not the PC is connected to the instrument.

  • When the PC is connected to the instrument, Instrument Setup software can retrieve the setup file from the instrument or open a default setup file or previously saved setup file from disk View Setup, then provides graphical user interface (GUI) screens with pull-down menus applicable to input, display, scaling, filtering, alarms, communications, analog output, and front panel lockouts. Fields that are not applicable to the instrument as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen for that item. After editing, the setup file can be downloaded, uploaded to the instrument, or saved to a disk. The same setup file can then be downloaded into multiple instruments.
  • When the PC is not connected to the instrument, the above GUI screens can be used to set up a virtual instrument. The setup file can then be saved to disk. Switching toView Menu then brings up a screen with the required front panel programming steps. This view can be printed out for use at the instrument site and to serve as a hard copy record.

    Download Free Instrument Setup Software


Installation

Set User Account Control (UAC) of MS Windows to "Never notifiy me" so that Instrument Setup Software can create directories. The UAC change screen can be reached as follows:

  • Under Windows 7, click on the Windows Start button in the lower left of the desktop and enter "UAC" in the search field.
  • Under Windows 8, navigate to Control Panel, then to the "User Accounts and Family Safety" section, and click on "Change User Account Control Settings."
  • Under Windows 10, click on the Windows Start button in the lower left of the desktop, then on "Settings", and enter "UAC" in the search field.
  • Reboot your computer for the changed UAC setting to take effect.
Meter board with USB Type-B connector

RJ11-to-DB9 cable with rear view of DB9 connector to PC

Laurel USB cable, P/N CBL05

RS232 cable, meter to PC, P/N CBL01

Laureate 1/8 DIN Laureate instruments must be equipped with a serial communications board and be connected to the computer via a serial communications cable. The connection can be via RS232, RS485, USB or Ethernet. Following setup, the serial communications board may be removed from the instrument if desired. The wiring of the RS232 cable is illustrated above with end views of the two connectors.

Laureate LT Series transmitters come standard with a 3-wire serial interface, which can be jumpered for RS232 or RS485.
Laureate LTE Series transmitters come standard with an Ethernet interface.

Meter Setup Screens

Click on any of the reduced screens below for a full-size screen view, then click on the Back button of your browser to return to this page. The screens examples below are for a fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS232. If the meter is a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different screens are brought up. Please see Series 1 setup screens.

Laurel Dual Channel Pulse Input Rate Meter
Welcome Screen
From the computer desktop, click on Start > Programs > IS2 > IS2. Or click on the IS icon on your desktop. This splash screen will be displayed for three seconds. The software revision number is in the lower right.
more
Setup Screen 02s for Digital Panel Meters and Digital Transmitters
Communications Selection Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 3 for Digital Panel Meters and Digital Transmitters
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port and Baud Rate, which should match the jumper setup of the instrument. Click on Establish. With the right settings, the Communications Established field will light up in green, and the Meter Type will be recognized. If so, click onMain Menu.
more
Setup Screen 4 for Digital Panel Meters and Digital Transmitters
Main Menu Screen
Click on File > Default Setup to retrieve the default setup file from disk for your type of meter. Click on File > Open Setupto retrieve a previously saved setup file from disk or on File > Save Setup to save your edited setup file to disk. Click onDPM > Get Setup to retrieve the setup file from your meter or on DPM > Put Setup to download your edited setup file into the meter.
more
Setup Screen 5 for Digital Panel Meters and Digital Transmitters
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on theInput+Display tab. You can now specify the meter hardware, signal type, display mode, and functions of control inputs A and B. Clicking on any item brings up a pull-down menu with the available choices.
more
Setup Screen 6 for Digital Panel Meters and Digital Transmitters
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 7 for Digital Panel Meters and Digital Transmitters
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital filter time constant (if any), the adaptive filter threshold, and whether Peak / Valley values are filtered or unfiltered. As for all setup screens, clicking on the F1 key while an item is highlighted brings up a Help screen for that item, as illustrated.
more
Setup Screen 8 for Digital Panel Meters and Digital Transmitters
DPM Relay Alarms Setup Screen
Click on the Relay Alarms tab, which allows you to set up Alarms 1 and 2 for the optional dual relay output board. Clicking on any of the four numeric fields changes these to green and brings up a special field to enter the desired numeric value, which is tied to the displayed reading.
more
Setup Screen 9 for Digital Panel Meters and Digital Transmitters
DPM Communications Setup Screen
Click on the Communications tab so set up serial communications. In particular, you can special the Serial Protocol and the meter address if multiple meters are to be addressed on the same serial data line.
more
Setup Screen 10 for Digital Panel Meters and Digital Transmitters
DPM Analog Output Setup Screen
Click on the Analog Out tab so set up the optional analog output board. Three output ranges are selectable, the endpoints of which can be tied to user-specified High and Low readings.
more
Setup Screen 11 for Digital Panel Meters and Digital Transmitters
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off menu items which will no longer be accessible from the front panel of the meter. This will simplify meter operation and prevent unintended setup changes.
more

Meter Setup Utilities

Setup Screen 12 for Digital Panel Meters and Digital Transmitters
DPM Front Panel Setup Screen
As an aid to programming the meter from the front panel when a serial connection is not available, you can return to the Main Menu and click on View > Menu. The required sequence of front panel screens will then be displayed. Click on any step in the sequence for the meaning of each digit, as illustrated for the FILtEr step. For a hardcopy, simply press on Print.
more
Setup Screen 13 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screen
Specify your desired communication protocol and the serial communications bus type, which should match the jumper setup of the instrument. Select None if the PC is not connected to the instrument.
more
Setup Screen 14 for Digital Panel Meters and Digital Transmitters
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented with the jumper positions and electrical connections for your selected board. This minimizes the need to refer to the printed manual.
more
Setup Screen 15 for Digital Panel Meters and Digital Transmitters
DPM Commands Screen
This page allows you set up external input, serial communications, an analog output proportional to the display (optional), and lockouts for Laureate digital counters. The grayed out area at the top right of the screen applies to Laureate remote displays.
more
Graphical Output Screens (not available with Ethernet)

From the Main Menu, click on Readings if your PC is connected to the meter. A pull-down menu then offers three choices: ListPlot and Graph.

  • List presents the latest readings in a 20-row by 10-column table. Press Pause at any time to freeze the display. This is one method to capture peak readings.   
  • Plot generates a plot of readings vs. time in seconds. It effectively turns the DPM-PC combination into a printing digital oscilloscope.
    more 
  • Graph generates a histogram where the horizontal axis is the reading and the vertical axis is the number of occurrences of readings. The display continually resizes itself as the number of readings increases.
    more
Setup Screen 18 for Digital Panel Meters and Digital Transmitters
DPM Calibration Screens
Click on the Scaling tab, which provides three scalClick on the Scaling tab, which provides three scaling methods to relate the signal to the displayed reading: 1) Scale and Offset method, 2) Coordinates of two points method, and 3) Reading Coordinates of Two Points method. The last method uses actual high and low signals, and the computer will prompt you.
more
Setup Screen 19 for Digital Panel Meters and Digital Transmitters
Frequency Meter Calibration Screen
Calibration of the quartz crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator. Apply the frequency, then enter the frequency in Hertz. Calibration will be automatic, with storage of the calibration factor stored in non-volatile memory.
more

 

Dimensions

Laurel transmitter case

Dimensioned CAD assembly drawings in EPRT, STEP, x_t, .dwg, pdf file formats: Laureate-transmitter-case.zip (zipping prevents browser from opening CAD files as text files).

 

Up/Down Totalizing
Dual-channel up/down totalizing using one channel for counts and one channel for count direction Up/down totalizing is provided by a mode of the Extended totalizing transmitter where pulses are either added or subtracted on Channel A based on a direction input on Channel B. The unit can also be programmed so that counting by Channel A is inhibited by an input on Channel B.
Combining Two Totals
Dual-channel up/down totalizing, where two channels are added or subtracted. A+B, A-B, A/B and AxB arithmetic functions are available with the Extended totalizing transmitter. A+B can sum two totals, while A-B subtracts the outflow total from inflow total. A/B ratio applied to two totals helps assure the proper mixing of components.
Up or Down Counting with Preset
Two repetitive fill operations handled by a single Laureate dual-channel digital counter and totalizer A single Laureate dual-channel totalizing transmitter will handle two repetitive fill operations by counting from zero up to a preset, or down from a preset to zero. The dual relay option is required.
Machine ON Time and Utilization
Dual-channel up/down totalizing to measure machine ON time by counting power line cycles. An easy way to measure the ON time of machines is to count AC line cycles and scale the total to hours. To display machine utilization or duty cycle in percent, use the Extended totalizing transmitter. Connect Channel A to switched AC and Channel B to the AC line, and apply a 100 multiplier to the A/B ratio.
Custom Curve Linearization
Displaying rate and total in the same Laureate totalizer and rate meter or transmitter The Extended version of the Laureate dual channel totalizing transmitter can transmit scaled rate or total for the same channel at the push of a button, and alarm both the rate and total. The Extended version can also do curve linearization, thereby extending the working range and accuracy of flow transducers.

 

 

CAL-Digital

Certificate of Calibration

$65.00

CBL02

USB-to-RS232 Adapter Cable

$47.00

CBL04

RS232 Cable for LT Transmitters

$47.00

CBL12

12-foot Power Cable

$47.00

CBL6

6-foot Power Cable

$41.00
Ordering Guide
Part Number as Configured: LTE80FR
Price as Configured: $613.00

Click on the Option Board Links for More Product Information

Base Item
$388.00
Main Board
$55.00
Power (Isolated)
$89.00
$89.00
Signal Input (Isolated)
$81.00
Part Number as Configured:
LTE80FR
Price as Configured:
$613.00
Quantity:
- +
Extended Price:
$613.00

Understanding the Laureate™ LTE Series DIN Rail Transmitter for Batch Controller Pulse Input

The Laureate™ LTE Series DIN rail transmitter for batch controller pulse input is a low cost, powerful, and highly accurate batching controller for repetitive fill operations. It uses the FR dual-channel pulse input signal conditioner for turbine flow meters and most industrial pulse-output transducers. Fill operations repeat continually with a programmable delay from 10 ms to 199.99 sec, or based on an external control input.

Item #1/#2/#3 Tracking

Three items are tracked by the batch control software, each scalable to engineering units and displayed on the controller's six-digit LED display. Item #1 is the current batch total, settable to count up from zero to a preset limit, or down from a preset limit to zero. Item #2 can be assigned to grand total or number of batches. Item #3 is the flow rate.

Dual-Channel Signal Conditioning

The same signal is applied in parallel to Channels A and B, used independently: either channel accepts pulse rates from 0.005 Hz to 250 kHz, exceeding the working range of turbine flow meters. Channel A is used for totalizing, scaled mathematically for volume in engineering units. Channel B is used for rate, with frequency determined by timing an integral number of periods over a specified gate time (plus 30 ms and 0-2 periods), then taking the inverse of period — this inverse-period approach allows much greater accuracy and faster update times (up to 25/sec) than conventional rate meters that count pulses over a fixed time interval.

Relay Options and Hysteresis Modes

Relay control is provided by two or four 8A contact relays, or two or four 120 mA AC/DC solid state relays. Relay latching modes are latching or non-latching; active modes are active on or off, active high or low. Hysteresis modes include QA passband mode, split hysteresis, and span hysteresis.

Communication Options

Communication boards support RS232, RS485, USB, High-Speed Ethernet, USB-to-RS485 gateway, High-Speed Ethernet-to-RS485 gateway, and WiFi options. Protocols supported are Laurel Custom ASCII (serial), Modbus RTU (serial), and Modbus TCP (Ethernet or WiFi), with digital addresses 247 for Modbus and 31 for Laurel ASCII.

Custom Curve Linearization

The Extended Laureate computer board can display rate based on successive readings and allows exceptionally accurate custom curve linearization — for example, reading out liquid volume or flow rate in a horizontal cylindrical tank based on level reported by a 4-20 mA transmitter. Up to 180 data points are entered into a spreadsheet or text file; the computer calculates spline-fit segments, which are downloaded into the transmitter.

Where Batch Controller Pulse Input DIN Rail Transmitters Are Used

  • Chemical & Liquid Batch Dosing — precise repetitive fill control via turbine flow meter pulses.
  • Multi-Component Mixing — combined-total ratio checks for accurate batch composition.
  • Dual-Station Fill Lines — two independent preset-based fill operations on one transmitter.
  • High-Cycle Production Batching — relay type selection for repetitive fill cycles.
  • Multi-Point RS485/Ethernet Batching Networks — daisy-chained or networked transmitters reporting to a central controller.
  • OEM Batch Control Instrumentation — DIN rail integration into existing control panels.

Batch Controller Pulse Input DIN Rail Transmitter Frequently Asked Questions

Why does this LTE-series batch controller page document RS232, RS485, and USB communication board options, when other LTE Series pages document Ethernet as the standard, built-in interface?

Documented specification on this specific page lists a broader Communication Boards selection (RS232, RS485, USB, High-Speed Ethernet, and gateway/WiFi options) as an optional board selection, distinct from the fixed, standard Ethernet interface documented on several other LTE Series product pages — this genuine documented difference indicates this particular batch controller model's communication interface is configured via selectable optional boards rather than a single fixed Ethernet interface, so the specific communication options should be confirmed against this page's own documented table rather than assumed to match other LTE Series pages.

Why does the same input signal get applied in parallel to both Channel A and Channel B, rather than each channel receiving a separate signal?

Documented description specifically explains this is intentional — Channel A and Channel B are used independently on the same pulse signal specifically so that one channel (A) can be dedicated to totalizing while the other (B) is independently dedicated to rate calculation via the inverse-period method; running both functions off the same physical sensor signal, but processed through two independent channels, is what allows simultaneous total and rate tracking from a single flow meter.

Why does the inverse-period method for Channel B rate calculation offer better accuracy than counting pulses over a fixed time interval?

Documented explanation specifically contrasts these two approaches — a conventional method counting pulses over a fixed time window is limited by how many whole pulses fall within that window, while the inverse-period method times an integral number of periods precisely and calculates the inverse, which is documented as allowing both greater accuracy and faster update times than the conventional counting approach, particularly at lower pulse rates where a fixed time window might capture very few pulses.

Does QA passband mode, split hysteresis, and span hysteresis all serve the same underlying purpose, or do they address different control needs?

Documented specification lists these as three distinct, separately selectable hysteresis modes rather than variations of one function — while all three relate to how a relay's on/off transition points are defined around a setpoint, they're documented as separate configuration options, implying each is suited to a different specific application need (such as the QA passband mode's specific fit for pass/fail testing around a target value) rather than being interchangeable descriptions of the same behavior.

Does Item #2 need to be configured as either grand total or number of batches, or can both be tracked simultaneously on one transmitter?

Documented description specifically frames Item #2 as assignable to either grand total or number of batches, phrased as an either/or configuration choice — this indicates Item #2 is a single configurable slot that displays one of these two related but distinct values at a time, rather than the transmitter simultaneously displaying both as separate tracked items.

Why does the batch delay range specifically span from 10 ms up to 199.99 seconds rather than allowing arbitrarily short or long delays?

The documented range (10 ms to 199.99 s) represents the specified selectable delay window between batches, without further detail on why these specific bounds were chosen — practically, the lower bound reflects a delay short enough to be effectively negligible for most fill cycles, while the upper bound of just under 200 seconds provides ample settling time for even fairly slow batch processes, without requiring a separate external timer for longer delays.

Does choosing magnetic relays over solid state relays for batch control change the transmitter's documented accuracy or timing specifications?

No — documented specifications for the relay output boards (contact rating, isolation, latching modes, hysteresis modes) are listed separately from the transmitter's core measurement accuracy specifications (input frequency range, inverse-period timing, output update rate); the relay type selected affects load-switching capability and mechanical/solid-state tradeoffs, not the underlying pulse measurement and totalizing accuracy documented for the FR signal conditioner itself.

Can custom curve linearization be applied specifically to the batch total (Item #1), or only to the flow rate (Item #3)?

Documented capability describes custom curve linearization as extending the working range and accuracy of flow transducers generally, without restricting it to only rate or only total — since Item #1 (batch total) and Item #3 (rate) are both derived from the same underlying signal conditioning and scaling process, the documented linearization capability is consistent with correcting either the totalized or the rate-based reading, depending on where the transducer's nonlinearity actually needs correction.

Does the documented 8A magnetic contact relay rating apply per relay, or as a combined total across all relays on a quad relay board?

Documented specification lists 8A as the rating for each documented relay type (dual or quad magnetic relays), phrased as a per-relay maximum rather than a combined system total — since each documented magnetic relay is specified with its own individual 8A max, 440Vac/125Vdc max, and 2500VA/300W ratings, this is consistent with each relay contact independently rated at these documented levels rather than the ratings representing a shared budget across multiple relays on the same board.

Does the documented 250 kHz maximum pulse rate on either channel limit which turbine flow meters can be used with this batch controller?

In practice, no — documented guidance specifically notes this 250 kHz maximum exceeds the working range of turbine flow meters, meaning the vast majority of real turbine flow meter pulse outputs fall well within this documented ceiling; the 250 kHz figure is presented as generous headroom above typical turbine meter output frequencies rather than a practical constraint most installations would need to work around.

Electromechanical vs. Solid State Relay Contact Life Questions From the Field

What is a documented typical difference in switching-cycle lifespan between a small electromechanical relay and a solid state relay?

Documented industry comparison specifically cites a typical electrical life of around 100,000 switching cycles for a small electromechanical relay, compared to millions of cycles for a solid state relay used within its specifications; this documented order-of-magnitude difference is specifically attributed to the SSR's lack of physical contacts subject to erosion and wear.

Why does contact arcing specifically limit an electromechanical relay's usable lifespan in a way that doesn't apply to solid state relays?

Documented explanation specifically identifies contact arcing and erosion as a physical, cumulative wear mechanism unique to mechanical contacts closing and opening under load; solid state relays are documented as having no physical contacts to arc or erode, since switching is instead performed by semiconductor devices, which is specifically why documented SSR lifespan estimates aren't limited by this particular wear mechanism.

Does a solid state relay's documented longer lifespan come without any tradeoff compared to an electromechanical relay?

No — documented technical comparison specifically notes solid state relays generate substantially more heat than electromechanical relays of similar current rating at full rated current (one documented source specifically citing roughly 10 times more heat), meaning documented SSR longevity in practice depends on proper heat sinking; without adequate heat management, this documented heat generation can itself become a limiting factor in achieving the SSR's theoretical extended lifespan.

Is there a documented typical cycle-count figure specifically for a mechanical relay/contactor used in demanding, high-power applications like those in electric vehicles?

Yes — one documented industry source specifically cites mechanical relays (contactors) enduring up to around 20,000 ON-OFF cycles depending on usage in this application context, a figure documented as notably lower than the general small-EMR figure cited elsewhere (100,000 cycles), illustrating that documented cycle-life figures for electromechanical relays can vary meaningfully depending on the specific load and application conditions.

Do solid state relays offer a genuinely unlimited number of switching cycles in practice, or is this a documented theoretical claim with real limits?

Documented analysis specifically frames "unlimited cycles" as a theoretical characterization based on the absence of physical contacts, while separately noting that electrical and thermal stress can still limit real-world SSR lifespan in practice; this documented nuance means the "millions of cycles" or "virtually forever" figures cited for SSRs should be understood as describing the absence of mechanical wear specifically, not an absolute guarantee against all possible failure modes.

Does an electromechanical relay's mechanical contact bounce create any documented electrical noise or interference concern that a solid state relay avoids?

Yes — documented comparison specifically identifies electromechanical relays as creating electromagnetic noise from contact arcing that can interfere with power lines, while solid state relays are documented as operating silently, without contact bounce issues, and producing little electrical interference; this is a documented, genuine electrical-noise distinction between the two relay technologies, separate from their differing mechanical lifespans.

Is total upfront relay cost or total cost of ownership documented as the more relevant comparison factor between EMRs and SSRs for high-cycle batch control applications?

Documented cost comparison specifically distinguishes these two framings — solid state relays are documented as having a higher upfront purchase cost, but documented total cost of ownership analysis notes SSRs can save money over time in high-use or harsh applications specifically because of their longer service life and reduced maintenance/replacement needs, making total cost of ownership the more relevant comparison specifically for demanding, high-cycle-count applications like repetitive batch control.

Does the number of independent output contacts differ in a documented way between electromechanical and solid state relays?

Yes — documented technical comparison specifically notes solid state relays normally provide only a single output contact per device, while electromechanical relays can offer multiple output contacts (such as the documented Form C dual or Form A quad configurations); this is a genuine, documented structural difference relevant when a specific application requires multiple independent switched outputs from a single relay unit.